Eon (Model 3716) IPG is a dual channel; 8-contact receiver designed to be connected to 4, 8, or 16 electrode leads or a lead extension. Product Usage: The Eon and Eon Mini Neurostimulation Systems are indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back surgery syndrome, intractable low back and leg pain.
Class I - DangerousWhat Should You Do?
- Check if you have this product: All lots of the Eon (Product Code 65-3716)
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Advanced Neuromodulation Systems Inc.
- Reason for Recall:
- As of 11/30/2011, St. Jude Medical has received 110 patient complaints of warmth or heating at the implantable pulse generators (IPGs) implant site during charging for the Eon IPG and 116 reports of similar symptoms for the Eon Mini IPG, respectively.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Eon (Model 3716) IPG is a dual channel; 8-contact receiver designed to be connected to 4, 8, or 16 electrode leads or a lead extension. Product Usage: The Eon and Eon Mini Neurostimulation Systems are indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back surgery syndrome, intractable low back and leg pain.
Product Codes/Lot Numbers:
All lots of the Eon (Product Code 65-3716)
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2280-2012
Related Recalls
Advanced Neuromodulation Systems, Brio DBS System implantable neurostimulator, IPG or R-IPG, Model 6788 Implantable Pulse Generator.
Advanced Neuromodulation Systems
Unintended Fluid ingress into the device header has been shown.
Eon Mini IPGs manufactured in April 2012 could potentially exhibit a sudden, brief surge in stimulation that would be felt by the patient as uncomfortable or painful. SJMs routine testing, monitoring, and analysis of manufacturing processes identified three lots of Eon Mini IPGs that the internal battery has the potential to come into contact with the internal microcontroller board.
As of 11/30/2011, St. Jude Medical has received 110 patient complaints of warmth or heating at the implantable pulse generators (IPGs) implant site during charging for the Eon IPG and 116 reports of similar symptoms for the Eon Mini IPG, respectively.