Aequalis Humeral Nail Targeting Jig, Product Code: 9020060. Orthopedic manual surgical instrument.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lots: DB0114069, DB0114349, CA46096, CA216052, DB0111264001, DB0112016, DB0113102.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Tornier, Inc
- Reason for Recall:
- Recall for the Aequalis IM Nail instrumentation set (Tray Number 9020000) due to several reports that the mast of the targeting jig (Part Number 9020060) is separating from the jig boom. Separation will impact alignment and can affect the ability to fixate Aequalis IM Nail screws.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Aequalis Humeral Nail Targeting Jig, Product Code: 9020060. Orthopedic manual surgical instrument.
Product Codes/Lot Numbers:
Lots: DB0114069, DB0114349, CA46096, CA216052, DB0111264001, DB0112016, DB0113102.
Distribution:
Distributed in: US, AL, AK, AZ, CA, CO, CT, FL, IL, IN, ME, MD, MI, MO, NC, OH, PA, TX
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2315-2015
Related Recalls
A specific lot of Latitude Humeral Trial Stems has the incorrect direct part marking. " Laser Marked: 9030103, Humeral Stem Trial 150mm Large / Right, Yellow color indicator (incorrect) " Device Actually: 9030092, Humeral Stem Trial 150mm Medium / Left, Red color indicator (correct)
Stryker Tornier Perform Humeral System, Reversed Insert, 45%, REF DWP2393, for shoulder arthroplasty
Tornier
Devices from one lot of Perform" Reversed Inserts may contain a locking ring that was assembled upside down which would prevent the humeral insert from securely seating in the stem or spacer.
One lot of Perform Humeral Nucleus may exceed the specification for bacterial endotoxin load (max 20 EU/device).