EnVision FLEX+, Mouse, High pH (Link), Product No. K8002, K8004. Intended for use in immunohistochemistry together with Autostainer Link Instruments.
Class I - DangerousWhat Should You Do?
- Check if you have this product: K8002 Lot no: 10112879 K8004 Lot no: 10111463
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Dako Denmark A/S
- Reason for Recall:
- The pH of Target Retrieval Solution in the kit measures 4-5 instead of the correct pH of 9.1.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
EnVision FLEX+, Mouse, High pH (Link), Product No. K8002, K8004. Intended for use in immunohistochemistry together with Autostainer Link Instruments.
Product Codes/Lot Numbers:
K8002 Lot no: 10112879 K8004 Lot no: 10111463
Distribution:
Distributed in: US
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2330-2016
Related Recalls
The Autostainer instrument may not have a syringe tray installed which could result in a small volume of excess buffer being applied to several slides. In addition to the syringe tray, the updated Basic User Guide will reduce the potential risk of false negative results in the case of leakage of buffer and/or reagents that may impact staining.
The expected life time of the CoverStainer slide racks (CS 10330) does not meet established performance requirements, originating from variations in the manufacturing process.