Impella RP Flex with SmartAssist intracardiac microaxial blood pump, Product Number 1000323
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI-DI: 00813502012811; Serial Numbers: 409541 409542 409544 411379 416645 416646 416647 416649 418933 418939 418940 419489 421953 421955 422974 422978 423999 425596 428726 428730 429616 429617 429623 429646 429648 429649 429650 429651 429652 429654 430363 430365 430366 430367 430923 430929 430931 431099 431106 431532 431535 431844 431846 431847 431848 431850 431868 431869 431870 432394 432395 432396 432397 432398 432400 432401 432402 432643 432644 432645 432648 432651 433755 433756 434540
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Abiomed, Inc.
- Reason for Recall:
- A higher than expected rate of thrombus formation or deposition has been observed. The risk is for thrombus formations or deposits on indwelling central venous lines or cannulas may break free and enter into the Impella RP Flex, resulting in reduced flow, loss of support, or hemolysis
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Impella RP Flex with SmartAssist intracardiac microaxial blood pump, Product Number 1000323
Product Codes/Lot Numbers:
UDI-DI: 00813502012811; Serial Numbers: 409541 409542 409544 411379 416645 416646 416647 416649 418933 418939 418940 419489 421953 421955 422974 422978 423999 425596 428726 428730 429616 429617 429623 429646 429648 429649 429650 429651 429652 429654 430363 430365 430366 430367 430923 430929 430931 431099 431106 431532 431535 431844 431846 431847 431848 431850 431868 431869 431870 432394 432395 432396 432397 432398 432400 432401 432402 432643 432644 432645 432648 432651 433755 433756 434540
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2281-2023
Related Recalls
Exposure to the low purge pressure occurrence may result in persistent low purge pressure alarms and, in some cases, interruption or loss of mechanical circulatory support.
Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console.
Potential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps.