OEC Elite Systems with 9-inch Image Intensifier

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI-DI: 00840682121729 Serial Numbers: FSXXTX00180, FSXXXX00020, F9XXTX00007, F9XXTX00008, F9XXXX00021, FSXXTX00068, FSXXTX00303, FSXXXX00054, FSXXTX00333, FSXXTX00272, F9XXXX00046, FSXXXX00009, FSXXXX00005, FSXXTX00350, F9XXTX00025, FSXXTX00131, FSXXTX00103, FSXXTX00116, FSXXTX00073, FSXXXX00056, FSXXTX00120, FSXXTX00171, FSXXTX00307, FSXXXX00006, FSXXTX00153, FSXXTX00154, FSXXTX00214, FSXXTX00048, FSXXTX00062, FSXXTX00170, FSXXTX00172, FSXXTX00362, FSXXTX00368, FSXXXX00024, F9XXTX00202, FSXXTX00326, FSXXTX00260, FSXXTX00133, FSXXXX00011, FSXXTX00286, F9XXTX00045, FSXXTX00241, FSXXTX00179, FSXXTX00276, FSXXTX00396, FSXXTX00034, FSXXTX00359, FSXXTX00248, FSXXXX00057, FSXXTX00079, FSXXTX00274, FSXXXX00032, FSXXTX00331, FSXXTX00042, FSXXTX00114, FSXXXX00007, FSXXTX00345, FSXXTX00065, FSXXTX00173, FSXXTX00191, FSXXTX00123, FSXXTX00376, F9XXTX00238, FSXXTX00234, FSXXTX00107, FSXXTX00298, F9XXXX00050, FSXXXX00036, FSXXXX00039, FSXXTX00373, FSXXTX00338, FSXXTX00382, FSXXTX00045, FSXXTX00118, F9XXTX00155, F9XXTX00179, FSXXTX00305, F9XXTX00053, FSXXTX00119, FSXXTX00135, FSXXTX00136, FSXXTX00221, FSXXXX00063, FSXXTX00067, FSXXTX00244, FSXXXX00062, FSXXTX00060, F9XXTX00059, FSXXTX00204, F9XXTX00040, FSXXTX00036, FSXXTX00099, F9XXTX00149, F9XXXX00013, F9XXXX00028, F9XXTX00019, FSXXTX00327, F9XXTX00017, F9XXTX00018, FSXXTX00216, FSXXTX00414, FSXXTX00315, FSXXTX00209, F9XXTX00121, F9XXTX00153, FSXXTX00027, FSXXTX00158, FSXXTX00231, FSXXTX00405, FSXXTX00406, FSXXTX00072, F9XXTX00030, FSXXTX00413, F9XXTX00196, F9XXTX00157, F9XXTX00158, F9XXTX00241, F9XXTX00248, FSXXTX00053, FSXXTX00054, FSXXTX00323, FSXXXX00010, FSXXTX00407, FSXXTX00049, FSXXTX00193, FSXXTX00220, FSXXTX00149, FSXXTX00150, FSXXTX00208, FSXXTX00168, FSXXTX00240, FSXXXX00061, FSXXTX00029, FSXXXX00045, FSXXXX00049, F9XXTX00220, F9XXTX00226, F9XXTX00127, FSXXTX00178, FSXXXX00017, FSXXTX00316, FSXXXX00023, FSXXTX00108, FSXXTX00330, FSXXXX00008, FSXXXX00052, FSXXTX00177, FSXXXX00048, FSXXTX00037, FSXXTX00278, FSXXTX00280, FSXXXX00035, F9XXXX00060, F9XXXX00038, F9XXXX00039, FSXXTX00102, FSXXTX00098, F9XXTX00081, FSXXTX00165, FSXXTX00057, FSXXTX00088, FSXXTX00197, FSXXTX00198, FSXXTX00412, FSXXTX00013, FSXXTX00014, FSXXTX00015, FSXXTX00016, FSXXTX00047, FSXXTX00337, FSXXTX00003, FSXXTX00189, FSXXXX00018, FSXXXX00019, FSXXTX00223, FSXXTX00306, FSXXTX00129, F9XXTX00071, FSXXTX00091, FSXXTX00166, FSXXTX00106, FSXXTX00117, FSXXTX00217, F9XXTX00198, FSXXTX00033, FSXXTX00109, FSXXTX00141, FSXXTX00282, FSXXTX00328, FSXXTX00341, FSXXTX00347, FSXXXX00003, FSXXTX00277, F9XXTX00180, F9XXXX00018, F9XXXX00020, F9XXXX00041, F9XXXX00042, F9XXTX00174, FSXXTX00090, F9XXXX00066, FSXXTX00010, FSXXTX00302, FSXXTX00146, FSXXTX00121, F9XXTX00020, F9XXTX00063, F9XXTX00064, FSXXTX00270, FSXXTX00322, F9XXTX00048, FSXXTX00294, FSXXTX00258, FSXXTX00259, FSXXTX00159, F9XXTX00188, FSXXTX00218, FSXXTX00230, F9XXTX00251, FSXXTX00319, F9XXXX00068, FSXXXX00064, FSXXTX00087, F9XXTX00023, F9XXXX00067, FSXXTX00264, FSXXXX00001, FSXXXX00059, FSXXTX00233, FSXXTX00398, FSXXTX00364, FSXXTX00339, FSXXTX00365, FSXXTX00349, F9XXTX00195, FSXXTX00283, FSXXTX00325, F9XXTX00111, F9XXTX00113, F9XXTX00244, F9XXTX00253, FSXXTX00012, FSXXTX00059, F9XXTX00154, FSXXTX00143, FSXXTX00186, FSXXTX00267, FSXXTX00271, FSXXTX00344, FSXXTX00055, FSXXTX00128, F9XXTX00231, FSXXTX00212, FSXXTX00355, F9XXXX00043, FSXXTX00066, FSXXTX00321, FSXXXX00043, F9XXTX00146, F9XXTX00201, F9XXTX00024, F9XXXX00019, FSXXTX00390, FSXXTX00096, FSXXTX00384, FSXXXX00021, FSXXTX00367, FSXXTX00155, FSXXTX00138, FSXXTX00207, FSXXTX00232, FSXXTX00236, F9XXTX00010, F9XXTX00011, F9XXTX00039, F9XXTX00181, F9XXXX00012, F9XXXX00014, FSXXTX00006, FSXXTX00009, FSXXTX00132, FSXXTX00265, FSXXTX00395, FSXXTX00239, FSXXTX00190, FSXXTX00052, FSXXTX00249, FSXXTX00004, FSXXTX00031, FSXXTX00032, FSXXTX00127, FSXXTX00343, FSXXTX00354, F9XXTX00137, F9XXTX00147, FSXXTX00295, F9XXTX00082, F9XXTX00044, FSXXTX00360, FSXXTX00374, F9XXTX00257, F9XXTX00112, F9XXTX00156, F9XXTX00085, FSXXTX00224, F9XXTX00159, FSXXTX00074, FSXXXX00066, FSXXTX00358, F9XXXX00051, FSXXTX00242, FSXXTX00213, FSXXXX00041, FSXXXX00053, F9XXTX00239, F9XXXX00036, FSXXTX00038, FSXXTX00317, F9XXTX00197, F9XXTX00041, F9XXTX00150, FSXXXX00027, FSXXTX00007, FSXXTX00332, FSXXTX00389, F9XXTX00107, FSXXTX00175, FSXXTX00195, FSXXTX00008, FSXXXX00033, FSXXTX00039, FSXXTX00030, FSXXTX00289, FSXXXX00026, F9XXTX00016, FSXXTX00163, FSXXTX00187, FSXXTX00261, FSXXTX00161, FSXXTX00100, FSXXTX00101, FSXXTX00095, FSXXTX00329, F9XXXX00037, F9XXTX00254, FSXXTX00263, F9XXXX00040, FSXXTX00071, FSXXXX00040, FSXXTX00035, FSXXTX00182, F9XXTX00177, FSXXXX00058, FSXXTX00352, FSXXTX00093, FSXXTX00082, FSXXTX00222, FSXXTX00112, F9XXTX00043, FSXXTX00069, FSXXTX00130, FSXXTX00134, FSXXTX00194, FSXXTX00199, FSXXTX00304, FSXXTX00125, FSXXTX00126, FSXXTX00361, FSXXTX00162, FSXXTX00288, FSXXTX00293, F9XXTX00009, F9XXXX00047, FSXXTX00342, FSXXTX00392, F9XXTX00218, F9XXTX00240, FSXXTX00185, F9XXTX00216, FSXXTX00061, FSXXTX00089, FSXXTX00144, FSXXTX00381, F9XXTX00243, FSXXTX00174, F9XXTX00242, FSXXTX00063, FSXXTX00064, FSXXTX00420, F9XXTX00031, FSXXTX00011, FSXXTX00164, FSXXTX00415, FSXXTX00416, FSXXTX00148, FSXXTX00375, F9XXTX00084, FSXXTX00046, FSXXTX00201, FSXXXX00065, F9XXXX00052, F9XXXX00053, FSXXXX00014, F9XXTX00138, FSXXTX00019, FSXXTX00084, FSXXTX00085, FSXXTX00269, FSXXTX00211, FSXXXX00029, FSXXTX00273, FSXXTX00372, FSXXTX00210, FSXXTX00092, FSXXTX00139, FSXXTX00243, F9XXTX00015, FSXXTX00238, FSXXTX00167, FSXXTX00157, FSXXTX00320, FSXXTX00394, F9XXTX00004, FSXXTX00005, FSXXTX00122, FSXXTX00237, FSXXTX00335, FSXXTX00137, FSXXTX00156, FSXXTX00215, FSXXTX00404, F9XXTX00217, F9XXTX00219, FSXXTX00113, FSXXTX00235, F9XXTX00027, F9XXTX00132, FSXXXX00034, FSXXTX00183, FSXXTX00043, FSXXTX00192, FSXXTX00336, F9XXTX00109, FSXXTX00299, F9XXTX00215, F9XXXX00033, F9XXXX00055, F9XXXX00056, FSXXTX00296, F9XXTX00099, F9XXTX00143, F9XXTX00144, F9XXTX00151, F9XXTX00161, F9XXTX00175, F9XXTX00178, F9XXTX00190, FSXXTX00203, FSXXTX00357, FSXXXX00055, F9XXTX00037, F9XXXX00023, F9XXXX00027, F9XXTX00098, F9XXTX00105, F9XXXX00049, FSXXTX00279, FSXXTX00290, FSXXTX00028, FSXXTX00219, FSXXTX00266, FSXXXX00025, F9XXXX00045, FSXXXX00050, FSXXTX00200, FSXXTX00371, FSXXTX00318, FSXXTX00385, FSXXXX00016, FSXXTX00245, FSXXTX00246, FSXXTX00292, FSXXTX00409, FSXXTX00419, FSXXTX00309, FSXXTX00334, F9XXTX00125, F9XXTX00097, F9XXTX00228, F9XXTX00038, FSXXTX00297, FSXXTX00408, FSXXTX00383, FSXXTX00247, FSXXTX00356, FSXXTX00399, FSXXTX00115, F9XXXX00065, FSXXXX00042, F9XXTX00078, F9XXTX00148, F9XXTX00221, F9XXTX00235, F9XXTX00236, FSXXTX00051, FSXXTX00058, FSXXTX00377, FSXXTX00387, FSXXTX00410, FSXXTX00417, FSXXTX00196, FSXXTX00275, FSXXXX00051, FSXXTX00301, FSXXTX00262, FSXXTX00397, FSXXTX00285, FSXXTX00353, FSXXTX00363, FSXXTX00366, F9XXTX00046, F9XXTX00160, F9XXTX00176, FSXXTX00340, FSXXTX00346, F9XXTX00122, F9XXTX00068, FSXXTX00268, FSXXTX00041, FSXXTX00142, FSXXTX00386, FSXXTX00284, FSXXTX00308, FSXXTX00202, FSXXTX00181, FSXXXX00047, FSXXTX00422, FSXXTX00086, FSXXTX00104, FSXXTX00056, F9XXXX00034, FSXXTX00169, F9XXTX00126, FSXXXX00013, FSXXXX00015, FSXXTX00351, FSXXXX00037, FSXXTX00300, FSXXTX00348, FSXXTX00391, FSXXTX00281, FSXXXX00038, FSXXTX00418, FSXXXX00044, FSXXTX00094, FSXXTX00097, F9XXTX00145, F9XXTX00200, F9XXXX00035, F9XXTX00021, F9XXTX00022, FSXXXX00002, FSXXTX00078, FSXXTX00070, FSXXTX00393, FSXXTX00105, FSXXXX00028, FSXXXX00031, F9XXTX00095, F9XXTX00255, FSXXXX00022, F9XXTX00123, F9XXTX00088, F9XXTX00087, FSXXTX00310, F9XXTX00199, F9XXTX00056, F9XXTX00069, FSXXTX00111, FSXXTX00077, F9XXTX00172, F9XXTX00185, F9XXTX00191, F9XXTX00205, F9XXTX00259, FSXXTX00369, F9XXTX00003, F9XXTX00058, F9XXTX00089, F9XXTX00094, F9XXTX00116, F9XXXX00048, FSXXXX00046, F9XXTX00100, F9XXXX00061, F9XXXX00001, F9XXTX00050, F9XXXX00032, F9XXXX00069, F9XXTX00194, F9XXTX00230, F9XXTX00252, F9XXTX00035, F9XXTX00036, F9XXTX00049, F9XXTX00142, F9XXTX00222, F9XXTX00232, F9XXXX00016, FSXXTX00021, FSXXTX00023, FSXXTX00024, FSXXTX00400, FSXXTX00022, FSXXTX00083, FSXXTX00160, F9XXXX00062, F9XXXX00063, FSXXTX00225, FSXXTX00257, FSXXTX00388, FSXXTX00176, FSXXTX00256, FSXXTX00380, FSXXTX00050, FSXXTX00205, FSXXTX00025, F9XXTX00171, FSXXTX00147, FSXXTX00403, F9XXTX00014, F9XXTX00029, F9XXTX00108, F9XXTX00130, F9XXTX00131, F9XXXX00064, F9XXTX00073, F9XXTX00074, F9XXTX00075, F9XXTX00028, F9XXTX00167, F9XXTX00135, F9XXTX00204, F9XXTX00152, F9XXXX00003, F9XXXX00007, FSXXXX00030, F9XXTX00057, FSXXTX00075, FSXXTX00018, F9XXTX00246, F9XXTX00213, F9XXTX00076, F9XXTX00224, F9XXTX00245, F9XXXX00044, F9XXXX00029, F9XXXX00022, F9XXXX00011, F9XXXX00008, FSXXTX00020, F9XXXX00025, F9XXTX00170, FSXXTX00152, F9XXTX00106, F9XXTX00139, FSXXTX00312, FSXXTX00252, FSXXXX00012, FSXXTX00081, FSXXTX00145, FSXXTX00151, FSXXXX00060, F9XXTX00203, FSXXTX00311, FSXXTX00378, FSXXTX00401, FSXXTX00402, FSXXTX00421, FSXXTX00076, F9XXTX00083, FSXXTX00206, FSXXTX00080, FSXXTX00229, F9XXTX00209, F9XXTX00208, F9XXTX00054, F9XXTX00065, F9XXTX00173, F9XXTX00193, F9XXTX00128, F9XXTX00129, F9XXTX00062, F9XXTX00067, F9XXTX00114, F9XXTX00162, F9XXTX00192, F9XXTX00133, F9XXTX00066, F9XXTX00115, F9XXTX00134, F9XXTX00164, F9XXTX00212, FSXXTX00140, F9XXTX00117, F9XXTX00227, FSXXTX00044, FSXXTX00040, F9XXTZ00001, F9XXTX00072, FSXXTX00026, F9XXXX00057, F9XXXX00058, F9XXXX00059, F9XXXX00004, F9XXTX00042, F9XXTX00005, F9XXTX00012, F9XXTX00013, F9XXXX00024, FSXXTX00226, FSXXTX00227, FSXXTX00250, FSXXTX00253, FSXXTX00254, FSXXXX00067, FSXXTX00228, FSXXTX00184, FSXXTX00287, FSXXTX00370, FSXXTX00379, FSXXXX00068, F9XXTX00006, FSXXTX00324, F9XXTX00233, F9XXTX00237, FSXXTX00313, FSXXTX00017, F9XXTX00165, F9XXTX00096, F9XXTX00234, FSXXTX00314, F9XXTX00077, F9XXTX00258, FSXXTX00110, F9XXTX00136, F9XXTX00141, F9XXTX00070, F9XXTX00214, FSXXTX00124, FSXXTX00188, F9XXTX00093, F9XXXX00054, F9XXTX00140, F9XXTX00186, F9XXTX00079, F9XXTX00223, F9XXTX00187, F9XXTX00086, F9XXXX00070, F9XXTX00225, F9XXTX00184, F9XXTX00206, F9XXTX00207, F9XXTX00256, F9XXTX00080, F9XXTX00103, F9XXTX00104, F9XXXX00015, F9XXXX00017, F9XXTX00061, F9XXTX00166, F9XXTX00052, F9XXTX00026, F9XXTX00183, FSXXTX00251, F9XXTX00060, F9XXTX00189, F9XXTX00211, F9XXXX00006, F9XXXX00031, F9XXTX00034, F9XXTX00055, F9XXTX00249, F9XXTX00101, F9XXTX00102, F9XXTX00110, F9XXTX00119, F9XXTX00090, F9XXTX00091, F9XXTX00092, F9XXTX00210, F9XXXX00009, F9XXXX00010, F9XXTX00169, F9XXTX00051, F9XXTX00118, F9XXTX00120, FSXXTX00291, F9XXTX00247, F9XXTX00032, FSXXTX00255, F9XXTX00182, F9XXXX00030, F9XXTX00033, F9XXTX00047, F9XXTX00124, F9XXTX00168, F9XXTX00250, F9XXXX00026, F9XXTX00229, F9XXTX00163
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
GE OEC Medical Systems, Inc
Reason for Recall:
Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

OEC Elite Systems with 9-inch Image Intensifier

Product Codes/Lot Numbers:

UDI-DI: 00840682121729 Serial Numbers: FSXXTX00180, FSXXXX00020, F9XXTX00007, F9XXTX00008, F9XXXX00021, FSXXTX00068, FSXXTX00303, FSXXXX00054, FSXXTX00333, FSXXTX00272, F9XXXX00046, FSXXXX00009, FSXXXX00005, FSXXTX00350, F9XXTX00025, FSXXTX00131, FSXXTX00103, FSXXTX00116, FSXXTX00073, FSXXXX00056, FSXXTX00120, FSXXTX00171, FSXXTX00307, FSXXXX00006, FSXXTX00153, FSXXTX00154, FSXXTX00214, FSXXTX00048, FSXXTX00062, FSXXTX00170, FSXXTX00172, FSXXTX00362, FSXXTX00368, FSXXXX00024, F9XXTX00202, FSXXTX00326, FSXXTX00260, FSXXTX00133, FSXXXX00011, FSXXTX00286, F9XXTX00045, FSXXTX00241, FSXXTX00179, FSXXTX00276, FSXXTX00396, FSXXTX00034, FSXXTX00359, FSXXTX00248, FSXXXX00057, FSXXTX00079, FSXXTX00274, FSXXXX00032, FSXXTX00331, FSXXTX00042, FSXXTX00114, FSXXXX00007, FSXXTX00345, FSXXTX00065, FSXXTX00173, FSXXTX00191, FSXXTX00123, FSXXTX00376, F9XXTX00238, FSXXTX00234, FSXXTX00107, FSXXTX00298, F9XXXX00050, FSXXXX00036, FSXXXX00039, FSXXTX00373, FSXXTX00338, FSXXTX00382, FSXXTX00045, FSXXTX00118, F9XXTX00155, F9XXTX00179, FSXXTX00305, F9XXTX00053, FSXXTX00119, FSXXTX00135, FSXXTX00136, FSXXTX00221, FSXXXX00063, FSXXTX00067, FSXXTX00244, FSXXXX00062, FSXXTX00060, F9XXTX00059, FSXXTX00204, F9XXTX00040, FSXXTX00036, FSXXTX00099, F9XXTX00149, F9XXXX00013, F9XXXX00028, F9XXTX00019, FSXXTX00327, F9XXTX00017, F9XXTX00018, FSXXTX00216, FSXXTX00414, FSXXTX00315, FSXXTX00209, F9XXTX00121, F9XXTX00153, FSXXTX00027, FSXXTX00158, FSXXTX00231, FSXXTX00405, FSXXTX00406, FSXXTX00072, F9XXTX00030, FSXXTX00413, F9XXTX00196, F9XXTX00157, F9XXTX00158, F9XXTX00241, F9XXTX00248, FSXXTX00053, FSXXTX00054, FSXXTX00323, FSXXXX00010, FSXXTX00407, FSXXTX00049, FSXXTX00193, FSXXTX00220, FSXXTX00149, FSXXTX00150, FSXXTX00208, FSXXTX00168, FSXXTX00240, FSXXXX00061, FSXXTX00029, FSXXXX00045, FSXXXX00049, F9XXTX00220, F9XXTX00226, F9XXTX00127, FSXXTX00178, FSXXXX00017, FSXXTX00316, FSXXXX00023, FSXXTX00108, FSXXTX00330, FSXXXX00008, FSXXXX00052, FSXXTX00177, FSXXXX00048, FSXXTX00037, FSXXTX00278, FSXXTX00280, FSXXXX00035, F9XXXX00060, F9XXXX00038, F9XXXX00039, FSXXTX00102, FSXXTX00098, F9XXTX00081, FSXXTX00165, FSXXTX00057, FSXXTX00088, FSXXTX00197, FSXXTX00198, FSXXTX00412, FSXXTX00013, FSXXTX00014, FSXXTX00015, FSXXTX00016, FSXXTX00047, FSXXTX00337, FSXXTX00003, FSXXTX00189, FSXXXX00018, FSXXXX00019, FSXXTX00223, FSXXTX00306, FSXXTX00129, F9XXTX00071, FSXXTX00091, FSXXTX00166, FSXXTX00106, FSXXTX00117, FSXXTX00217, F9XXTX00198, FSXXTX00033, FSXXTX00109, FSXXTX00141, FSXXTX00282, FSXXTX00328, FSXXTX00341, FSXXTX00347, FSXXXX00003, FSXXTX00277, F9XXTX00180, F9XXXX00018, F9XXXX00020, F9XXXX00041, F9XXXX00042, F9XXTX00174, FSXXTX00090, F9XXXX00066, FSXXTX00010, FSXXTX00302, FSXXTX00146, FSXXTX00121, F9XXTX00020, F9XXTX00063, F9XXTX00064, FSXXTX00270, FSXXTX00322, F9XXTX00048, FSXXTX00294, FSXXTX00258, FSXXTX00259, FSXXTX00159, F9XXTX00188, FSXXTX00218, FSXXTX00230, F9XXTX00251, FSXXTX00319, F9XXXX00068, FSXXXX00064, FSXXTX00087, F9XXTX00023, F9XXXX00067, FSXXTX00264, FSXXXX00001, FSXXXX00059, FSXXTX00233, FSXXTX00398, FSXXTX00364, FSXXTX00339, FSXXTX00365, FSXXTX00349, F9XXTX00195, FSXXTX00283, FSXXTX00325, F9XXTX00111, F9XXTX00113, F9XXTX00244, F9XXTX00253, FSXXTX00012, FSXXTX00059, F9XXTX00154, FSXXTX00143, FSXXTX00186, FSXXTX00267, FSXXTX00271, FSXXTX00344, FSXXTX00055, FSXXTX00128, F9XXTX00231, FSXXTX00212, FSXXTX00355, F9XXXX00043, FSXXTX00066, FSXXTX00321, FSXXXX00043, F9XXTX00146, F9XXTX00201, F9XXTX00024, F9XXXX00019, FSXXTX00390, FSXXTX00096, FSXXTX00384, FSXXXX00021, FSXXTX00367, FSXXTX00155, FSXXTX00138, FSXXTX00207, FSXXTX00232, FSXXTX00236, F9XXTX00010, F9XXTX00011, F9XXTX00039, F9XXTX00181, F9XXXX00012, F9XXXX00014, FSXXTX00006, FSXXTX00009, FSXXTX00132, FSXXTX00265, FSXXTX00395, FSXXTX00239, FSXXTX00190, FSXXTX00052, FSXXTX00249, FSXXTX00004, FSXXTX00031, FSXXTX00032, FSXXTX00127, FSXXTX00343, FSXXTX00354, F9XXTX00137, F9XXTX00147, FSXXTX00295, F9XXTX00082, F9XXTX00044, FSXXTX00360, FSXXTX00374, F9XXTX00257, F9XXTX00112, F9XXTX00156, F9XXTX00085, FSXXTX00224, F9XXTX00159, FSXXTX00074, FSXXXX00066, FSXXTX00358, F9XXXX00051, FSXXTX00242, FSXXTX00213, FSXXXX00041, FSXXXX00053, F9XXTX00239, F9XXXX00036, FSXXTX00038, FSXXTX00317, F9XXTX00197, F9XXTX00041, F9XXTX00150, FSXXXX00027, FSXXTX00007, FSXXTX00332, FSXXTX00389, F9XXTX00107, FSXXTX00175, FSXXTX00195, FSXXTX00008, FSXXXX00033, FSXXTX00039, FSXXTX00030, FSXXTX00289, FSXXXX00026, F9XXTX00016, FSXXTX00163, FSXXTX00187, FSXXTX00261, FSXXTX00161, FSXXTX00100, FSXXTX00101, FSXXTX00095, FSXXTX00329, F9XXXX00037, F9XXTX00254, FSXXTX00263, F9XXXX00040, FSXXTX00071, FSXXXX00040, FSXXTX00035, FSXXTX00182, F9XXTX00177, FSXXXX00058, FSXXTX00352, FSXXTX00093, FSXXTX00082, FSXXTX00222, FSXXTX00112, F9XXTX00043, FSXXTX00069, FSXXTX00130, FSXXTX00134, FSXXTX00194, FSXXTX00199, FSXXTX00304, FSXXTX00125, FSXXTX00126, FSXXTX00361, FSXXTX00162, FSXXTX00288, FSXXTX00293, F9XXTX00009, F9XXXX00047, FSXXTX00342, FSXXTX00392, F9XXTX00218, F9XXTX00240, FSXXTX00185, F9XXTX00216, FSXXTX00061, FSXXTX00089, FSXXTX00144, FSXXTX00381, F9XXTX00243, FSXXTX00174, F9XXTX00242, FSXXTX00063, FSXXTX00064, FSXXTX00420, F9XXTX00031, FSXXTX00011, FSXXTX00164, FSXXTX00415, FSXXTX00416, FSXXTX00148, FSXXTX00375, F9XXTX00084, FSXXTX00046, FSXXTX00201, FSXXXX00065, F9XXXX00052, F9XXXX00053, FSXXXX00014, F9XXTX00138, FSXXTX00019, FSXXTX00084, FSXXTX00085, FSXXTX00269, FSXXTX00211, FSXXXX00029, FSXXTX00273, FSXXTX00372, FSXXTX00210, FSXXTX00092, FSXXTX00139, FSXXTX00243, F9XXTX00015, FSXXTX00238, FSXXTX00167, FSXXTX00157, FSXXTX00320, FSXXTX00394, F9XXTX00004, FSXXTX00005, FSXXTX00122, FSXXTX00237, FSXXTX00335, FSXXTX00137, FSXXTX00156, FSXXTX00215, FSXXTX00404, F9XXTX00217, F9XXTX00219, FSXXTX00113, FSXXTX00235, F9XXTX00027, F9XXTX00132, FSXXXX00034, FSXXTX00183, FSXXTX00043, FSXXTX00192, FSXXTX00336, F9XXTX00109, FSXXTX00299, F9XXTX00215, F9XXXX00033, F9XXXX00055, F9XXXX00056, FSXXTX00296, F9XXTX00099, F9XXTX00143, F9XXTX00144, F9XXTX00151, F9XXTX00161, F9XXTX00175, F9XXTX00178, F9XXTX00190, FSXXTX00203, FSXXTX00357, FSXXXX00055, F9XXTX00037, F9XXXX00023, F9XXXX00027, F9XXTX00098, F9XXTX00105, F9XXXX00049, FSXXTX00279, FSXXTX00290, FSXXTX00028, FSXXTX00219, FSXXTX00266, FSXXXX00025, F9XXXX00045, FSXXXX00050, FSXXTX00200, FSXXTX00371, FSXXTX00318, FSXXTX00385, FSXXXX00016, FSXXTX00245, FSXXTX00246, FSXXTX00292, FSXXTX00409, FSXXTX00419, FSXXTX00309, FSXXTX00334, F9XXTX00125, F9XXTX00097, F9XXTX00228, F9XXTX00038, FSXXTX00297, FSXXTX00408, FSXXTX00383, FSXXTX00247, FSXXTX00356, FSXXTX00399, FSXXTX00115, F9XXXX00065, FSXXXX00042, F9XXTX00078, F9XXTX00148, F9XXTX00221, F9XXTX00235, F9XXTX00236, FSXXTX00051, FSXXTX00058, FSXXTX00377, FSXXTX00387, FSXXTX00410, FSXXTX00417, FSXXTX00196, FSXXTX00275, FSXXXX00051, FSXXTX00301, FSXXTX00262, FSXXTX00397, FSXXTX00285, FSXXTX00353, FSXXTX00363, FSXXTX00366, F9XXTX00046, F9XXTX00160, F9XXTX00176, FSXXTX00340, FSXXTX00346, F9XXTX00122, F9XXTX00068, FSXXTX00268, FSXXTX00041, FSXXTX00142, FSXXTX00386, FSXXTX00284, FSXXTX00308, FSXXTX00202, FSXXTX00181, FSXXXX00047, FSXXTX00422, FSXXTX00086, FSXXTX00104, FSXXTX00056, F9XXXX00034, FSXXTX00169, F9XXTX00126, FSXXXX00013, FSXXXX00015, FSXXTX00351, FSXXXX00037, FSXXTX00300, FSXXTX00348, FSXXTX00391, FSXXTX00281, FSXXXX00038, FSXXTX00418, FSXXXX00044, FSXXTX00094, FSXXTX00097, F9XXTX00145, F9XXTX00200, F9XXXX00035, F9XXTX00021, F9XXTX00022, FSXXXX00002, FSXXTX00078, FSXXTX00070, FSXXTX00393, FSXXTX00105, FSXXXX00028, FSXXXX00031, F9XXTX00095, F9XXTX00255, FSXXXX00022, F9XXTX00123, F9XXTX00088, F9XXTX00087, FSXXTX00310, F9XXTX00199, F9XXTX00056, F9XXTX00069, FSXXTX00111, FSXXTX00077, F9XXTX00172, F9XXTX00185, F9XXTX00191, F9XXTX00205, F9XXTX00259, FSXXTX00369, F9XXTX00003, F9XXTX00058, F9XXTX00089, F9XXTX00094, F9XXTX00116, F9XXXX00048, FSXXXX00046, F9XXTX00100, F9XXXX00061, F9XXXX00001, F9XXTX00050, F9XXXX00032, F9XXXX00069, F9XXTX00194, F9XXTX00230, F9XXTX00252, F9XXTX00035, F9XXTX00036, F9XXTX00049, F9XXTX00142, F9XXTX00222, F9XXTX00232, F9XXXX00016, FSXXTX00021, FSXXTX00023, FSXXTX00024, FSXXTX00400, FSXXTX00022, FSXXTX00083, FSXXTX00160, F9XXXX00062, F9XXXX00063, FSXXTX00225, FSXXTX00257, FSXXTX00388, FSXXTX00176, FSXXTX00256, FSXXTX00380, FSXXTX00050, FSXXTX00205, FSXXTX00025, F9XXTX00171, FSXXTX00147, FSXXTX00403, F9XXTX00014, F9XXTX00029, F9XXTX00108, F9XXTX00130, F9XXTX00131, F9XXXX00064, F9XXTX00073, F9XXTX00074, F9XXTX00075, F9XXTX00028, F9XXTX00167, F9XXTX00135, F9XXTX00204, F9XXTX00152, F9XXXX00003, F9XXXX00007, FSXXXX00030, F9XXTX00057, FSXXTX00075, FSXXTX00018, F9XXTX00246, F9XXTX00213, F9XXTX00076, F9XXTX00224, F9XXTX00245, F9XXXX00044, F9XXXX00029, F9XXXX00022, F9XXXX00011, F9XXXX00008, FSXXTX00020, F9XXXX00025, F9XXTX00170, FSXXTX00152, F9XXTX00106, F9XXTX00139, FSXXTX00312, FSXXTX00252, FSXXXX00012, FSXXTX00081, FSXXTX00145, FSXXTX00151, FSXXXX00060, F9XXTX00203, FSXXTX00311, FSXXTX00378, FSXXTX00401, FSXXTX00402, FSXXTX00421, FSXXTX00076, F9XXTX00083, FSXXTX00206, FSXXTX00080, FSXXTX00229, F9XXTX00209, F9XXTX00208, F9XXTX00054, F9XXTX00065, F9XXTX00173, F9XXTX00193, F9XXTX00128, F9XXTX00129, F9XXTX00062, F9XXTX00067, F9XXTX00114, F9XXTX00162, F9XXTX00192, F9XXTX00133, F9XXTX00066, F9XXTX00115, F9XXTX00134, F9XXTX00164, F9XXTX00212, FSXXTX00140, F9XXTX00117, F9XXTX00227, FSXXTX00044, FSXXTX00040, F9XXTZ00001, F9XXTX00072, FSXXTX00026, F9XXXX00057, F9XXXX00058, F9XXXX00059, F9XXXX00004, F9XXTX00042, F9XXTX00005, F9XXTX00012, F9XXTX00013, F9XXXX00024, FSXXTX00226, FSXXTX00227, FSXXTX00250, FSXXTX00253, FSXXTX00254, FSXXXX00067, FSXXTX00228, FSXXTX00184, FSXXTX00287, FSXXTX00370, FSXXTX00379, FSXXXX00068, F9XXTX00006, FSXXTX00324, F9XXTX00233, F9XXTX00237, FSXXTX00313, FSXXTX00017, F9XXTX00165, F9XXTX00096, F9XXTX00234, FSXXTX00314, F9XXTX00077, F9XXTX00258, FSXXTX00110, F9XXTX00136, F9XXTX00141, F9XXTX00070, F9XXTX00214, FSXXTX00124, FSXXTX00188, F9XXTX00093, F9XXXX00054, F9XXTX00140, F9XXTX00186, F9XXTX00079, F9XXTX00223, F9XXTX00187, F9XXTX00086, F9XXXX00070, F9XXTX00225, F9XXTX00184, F9XXTX00206, F9XXTX00207, F9XXTX00256, F9XXTX00080, F9XXTX00103, F9XXTX00104, F9XXXX00015, F9XXXX00017, F9XXTX00061, F9XXTX00166, F9XXTX00052, F9XXTX00026, F9XXTX00183, FSXXTX00251, F9XXTX00060, F9XXTX00189, F9XXTX00211, F9XXXX00006, F9XXXX00031, F9XXTX00034, F9XXTX00055, F9XXTX00249, F9XXTX00101, F9XXTX00102, F9XXTX00110, F9XXTX00119, F9XXTX00090, F9XXTX00091, F9XXTX00092, F9XXTX00210, F9XXXX00009, F9XXXX00010, F9XXTX00169, F9XXTX00051, F9XXTX00118, F9XXTX00120, FSXXTX00291, F9XXTX00247, F9XXTX00032, FSXXTX00255, F9XXTX00182, F9XXXX00030, F9XXTX00033, F9XXTX00047, F9XXTX00124, F9XXTX00168, F9XXTX00250, F9XXXX00026, F9XXTX00229, F9XXTX00163

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2278-2023

Related Recalls

Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.

Nov 21, 2024 Diagnostic Equipment Nationwide View Details →

Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.

Nov 21, 2024 Diagnostic Equipment Nationwide View Details →

Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.

Nov 21, 2024 Diagnostic Equipment Nationwide View Details →