MAJOR KNEE - 302497- Procedure tray Catalog Number: ASKN23B
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI-DI: 00191072216885 LOT#"s: 946241
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- American Contract Systems Inc
- Reason for Recall:
- Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
MAJOR KNEE - 302497- Procedure tray Catalog Number: ASKN23B
Product Codes/Lot Numbers:
UDI-DI: 00191072216885 LOT#"s: 946241
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2276-2024
Related Recalls
Medical convenience kits Item Number/Description HSAN63N ANGIO PACK IHAN02W ANGIOGRAPHY PACK MHCA98AK CORONARY ANGIO PACK- 205991 RCAN45C ANGIO CUSTOM RCH UDCO10F CVL CORONARY PACK 219224
American Contract Systems
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Medical convenience kits Item Number/Description BUCY78E CYSTO PACK CMCP10X CYSTO PACK - 205972 IHCY10Y CYSTO PACK MHCY35AE CYSTO PACK- 212487 PLBY52E CYSTO PACK UDCY31Z CYSTO PACK - 206047 UICY28L MOR CYSTO PACK
American Contract Systems
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.