Product Name: AVEA¿ Ventilator. All Models. The affected devices are cleared under K073069 pm 1/25/2008 and K103211 on 05/06/2011. This device is a Class II medical Device. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
Class I - DangerousWhat Should You Do?
- Check if you have this product: All Models. Serial Numbers: BCV01807 AEV01754 AFV01269 AFV01692 AFV01694 AFV02330 AFV02338 AFV02339 AFV02349 AFV02350 AFV02515 AGV01817 AGV01818 AGV02135 AGV02189 AGV02296 AGV02306 AGV02344 AHV03372 AJV01001 AJV01242 AJV01292 AJV01297 AJV01300 AJV01301 AJV01701 AJV01702 AJV01703 AJV01716 AJV01718 AJV01732 AJV01733 AJV01746 AJV01748 AJV01749 AJV01819 AJV01820 AJV01824 AJV02086 AJV02088 AJV02089 AJV02090 AJV02091 AJV02093 AJV02097 AJV02099 AJV02208 AJV02306 AJV02311 AJV02320 AJV02321 AJV02322 AJV02323 AJV02324 AJV02325 AJV02326 AJV02327 AJV02328 AJV02329 AJV02330 AJV02331 AJV02332 AJV02333 AJV02334 AJV02335 AJV02337 AJV02338 AJV02341 AJV02342 AJV02343 AJV02344 AJV02345 AJV02346 AJV02347 AJV02349 AJV02350 AJV02351 AJV02352 AJV02353 AJV02354 AJV02355 AJV02356 AJV02357 AJV02358 AJV02359 AJV02965 AKV01257 AKV01272 AKV01303 AKV01304 AKV01305 AKV01306 AKV01307 AKV01308 AKV01309 AKV01310 AKV01311 AKV01312 AKV01313 AKV01323 AKV01324 AKV01325 AKV01326 AKV01327 AKV01328 AKV01329 AKV01333 AKV01334 AKV01336 AKV01337 AKV01338 AKV01339 AKV01345 AKV01347 AKV01348 AKV01350 AKV01351 AKV01355 AKV01358 AKV01362 AKV01920 AKV01944 BAV04617 BAV04618 BAV04619 BAV04620 BAV04621 BAV04623 BAV04624 BAV04625 BAV04626 BAV04629 BAV04799 BAV04800 BAV04801 BAV04802 BAV04803 BBV01262 BBV01263 BBV01264 BBV01265 BBV01266 BBV01267 BBV01268 BBV01269 BBV01270 BBV01271 BBV03347 BBV03353 BBV03354 BBV03356 BBV03357 BBV03361 BBV03366 BBV03555 BBV03627 BBV03628 BBV03629 BBV03630 BBV03631 BBV03632 BBV03633 BBV03634 BBV03635 BBV03636 BBV03637 BBV03638 BBV03639 BBV03640 BBV03641 BBV03642 BBV03643 BBV03644 BBV03645 BBV03646 BBV03647 BBV03648 BBV03649 BBV03652 BBV03653 BBV03654 BBV03655 BBV03656 BBV03657 BBV03658 BBV03659 BBV03710 BBV03711 BBV03712 BCV02567 BCV02590 BCV02598 BCV02604 BCV02607 BCV02608 ACV01035 ACV01054 ACV01058 ACV01061 ACV01092 ACV01099 ACV01500 ACV01501 ACV01503 ACV01504 ACV01505 ACV01510 ACV01511 ACV01512 ACV01515 ACV01519 ACV01525 ACV01526 ACV01529 ACV01537 ACV01538 ACV01539 ACV01540 ACV01542 ACV01545 ACV01547 ACV01552 ACV01554 ACV01556 ACV01557 ACV01560 ACV01563 ACV01565 ACV01566 ACV01571 ACV01575 ACV01577 ACV01580 ACV01582 ACV01583 ACV01584 ACV01592 ACV01593 ACV01594 ACV01601 ACV01602 ACV01603 ACV01604 ACV01605 ACV01606 ACV01607 ACV01608 ACV01610 ACV01611 ACV01616 ACV01620 ACV01623 ACV01625 ACV01629 ACV01630 ACV01632 ACV01633 ACV01634 ACV01635 ACV01637 ACV01638 ACV01639 ACV01640 ACV01646 ACV01647 ACV01649 ACV01650 ACV01653 ACV01654 ACV01655 ACV01658 ACV01668 ACV01679 ACV01681 ACV01682 ACV01691 ACV01692 ACV01696 ACV01704 ACV01708 ACV01709 ACV01711 ACV01716 ACV01717 ACV01719 ACV01720 ACV01726 ACV01739 ACV01744 ACV01751 ACV01752 ACV01757 ACV01759 ADV01001 ADV01006 ADV01025 ADV01031 ADV01037 ADV01044 ADV01048 ADV01050 ADV01052 ADV01055 ADV01057 ADV01059 ADV01061 ADV01186 ADV01192 ADV01207 ADV01226 ADV01241 ADV01251 ADV01257 ADV01259 ADV01263 ADV01265 ADV01282 ADV01283 ADV01289 ADV01292 ADV01293 ADV01297 ADV01299 ADV01304 ADV01318 ADV01319 ADV01326 ADV01327 ADV01328 ADV01329 ADV01331 ADV01336 ADV01338 ADV01342 ADV01349 ADV01351 ADV01354 ADV01369 ADV01372 ADV01390 ADV01414 ADV01422 ADV01435 ADV01436 ADV01437 ADV01451 ADV01501 ADV01505 ADV01513 ADV01516 ADV01522 ADV01530 ADV01531 ADV01534 ADV01535 ADV01537 ADV01538 ADV01539 ADV01540 ADV01544 ADV01546 ADV01552 ADV01553 ADV01554 ADV01555 ADV01556 ADV01557 ADV01559 ADV01563 ADV01564 ADV01565 ADV01568 ADV01570 ADV01571 ADV01572 ADV01573 ADV01575 ADV01578 ADV01581 ADV01583 ADV01585 ADV01587 ADV01589 ADV01592 ADV01595 ADV01597 ADV01603 ADV01615 ADV01616 ADV01620 ADV01622 ADV01635 ADV01637 ADV01639 ADV01640 ADV01641 ADV01645 ADV01647 ADV01651 ADV01655 ADV01657 ADV01660 ADV01661 ADV01664 ADV01666 ADV01667 ADV01668 ADV01670 ADV01672 ADV01674 ADV01675 ADV01676 ADV01677 ADV01680 ADV01681 ADV01682 ADV01685 ADV01687 ADV01689 ADV01692 ADV01696 ADV01698 ADV01699 ADV01702 ADV01703 ADV01706 ADV01708 ADV01715 ADV01719 ADV01720 ADV01729 ADV01730 ADV01738 ADV01744 ADV01746 ADV01748 ADV01758 ADV01760 ADV01765 ADV01781 ADV01790 ADV01794 ADV01796 ADV01803 ADV01807 ADV01808 ADV01809 ADV01811 ADV01813 ADV01822 ADV01833 ADV01846 ADV01856 ADV01860 ADV01865 ADV01867 ADV01872 ADV01873 ADV01875 ADV01876 ADV01878 ADV01881 ADV01884 ADV01886 ADV01888 ADV02002 ADV02004 ADV02006 ADV02007 ADV02011 ADV02012 ADV02015 ADV02019 ADV02020 ADV02021 ADV02023 ADV02025 ADV02030 ADV02032 ADV02033 ADV02037 ADV02042 ADV02049 ADV02050 ADV02051 ADV02055 ADV02057 ADV02058 ADV02059 ADV02062 ADV02065 ADV02066 ADV02068 ADV02069 ADV02070 ADV02071 ADV02073 ADV02081 ADV02083 ADV02092 ADV02093 ADV02096 ADV02097 ADV02098 ADV02100 ADV02101 ADV02104 ADV02107 ADV02109 ADV02113 ADV02116 ADV02119 ADV02120 ADV02121 ADV02123 ADV02127 ADV02128 ADV02129 ADV02130 ADV02131 ADV02133 ADV02134 ADV02135 ADV02138 ADV02142 ADV02148 ADV02151 ADV02153 ADV02154 ADV02160 ADV02161 ADV02163 ADV02180 ADV02185 ADV02207 ADV02220 ADV03000 ADV03001 ADV03002 ADV03003 ADV03005 ADV03006 ADV03007 ADV03008 ADV03009 ADV03010 ADV03018 ADV03019 ADV03021 ADV03022 ADV03024 ADV03025 ADV03026 ADV03027 ADV03031 ADV03033 ADV03034 ADV03035 ADV03036 ADV03037 ADV03039 ADV03042 ADV03045 ADV03047 ADV03048 ADV03049 ADV03050 ADV03051 ADV03052 ADV03053 ADV03055 ADV03056 ADV03059 ADV03065 ADV03067 ADV03069 ADV03099 ADV03100 ADV03103 ADV03104 ADV03108 ADV03110 ADV03111 ADV03112 ADV03113 ADV03114 ADV03115 ADV04001 ADV04002 ADV04004 ADV04006 ADV04008 ADV04009 ADV04010 ADV04013 ADV04015 ADV04017 ADV04018 ADV04020 ADV04021 ADV04029 ADV04031 ADV04033 ADV04041 ADV04050 ADV04051 ADV04054 ADV04055 ADV04056 ADV04057 ADV04058 ADV04098 ADV04100 ADV04101 ADV04102 ADV04103 ADV04105 ADV04107 ADV04110 ADV04111 ADV04114 ADV04115 ADV04116 ADV04117 ADV04120 ADV04121 ADV04123 ADV04124 ADV04125 ADV04126 ADV04127 ADV04129 ADV04130 ADV04132 ADV04136 ADV04137 ADV04457 ADV04458 ADV04459 ADV04460 ADV04461 ADV05000 ADV05001 ADV05002 ADV05003 ADV05004 ADV05006 ADV05007 ADV05008 ADV05015 ADV05017 ADV05020 ADV05022 ADV05023 ADV05024 ADV05026 ADV05027 ADV05028 ADV05029 ADV05031 ADV05033 ADV05035 ADV05037 ADV05045 ADV05047 ADV05147 ADV05157 ADV05158 ADV05161 ADV05162 ADV05168 ADV05172 ADV05173 ADV05174 ADV05177 ADV05178 ADV05179 ADV05180 ADV05181 ADV05182 ADV05183 ADV05184 ADV05185 ADV05188 ADV05189 ADV05192 ADV05194 ADV05196 ADV05200 ADV06000 ADV06002 ADV06003 ADV06005 ADV06006 ADV06008 ADV06010 ADV06017 ADV06029 ADV06032 ADV06033 ADV06034 ADV06035 ADV06038 ADV06041 ADV06047 ADV06048 ADV06050 ADV06052 ADV06054 ADV06065 ADV06066 ADV06070 ADV06072 ADV06073 ADV06077 ADV06079 ADV06081 ADV06082 ADV06084 ADV06088 ADV06091 ADV06093 ADV06095 ADV06096 ADV06098 ADV06101 ADV06102 ADV06103 ADV06105 ADV06109 ADV06111 ADV06115 ADV06119 ADV06123 ADV06129 ADV06132 ADV06133 ADV06140 ADV06144 ADV06145 ADV06146 ADV06147 ADV06152 ADV06155 ADV06162 ADV06164 ADV06165 ADV06168 ADV06169 ADV06173 ADV06174 ADV06177 ADV06179 ADV06180 ADV06181 ADV06183 ADV06184 ADV06186 ADV06187 ADV06189 ADV06190 ADV06195 ADV06196 ADV06199 ADV06201 ADV06204 ADV06205 ADV06207 ADV06209 ADV06210 ADV06212 ADV06214 ADV06216 ADV06219 ADV06221 ADV06223 ADV06224 ADV06226 ADV06227 ADV06230 ADV06231 ADV06232 ADV06234 ADV06235 ADV06237 ADV06238 ADV06239 ADV06240 ADV06241 ADV06242 ADV06243 ADV06244 ADV06246 ADV06247 ADV06248 ADV06249 ADV06252 ADV06258 ADV06261 ADV06263 ADV06267 ADV06268 ADV06270 ADV06275 ADV06277 ADV06278 ADV06281 ADV06282 ADV06283 ADV06284 ADV06287 ADV06290 ADV06291 ADV06292 ADV06293 ADV06294 ADV06298 ADV06299 ADV06300 ADV06301 ADV06304 ADV06305 ADV06306 ADV06308 ADV06309 ADV06310 ADV06313 ADV06314 ADV06315 ADV06321 ADV06323 ADV06326 ADV06327 ADV06328 ADV06329 ADV06330 ADV06332 ADV06333 ADV06336 ADV06337 ADV06338 ADV06342 ADV06343 ADV06347 ADV06349 ADV06350 ADV06353 ADV06354 ADV06355 ADV06356 ADV06358 ADV06359 ADV06361 ADV06362 ADV06363 ADV06365 ADV06367 ADV06370 ADV06371 ADV06379 ADV06381 ADV06393 ADV06394 ADV06395 ADV06396 ADV06398 ADV06399 ADV06405 ADV06406 ADV06407 ADV06408 ADV06409 ADV06420 AEV01002 AEV01009 AEV01011 AEV01020 AEV01032 AEV01036 AEV01037 AEV01038 AEV01039 AEV01040 AEV01046 AEV01049 AEV01052 AEV01057 AEV01500 AEV01503 AEV01505 AEV01508 AEV01509 AEV01512 AEV01515 AEV01518 AEV01520 AEV01521 AEV01527 AEV01528 AEV01530 AEV01532 AEV01533 AEV01534 AEV01535 AEV01536 AEV01537 AEV01538 AEV01539 AEV01540 AEV01541 AEV01542 AEV01543 AEV01544 AEV01545 AEV01549 AEV01551 AEV01552 AEV01553 AEV01558 AEV01561 AEV01562 AEV01563 AEV01564 AEV01566 AEV01568 AEV01569 AEV01570 AEV01572 AEV01574 AEV01575 AEV01576 AEV01577 AEV01578 AEV01579 AEV01580 AEV01583 AEV01584 AEV01586 AEV01587 AEV01589 AEV01590 AEV01591 AEV01592 AEV01593 AEV01594 AEV01595 AEV01597 AEV01600 AEV01602 AEV01603 AEV01604 AEV01605 AEV01606 AEV01607 AEV01608 AEV01609 AEV01610 AEV01611 AEV01612 AEV01613 AEV01614 AEV01615 AEV01616 AEV01617 AEV01618 AEV01619 AEV01620 AEV01621 AEV01622 AEV01624 AEV01625 AEV01626 AEV01627 AEV01628 AEV01629 AEV01631 AEV01632 AEV01633 AEV01638 AEV01639 AEV01640 AEV01641 AEV01642 AEV01645 AEV01646 AEV01647 AEV01648 AEV01651 AEV01652 AEV01653 AEV01654 AEV01656 AEV01657 AEV01658 AEV01660 AEV01661 AEV01662 AEV01663 AEV01664 AEV01665 AEV01670 AEV01674 AEV01675 AEV01677 AEV01679 AEV01682 AEV01683 AEV01684 AEV01685 AEV01687 AEV01688 AEV01689 AEV01690 AEV01691 AEV01692 AEV01694 AEV01695 AEV01696 AEV01697 AEV01701 AEV01703 AEV01704 AEV01708 AEV01709 AEV01710 AEV01711 AEV01712 AEV01713 AEV01714 AEV01715 AEV01716 AEV01718 AEV01719 AEV01720 AEV01721 AEV01723 AEV01724 AEV01725 AEV01726 AEV01727 AEV01728 AEV01729 AEV01730 AEV01731 AEV01733 AEV01741 AEV01742 AEV01744 AEV01745 AEV01749 AEV01753 AEV01756 AEV01757 AEV01761 AEV01762 AEV01763 AEV01764 AEV01765 AEV01766 AEV01768 AEV01770 AEV01771 AEV01772 AEV01773 AEV01774 AEV01775 AEV01776 AEV01777 AEV01779 AEV01780 AEV01781 AEV01782 AEV01783 AEV01785 AEV01788 AEV01791 AEV01792 AEV01793 AEV01796 AEV01797 AEV01798 AEV01799 AEV01801 AEV01802 AEV01806 AEV01807 AEV01809 AEV01810 AEV01812 AEV01813 AEV01814 AEV01815 AEV01816 AEV01817 AEV01818 AEV01821 AEV01822 AEV01825 AEV01829 AEV01831 AEV01834 AEV01835 AEV01837 AEV01840 AEV01841 AEV01843 AEV01844 AEV01845 AEV01846 AEV01847 AEV01851 AEV01854 AEV01855 AEV01859 AEV01865 AEV01866 AEV01870 AEV01876 AEV01883 AEV01884 AEV01886 AEV01887 AEV01888 AEV01894 AEV01896 AEV01897 AEV01903 AEV01906 AEV01908 AEV01909 AEV01911 AEV01912 AEV01914 AEV01919 AEV01921 AEV01922 AEV01923 AEV01926 AEV01928 AEV01929 AEV01930 AEV01931 AEV01932 AEV01933 AEV01934 AEV01936 AEV01937 AEV01939 AEV01942 AEV01943 AEV01944 AEV01945 AEV01946 AEV01947 AEV01948 AEV01951 AEV01952 AEV01953 AEV01955 AEV01958 AEV01961 AEV01962 AEV01964 AEV01968 AEV01969 AEV01972 AEV01974 AEV01978 AEV01979 AEV01981 AEV01984 AEV01986 AEV01987 AEV01993 AEV01995 AEV01996 AEV01997 AEV02002 AEV02003 AEV02011 AEV02018 AEV02019 AEV02021 AEV02023 AEV02026 AEV02029 AEV02030 AEV02032 AEV02033 AEV02034 AEV02037 AEV02041 AEV02043 AEV02046 AEV02050 AEV02051 AEV02052 AEV02056 AEV02060 AEV02062 AEV02063 AEV02064 AEV02066 AEV02070 AEV02071 AEV02072 AEV02076 AEV02078 AEV02079 AEV02080 AEV02082 AEV02087 AEV02088 AEV02089 AEV02091 AEV02092 AEV02094 AEV02095 AEV02099 AEV02102 AEV02106 AEV02108 AEV02110 AEV02114 AEV02117 AEV02118 AEV02119 AEV02120 AEV02122 AEV02123 AEV02124 AEV02125 AEV02130 AEV02136 AEV02138 AEV02141 AEV02142 AEV02144 AEV02145 AEV02148 AEV02151 AEV02154 AEV02158 AEV02160 AEV02161 AEV02162 AEV02164 AEV02165 AEV02166 AEV02168 AEV02169 AEV02170 AEV02171 AEV02172 AEV02174 AEV02175 AEV02177 AEV02178 AEV02179 AEV02180 AEV02181 AEV02182 AEV02183 AEV02184 AEV02185 AEV02186 AEV02187 AEV02188 AEV02192 AEV02197 AEV02202 AEV02204 AEV02205 AEV02206 AEV02207 AEV02208 AEV02209 AEV02211 AEV02213 AEV02214 AEV02221 AEV02222 AEV02225 AEV02226 AEV02228 AEV02229 AEV02230 AEV02233 AEV02234 AEV02236 AEV02237 AEV02239 AEV02243 AEV02244 AEV02248 AEV02249 AEV02252 AEV02253 AEV02254 AEV02255 AEV02258 AEV02259 AEV02260 AEV02261 AEV02262 AEV02264 AEV02265 AEV02266 AEV02268 AEV02269 AEV02272 AEV02274 AEV02275 AEV02276 AEV02277 AEV02279 AEV02283 AEV02285 AEV02287 AEV02288 AEV02292 AEV02293 AEV02296 AEV02298 AEV02299 AEV02300 AEV02301 AEV02303 AEV02306 AEV02309 AEV02310 AEV02311 AEV02313 AEV02318 AEV02320 AEV02321 AEV02323 AEV02325 AEV02326 AEV02327 AEV02328 AEV02329 AEV02331 AEV02332 AEV02333 AEV02334 AEV02337 AEV02339 AEV02340 AEV02341 AEV02342 AEV02347 AEV02350 AEV02351 AEV02355 AEV02356 AEV02357 AEV02358 AEV02359 AEV02360 AEV02361 AEV02362 AEV02363 AEV02364 AEV02365 AEV02366 AEV02367 AEV02369 AEV02370 AEV02371 AEV02372 AEV02373 AEV02375 AEV02376 AEV02378 AEV02380 AEV02381 AEV02383 AEV02384 AEV02385 AEV02386 AEV02387 AEV02390 AEV02391 AEV02394 AEV02395 AEV02406 AEV02407 AEV02409 AEV02411 AEV02412 AEV02414 AEV02415 AEV02417 AEV02419 AEV02420 AEV02421 AEV02422 AEV02423 AEV02424 AEV02425 AEV02426 AEV02427 AEV02428 AEV02430 AEV02432 AEV02435 AEV02440 AEV02441 AEV02442 AEV02443 AEV02445 AEV02446 AEV02447 AEV02448 AEV02449 AEV02450 AEV02455 AEV02456 AEV02461 AEV02466 AEV02471 AEV02472 AEV02473 AEV02474 AEV02475 AEV02476 AEV02477 AEV02478 AEV02480 AEV02482 AEV02484 AEV02488 AEV02491 AEV02495 AEV02497 AEV02499 AEV02500 AEV02504 AEV02505 AEV02506 AEV02507 AEV02513 AEV02515 AEV02518 AEV02521 AEV02523 AEV02524 AEV02525 AEV02526 AEV02537 AEV02544 AEV02545 AEV02547 AEV02548 AEV02549 AEV02551 AEV02552 AEV02554 AEV02556 AEV02557 AEV02561 AEV02562 AEV02565 AEV02566 AEV02571 AEV02574 AEV02575 AEV02576 AEV02577 AEV02580 AEV02581 AEV02585 AEV02586 AEV02587 AEV02588 AEV02589 AEV02590 AEV02591 AEV02594 AEV02601 AEV02602 AEV02608 AEV02612 AEV02613 AEV02614 AEV02617 AEV02620 AEV02627 AEV02629 AEV02631 AEV02635 AEV02636 AEV02643 AEV02644 AEV02648 AEV02653 AEV02654 AEV02656 AEV02657 AEV02662 AEV02665 AEV02666 AEV02667 AEV02670 AEV02674 AEV02675 AEV02679 AEV02681 AEV02682 AEV02684 AEV02685 AEV02691 AEV02693 AEV02694 AEV02695 AEV02696 AEV02697 AEV02699 AEV02700 AEV02704 AEV02706 AEV02707 AEV02708 AEV02709 AEV02710 AEV02712 AEV02713 AEV02717 AEV02718 AEV02719 AEV02720 AEV02721 AEV02723 AEV02725 AEV02726 AEV02727 AEV02728 AEV02729 AEV02730 AEV02736 AEV02738 AEV02739 AEV02743 AEV02744 AEV02745 AEV02746 AEV02747 AEV02750 AEV02752 AEV02753 AEV02755 AEV02757 AEV02760 AEV02761 AEV02763 AEV02766 AEV02768 AFV01001 AFV01002 AFV01003 AFV01005 AFV01006 AFV01008 AFV01010 AFV01013 AFV01017 AFV01022 AFV01028 AFV01029 AFV01030 AFV01034 AFV01037 AFV01040 AFV01042 AFV01043 AFV01047 AFV01048 AFV01053 AFV01054 AFV01055 AFV01057 AFV01062 AFV01064 AFV01065 AFV01068 AFV01072 AFV01075 AFV01076 AFV01078 AFV01079 AFV01081 AFV01090 AFV01092 AFV01097 AFV01105 AFV01114 AFV01117 AFV01142 AFV01150 AFV01151 AFV01157 AFV01158 AFV01159 AFV01163 AFV01166 AFV01169 AFV01175 AFV01176 AFV01177 AFV01179 AFV01181 AFV01184 AFV01185 AFV01189 AFV01190 AFV01191 AFV01192 AFV01193 AFV01194 AFV01195 AFV01200 AFV01202 AFV01203 AFV01209 AFV01211 AFV01212 AFV01214 AFV01216 AFV01217 AFV01220 AFV01221 AFV01222 AFV01225 AFV01230 AFV01233 AFV01237 AFV01238 AFV01241 AFV01242 AFV01245 AFV01251 AFV01255 AFV01256 AFV01257 AFV01258 AFV01259 AFV01260 AFV01262 AFV01264 AFV01265 AFV01267 AFV01268 AFV01271 AFV01272 AFV01275 AFV01277 AFV01280 AFV01284 AFV01285 AFV01286 AFV01287 AFV01289 AFV01291 AFV01292 AFV01293 AFV01294 AFV01295 AFV01296 AFV01298 AFV01302 AFV01309 AFV01311 AFV01313 AFV01315 AFV01320 AFV01321 AFV01322 AFV01323 AFV01324 AFV01325 AFV01327 AFV01331 AFV01333 AFV01334 AFV01335 AFV01336 AFV01337 AFV01338 AFV01339 AFV01341 AFV01343 AFV01346 AFV01352 AFV01355 AFV01359 AFV01360 AFV01361 AFV01362 AFV01363 AFV01365 AFV01367 AFV01368 AFV01369 AFV01370 AFV01371 AFV01372 AFV01373 AFV01374 AFV01375 AFV01379 AFV01380 AFV01381 AFV01382 AFV01383 AFV01385 AFV01386 AFV01387 AFV01388 AFV01389 AFV01390 AFV01392 AFV01393 AFV01394 AFV01395 AFV01396 AFV01398 AFV01400 AFV01401 AFV01402 AFV01403 AFV01404 AFV01406 AFV01407 AFV01408 AFV01409 AFV01413 AFV01416 AFV01417 AFV01418 AFV01419 AFV01421 AFV01422 AFV01423 AFV01424 AFV01425 AFV01427 AFV01428 AFV01429 AFV01430 AFV01431 AFV01433 AFV01434 AFV01435 AFV01436 AFV01437 AFV01438 AFV01440 AFV01441 AFV01442 AFV01443 AFV01444 AFV01447 AFV01449 AFV01451 AFV01452 AFV01454 AFV01456 AFV01458 AFV01460 AFV01462 AFV01463 AFV01464 AFV01466 AFV01467 AFV01468 AFV01471 AFV01472 AFV01473 AFV01475 AFV01476 AFV01477 AFV01478 AFV01479 AFV01480 AFV01481 AFV01482 AFV01484 AFV01485 AFV01487 AFV01488 AFV01489 AFV01490 AFV01491 AFV01494 AFV01495 AFV01496 AFV01499 AFV01500 AFV01503 AFV01504 AFV01505 AFV01510 AFV01511 AFV01514 AFV01521 AFV01523 AFV01524 AFV01528 AFV01530 AFV01532 AFV01533 AFV01534 AFV01536 AFV01537 AFV01540 AFV01541 AFV01543 AFV01544 AFV01545 AFV01546 AFV01549 AFV01550 AFV01551 AFV01552 AFV01553 AFV01554 AFV01556 AFV01557 AFV01558 AFV01559 AFV01560 AFV01563 AFV01565 AFV01566 AFV01567 AFV01568 AFV01569 AFV01570 AFV01571 AFV01572 AFV01573 AFV01574 AFV01575 AFV01576 AFV01577 AFV01578 AFV01579 AFV01580 AFV01582 AFV01584 AFV01585 AFV01586 AFV01587 AFV01588 AFV01589 AFV01591 AFV01599 AFV01600 AFV01601 AFV01602 AFV01603 AFV01610 AFV01612 AFV01616 AFV01618 AFV01619 AFV01620 AFV01622 AFV01623 AFV01626 AFV01627 AFV01628 AFV01629 AFV01630 AFV01631 AFV01632 AFV01633 AFV01634 AFV01635 AFV01637 AFV01640 AFV01641 AFV01643 AFV01644 AFV01645 AFV01647 AFV01649 AFV01652 AFV01653 AFV01654 AFV01655 AFV01656 AFV01657 AFV01658 AFV01661 AFV01662 AFV01665 AFV01666 AFV01668 AFV01677 AFV01685 AFV01686 AFV01688 AFV01690 AFV01695 AFV01697 AFV01704 AFV01705 AFV01706 AFV01717 AFV01719 AFV01720 AFV01721 AFV01722 AFV01729 AFV01732 AFV01736 AFV01742 AFV01744 AFV01745 AFV01746 AFV01747 AFV01748 AFV01756 AFV01757 AFV01758 AFV01761 AFV01762 AFV01767 AFV01769 AFV01776 AFV01780 AFV01781 AFV01787 AFV01789 AFV01790 AFV01791 AFV01792 AFV01793 AFV01794 AFV01795 AFV01798 AFV01799 AFV01800 AFV01801 AFV01802 AFV01803 AFV01804 AFV01806 AFV01808 AFV01810 AFV01811 AFV01815 AFV01818 AFV01819 AFV01822 AFV01824 AFV01828 AFV01830 AFV01831 AFV01834 AFV01835 AFV01836 AFV01837 AFV01838 AFV01840 AFV01842 AFV01844 AFV01845 AFV01846 AFV01848 AFV01849 AFV01851 AFV01852 AFV01854 AFV01856 AFV01857 AFV01858 AFV01860 AFV01862 AFV01863 AFV01864 AFV01865 AFV01866 AFV01867 AFV01868 AFV01869 AFV01870 AFV01871 AFV01873 AFV01874 AFV01875 AFV01876 AFV01877 AFV01878 AFV01879 AFV01880 AFV01885 AFV01887 AFV01888 AFV01889 AFV01891 AFV01892 AFV01893 AFV01895 AFV01898 AFV01899 AFV01902 AFV01905 AFV01907 AFV01908 AFV01909 AFV01911 AFV01917 AFV01919 AFV01922 AFV01928 AFV01932 AFV01933 AFV01934 AFV01941 AFV01944 AFV01946 AFV01947 AFV01954 AFV01958 AFV01960 AFV01962 AFV01963 AFV01964 AFV01981 AFV01985 AFV01986 AFV01988 AFV01990 AFV01991 AFV01992 AFV01994 AFV01995 AFV02000 AFV02003 AFV02004 AFV02008 AFV02011 AFV02012 AFV02013 AFV02014 AFV02015 AFV02016 AFV02017 AFV02019 AFV02020 AFV02021 AFV02022 AFV02023 AFV02025 AFV02026 AFV02028 AFV02030 AFV02031 AFV02034 AFV02035 AFV02036 AFV02037 AFV02038 AFV02039 AFV02040 AFV02041 AFV02043 AFV02044 AFV02045 AFV02047 AFV02048 AFV02052 AFV02053 AFV02054 AFV02055 AFV02056 AFV02058 AFV02059 AFV02060 AFV02061 AFV02062 AFV02063 AFV02065 AFV02067 AFV02070 AFV02072 AFV02074 AFV02075 AFV02076 AFV02078 AFV02079 AFV02080 AFV02082 AFV02083 AFV02087 AFV02088 AFV02090 AFV02092 AFV02093 AFV02095 AFV02096 AFV02097 AFV02098 AFV02099 AFV02101 AFV02102 AFV02104 AFV02105 AFV02106 AFV02109 AFV02113 AFV02114 AFV02115 AFV02117 AFV02119 AFV02120 AFV02122 AFV02123 AFV02126 AFV02127 AFV02129 AFV02130 AFV02132 AFV02133 AFV02134 AFV02136 AFV02137 AFV02141 AFV02142 AFV02143 AFV02144 AFV02145 AFV02146 AFV02147 AFV02150 AFV02152 AFV02153 AFV02155 AFV02156 AFV02157 AFV02159 AFV02162 AFV02164 AFV02165 AFV02168 AFV02170 AFV02171 AFV02172 AFV02173 AFV02174 AFV02176 AFV02177 AFV02178 AFV02179 AFV02182 AFV02183 AFV02187 AFV02188 AFV02190 AFV02191 AFV02192 AFV02193 AFV02196 AFV02197 AFV02201 AFV02205 AFV02206 AFV02207 AFV02208 AFV02209 AFV02210 AFV02211 AFV02214 AFV02215 AFV02217 AFV02225 AFV02226 AFV02227 AFV02228 AFV02229 AFV02231 AFV02232 AFV02233 AFV02234 AFV02235 AFV02236 AFV02237 AFV02238 AFV02239 AFV02240 AFV02242 AFV02244 AFV02245 AFV02246 AFV02247 AFV02248 AFV02250 AFV02251 AFV02253 AFV02254 AFV02257 AFV02260 AFV02261 AFV02262 AFV02263 AFV02264 AFV02266 AFV02268 AFV02269 AFV02270 AFV02273 AFV02274 AFV02275 AFV02277 AFV02279 AFV02280 AFV02281 AFV02282 AFV02283 AFV02284 AFV02287 AFV02288 AFV02289 AFV02291 AFV02292 AFV02293 AFV02294 AFV02295 AFV02296 AFV02297 AFV02298 AFV02299 AFV02301 AFV02302 AFV02303 AFV02306 AFV02307 AFV02308 AFV02310 AFV02311 AFV02317 AFV02318 AFV02319 AFV02322 AFV02323 AFV02324 AFV02325 AFV02326 AFV02328 AFV02329 AFV02332 AFV02334 AFV02341 AFV02344 AFV02346 AFV02347 AFV02352 AFV02353 AFV02354 AFV02355 AFV02356 AFV02357 AFV02359 AFV02363 AFV02365 AFV02366 AFV02368 AFV02369 AFV02371 AFV02372 AFV02376 AFV02378 AFV02379 AFV02380 AFV02381 AFV02382 AFV02385 AFV02387 AFV02388 AFV02394 AFV02395 AFV02397 AFV02398 AFV02399 AFV02401 AFV02402 AFV02404 AFV02405 AFV02407 AFV02412 AFV02413 AFV02415 AFV02416 AFV02417 AFV02418 AFV02420 AFV02421 AFV02422 AFV02423 AFV02424 AFV02427 AFV02429 AFV02433 AFV02434 AFV02435 AFV02438 AFV02439 AFV02441 AFV02443 AFV02444 AFV02446 AFV02447 AFV02448 AFV02449 AFV02450 AFV02451 AFV02453 AFV02455 AFV02458 AFV02460 AFV02461 AFV02462 AFV02463 AFV02464 AFV02466 AFV02473 AFV02479 AFV02480 AFV02481 AFV02483 AFV02484 AFV02485 AFV02486 AFV02487 AFV02488 AFV02490 AFV02491 AFV02492 AFV02493 AFV02494 AFV02495 AFV02497 AFV02499 AFV02500 AFV02501 AFV02502 AFV02504 AFV02505 AFV02506 AFV02507 AFV02508 AFV02511 AFV02512 AFV02513 AFV02514 AFV02516 AFV02517 AFV02518 AFV02519 AFV02520 AFV02521 AFV02522 AFV02523 AFV02524 AFV02525 AFV02526 AFV02527 AFV02528 AFV02529 AFV02530 AFV02531 AFV02532 AFV02533 AFV02534 AFV02535 AFV02536 AFV02537 AFV02538 AFV02539 AFV02540 AFV02542 AFV02543 AFV02545 AFV02547 AFV02548 AFV02549 AFV02550 AFV02551 AFV02552 AFV02553 AFV02554 AFV02555 AFV02556 AFV02557 AFV02558 AFV02559 AFV02560 AFV02561 AFV02562 AFV02563 AFV02564 AFV02565 AFV02566 AFV02567 AFV02568 AFV02569 AFV02570 AFV02571 AFV02572 AFV02573 AFV02574 AFV02575 AFV02576 AFV02577 AFV02578 AFV02579 AFV02580 AFV02581 AFV02582 AFV02583 AFV02584 AFV02585 AFV02586 AFV02587 AFV02588 AFV02589 AFV02590 AFV02591 AFV02592 AFV02593 AFV02594 AFV02595 AFV02596 AFV02597 AFV02598 AFV02599 AFV02600 AFV02601 AFV02602 AFV02603 AFV02604 AFV02605 AFV02606 AFV02608 AFV02609 AFV02610 AFV02611 AFV02612 AFV02613 AFV02614 AFV02615 AFV02616 AFV02617 AFV02619 AFV02621 AFV02623 AFV02624 AFV02626 AFV02770 AGV01000 AGV01001 AGV01002 AGV01003 AGV01004 AGV01005 AGV01006 AGV01007 AGV01009 AGV01010 AGV01014 AGV01020 AGV01022 AGV01023 AGV01024 AGV01025 AGV01027 AGV01031 AGV01034 AGV01037 AGV01038 AGV01040 AGV01041 AGV01042 AGV01043 AGV01044 AGV01045 AGV01046 AGV01047 AGV01048 AGV01049 AGV01050 AGV01051 AGV01052 AGV01053 AGV01055 AGV01056 AGV01061 AGV01064 AGV01068 AGV01071 AGV01074 AGV01082 AGV01087 AGV01095 AGV01098 AGV01100 AGV01104 AGV01106 AGV01109 AGV01113 AGV01116 AGV01119 AGV01120 AGV01121 AGV01126 AGV01127 AGV01128 AGV01129 AGV01131 AGV01132 AGV01136 AGV01138 AGV01140 AGV01146 AGV01149 AGV01151 AGV01152 AGV01155 AGV01156 AGV01157 AGV01160 AGV01161 AGV01162 AGV01163 AGV01165 AGV01166 AGV01169 AGV01170 AGV01173 AGV01175 AGV01177 AGV01178 AGV01179 AGV01180 AGV01183 AGV01186 AGV01188 AGV01193 AGV01195 AGV01196 AGV01197 AGV01198 AGV01199 AGV01200 AGV01201 AGV01202 AGV01203 AGV01204 AGV01205 AGV01206 AGV01207 AGV01209 AGV01210 AGV01212 AGV01214 AGV01215 AGV01216 AGV01217 AGV01219 AGV01220 AGV01221 AGV01222 AGV01225 AGV01226 AGV01227 AGV01228 AGV01229 AGV01230 AGV01233 AGV01235 AGV01237 AGV01238 AGV01239 AGV01241 AGV01242 AGV01243 AGV01244 AGV01248 AGV01249 AGV01250 AGV01251 AGV01252 AGV01254 AGV01255 AGV01256 AGV01257 AGV01258 AGV01259 AGV01261 AGV01262 AGV01263 AGV01268 AGV01269 AGV01270 AGV01271 AGV01275 AGV01277 AGV01278 AGV01279 AGV01280 AGV01281 AGV01284 AGV01286 AGV01287 AGV01288 AGV01293 AGV01295 AGV01299 AGV01300 AGV01301 AGV01304 AGV01305 AGV01312 AGV01314 AGV01317 AGV01318 AGV01321 AGV01322 AGV01323 AGV01327 AGV01328 AGV01329 AGV01333 AGV01334 AGV01335 AGV01338 AGV01349 AGV01352 AGV01353 AGV01358 AGV01362 AGV01379 AGV01382 AGV01385 AGV01390 AGV01393 AGV01402 AGV01405 AGV01410 AGV01412 AGV01413 AGV01414 AGV01415 AGV01426 AGV01429 AGV01441 AGV01443 AGV01445 AGV01446 AGV01447 AGV01449 AGV01452 AGV01453 AGV01454 AGV01455 AGV01457 AGV01458 AGV01459 AGV01460 AGV01461 AGV01462 AGV01463 AGV01464 AGV01465 AGV01468 AGV01469 AGV01470 AGV01477 AGV01478 AGV01480 AGV01486 AGV01487 AGV01488 AGV01491 AGV01492 AGV01493 AGV01494 AGV01495 AGV01497 AGV01498 AGV01502 AGV01503 AGV01505 AGV01506 AGV01509 AGV01511 AGV01512 AGV01513 AGV01514 AGV01517 AGV01519 AGV01520 AGV01521 AGV01522 AGV01523 AGV01524 AGV01525 AGV01526 AGV01527 AGV01528 AGV01530 AGV01531 AGV01532 AGV01533 AGV01534 AGV01535 AGV01536 AGV01537 AGV01538 AGV01539 AGV01541 AGV01543 AGV01544 AGV01546 AGV01547 AGV01550 AGV01552 AGV01553 AGV01554 AGV01557 AGV01561 AGV01563 AGV01569 AGV01570 AGV01571 AGV01576 AGV01577 AGV01580 AGV01581 AGV01584 AGV01586 AGV01587 AGV01589 AGV01590 AGV01592 AGV01593 AGV01594 AGV01595 AGV01596 AGV01597 AGV01605 AGV01616 AGV01621 AGV01629 AGV01632 AGV01633 AGV01636 AGV01642 AGV01653 AGV01655 AGV01656 AGV01657 AGV01659 AGV01667 AGV01668 AGV01672 AGV01675 AGV01676 AGV01677 AGV01681 AGV01682 AGV01684 AGV01685 AGV01686 AGV01689 AGV01704 AGV01707 AGV01708 AGV01709 AGV01710 AGV01711 AGV01712 AGV01714 AGV01719 AGV01720 AGV01721 AGV01722 AGV01725 AGV01726 AGV01727 AGV01728 AGV01729 AGV01730 AGV01731 AGV01732 AGV01733 AGV01734 AGV01736 AGV01737 AGV01738 AGV01739 AGV01740 AGV01741 AGV01742 AGV01743 AGV01744 AGV01745 AGV01746 AGV01748 AGV01749 AGV01750 AGV01751 AGV01752 AGV01753 AGV01754 AGV01755 AGV01756 AGV01757 AGV01758 AGV01759 AGV01760 AGV01761 AGV01762 AGV01763 AGV01764 AGV01765 AGV01766 AGV01767 AGV01768 AGV01769 AGV01770 AGV01771 AGV01772 AGV01773 AGV01775 AGV01776 AGV01777 AGV01780 AGV01781 AGV01782 AGV01784 AGV01786 AGV01787 AGV01789 AGV01791 AGV01792 AGV01793 AGV01794 AGV01795 AGV01796 AGV01797 AGV01798 AGV01799 AGV01800 AGV01801 AGV01802 AGV01803 AGV01804 AGV01806 AGV01809 AGV01810 AGV01811 AGV01812 AGV01834 AGV01835 AGV01837 AGV01838 AGV01842 AGV01846 AGV01847 AGV01848 AGV01849 AGV01850 AGV01851 AGV01852 AGV01853 AGV01854 AGV01855 AGV01856 AGV01857 AGV01858 AGV01859 AGV01860 AGV01861 AGV01862 AGV01863 AGV01864 AGV01865 AGV01866 AGV01867 AGV01868 AGV01869 AGV01870 AGV01871 AGV01872 AGV01873 AGV01875 AGV01876 AGV01877 AGV01878 AGV01880 AGV01881 AGV01882 AGV01883 AGV01884 AGV01885 AGV01886 AGV01887 AGV01890 AGV01891 AGV01892 AGV01893 AGV01894 AGV01898 AGV01899 AGV01900 AGV01901 AGV01902 AGV01903 AGV01904 AGV01909 AGV01911 AGV01912 AGV01918 AGV01920 AGV01922 AGV01928 AGV01929 AGV01930 AGV01931 AGV01932 AGV01933 AGV01938 AGV01940 AGV01942 AGV01953 AGV01960 AGV01964 AGV01968 AGV01971 AGV01972 AGV01978 AGV01979 AGV01980 AGV01984 AGV01985 AGV01986 AGV01987 AGV01988 AGV01991 AGV01992 AGV01993 AGV01994 AGV01995 AGV01996 AGV01997 AGV01998 AGV02000 AGV02001 AGV02002 AGV02003 AGV02004 AGV02008 AGV02009 AGV02010 AGV02011 AGV02012 AGV02013 AGV02014 AGV02015 AGV02016 AGV02017 AGV02018 AGV02019 AGV02020 AGV02021 AGV02022 AGV02024 AGV02025 AGV02026 AGV02027 AGV02028 AGV02029 AGV02031 AGV02034 AGV02035 AGV02036 AGV02038 AGV02040 AGV02041 AGV02043 AGV02044 AGV02046 AGV02048 AGV02049 AGV02057 AGV02060 AGV02062 AGV02065 AGV02066 AGV02067 AGV02068 AGV02070 AGV02071 AGV02072 AGV02073 AGV02075 AGV02077 AGV02078 AGV02081 AGV02082 AGV02086 AGV02087 AGV02088 AGV02089 AGV02090 AGV02093 AGV02094 AGV02096 AGV02097 AGV02098 AGV02101 AGV02104 AGV02106 AGV02108 AGV02111 AGV02113 AGV02114 AGV02115 AGV02116 AGV02117 AGV02119 AGV02121 AGV02122 AGV02124 AGV02126 AGV02127 AGV02128 AGV02130 AGV02131 AGV02134 AGV02136 AGV02139 AGV02142 AGV02144 AGV02147 AGV02149 AGV02150 AGV02152 AGV02153 AGV02155 AGV02156 AGV02158 AGV02161 AGV02162 AGV02163 AGV02170 AGV02171 AGV02177 AGV02179 AGV02183 AGV02185 AGV02186 AGV02187 AGV02188 AGV02190 AGV02192 AGV02193 AGV02194 AGV02195 AGV02197 AGV02198 AGV02200 AGV02201 AGV02203 AGV02204 AGV02205 AGV02206 AGV02207 AGV02209 AGV02210 AGV02211 AGV02213 AGV02216 AGV02217 AGV02218 AGV02222 AGV02223 AGV02224 AGV02226 AGV02231 AGV02232 AGV02234 AGV02237 AGV02239 AGV02241 AGV02243 AGV02244 AGV02247 AGV02249 AGV02252 AGV02260 AGV02266 AGV02272 AGV02278 AGV02282 AGV02284 AGV02286 AGV02289 AGV02291 AGV02293 AGV02297 AGV02300 AGV02302 AGV02304 AGV02310 AGV02316 AGV02321 AGV02324 AGV02325 AGV02330 AGV02331 AGV02332 AGV02333 AGV02336 AGV02338 AGV02345 AGV02349 AGV02351 AGV02354 AGV02355 AGV02357 AGV02358 AGV02359 AGV02362 AGV02364 AGV02366 AGV02368 AGV02369 AGV02372 AGV02373 AGV02375 AGV02377 AGV02378 AGV02383 AGV02384 AGV02387 AGV02388 AGV02391 AGV02394 AGV02397 AGV02400 AGV02402 AGV02404 AGV02405 AGV02406 AGV02407 AGV02409 AGV02410 AGV02412 AGV02413 AGV02416 AGV02417 AGV02419 AGV02420 AGV02423 AGV02425 AGV02426 AGV02427 AGV02429 AGV02432 AGV02435 AGV02436 AGV02440 AGV02441 AGV02443 AGV02444 AGV02446 AGV02447 AGV02450 AGV02451 AGV02452 AGV02454 AGV02455 AGV02459 AGV02466 AGV02471 AGV02476 AGV02480 AGV02481 AGV02482 AGV02485 AGV02486 AGV02487 AGV02488 AGV02489 AGV02490 AGV02491 AGV02492 AGV02493 AGV02494 AGV02495 AGV02496 AGV02497 AGV02499 AGV02500 AGV02503 AGV02508 AGV02509 AGV02510 AGV02511 AGV02513 AGV02514 AGV02515 AGV02516 AGV02518 AGV02519 AGV02520 AGV02521 AGV02522 AGV02523 AGV02525 AGV02527 AGV02528 AGV02530 AGV02531 AGV02532 AGV02533 AGV02534 AGV02535 AGV02536 AGV02538 AGV02540 AGV02541 AGV02542 AGV02545 AGV02546 AGV02549 AGV02550 AGV02551 AGV02552 AGV02553 AGV02555 AGV02556 AGV02557 AGV02558 AGV02559 AGV02560 AGV02561 AGV02562 AGV02563 AGV02564 AGV02565 AGV02566
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Carefusion 211 Inc dba Carefusion
- Reason for Recall:
- CareFusion has identified a potential risk associated with AVEA¿ ventilators when used at higher altitudes. A proactive complaint review identified an error in the barometric pressure compensation. CareFusion is voluntarily initiating a field correction of the affected devices to correct the error.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Product Name: AVEA¿ Ventilator. All Models. The affected devices are cleared under K073069 pm 1/25/2008 and K103211 on 05/06/2011. This device is a Class II medical Device. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
Product Codes/Lot Numbers:
All Models. Serial Numbers: BCV01807 AEV01754 AFV01269 AFV01692 AFV01694 AFV02330 AFV02338 AFV02339 AFV02349 AFV02350 AFV02515 AGV01817 AGV01818 AGV02135 AGV02189 AGV02296 AGV02306 AGV02344 AHV03372 AJV01001 AJV01242 AJV01292 AJV01297 AJV01300 AJV01301 AJV01701 AJV01702 AJV01703 AJV01716 AJV01718 AJV01732 AJV01733 AJV01746 AJV01748 AJV01749 AJV01819 AJV01820 AJV01824 AJV02086 AJV02088 AJV02089 AJV02090 AJV02091 AJV02093 AJV02097 AJV02099 AJV02208 AJV02306 AJV02311 AJV02320 AJV02321 AJV02322 AJV02323 AJV02324 AJV02325 AJV02326 AJV02327 AJV02328 AJV02329 AJV02330 AJV02331 AJV02332 AJV02333 AJV02334 AJV02335 AJV02337 AJV02338 AJV02341 AJV02342 AJV02343 AJV02344 AJV02345 AJV02346 AJV02347 AJV02349 AJV02350 AJV02351 AJV02352 AJV02353 AJV02354 AJV02355 AJV02356 AJV02357 AJV02358 AJV02359 AJV02965 AKV01257 AKV01272 AKV01303 AKV01304 AKV01305 AKV01306 AKV01307 AKV01308 AKV01309 AKV01310 AKV01311 AKV01312 AKV01313 AKV01323 AKV01324 AKV01325 AKV01326 AKV01327 AKV01328 AKV01329 AKV01333 AKV01334 AKV01336 AKV01337 AKV01338 AKV01339 AKV01345 AKV01347 AKV01348 AKV01350 AKV01351 AKV01355 AKV01358 AKV01362 AKV01920 AKV01944 BAV04617 BAV04618 BAV04619 BAV04620 BAV04621 BAV04623 BAV04624 BAV04625 BAV04626 BAV04629 BAV04799 BAV04800 BAV04801 BAV04802 BAV04803 BBV01262 BBV01263 BBV01264 BBV01265 BBV01266 BBV01267 BBV01268 BBV01269 BBV01270 BBV01271 BBV03347 BBV03353 BBV03354 BBV03356 BBV03357 BBV03361 BBV03366 BBV03555 BBV03627 BBV03628 BBV03629 BBV03630 BBV03631 BBV03632 BBV03633 BBV03634 BBV03635 BBV03636 BBV03637 BBV03638 BBV03639 BBV03640 BBV03641 BBV03642 BBV03643 BBV03644 BBV03645 BBV03646 BBV03647 BBV03648 BBV03649 BBV03652 BBV03653 BBV03654 BBV03655 BBV03656 BBV03657 BBV03658 BBV03659 BBV03710 BBV03711 BBV03712 BCV02567 BCV02590 BCV02598 BCV02604 BCV02607 BCV02608 ACV01035 ACV01054 ACV01058 ACV01061 ACV01092 ACV01099 ACV01500 ACV01501 ACV01503 ACV01504 ACV01505 ACV01510 ACV01511 ACV01512 ACV01515 ACV01519 ACV01525 ACV01526 ACV01529 ACV01537 ACV01538 ACV01539 ACV01540 ACV01542 ACV01545 ACV01547 ACV01552 ACV01554 ACV01556 ACV01557 ACV01560 ACV01563 ACV01565 ACV01566 ACV01571 ACV01575 ACV01577 ACV01580 ACV01582 ACV01583 ACV01584 ACV01592 ACV01593 ACV01594 ACV01601 ACV01602 ACV01603 ACV01604 ACV01605 ACV01606 ACV01607 ACV01608 ACV01610 ACV01611 ACV01616 ACV01620 ACV01623 ACV01625 ACV01629 ACV01630 ACV01632 ACV01633 ACV01634 ACV01635 ACV01637 ACV01638 ACV01639 ACV01640 ACV01646 ACV01647 ACV01649 ACV01650 ACV01653 ACV01654 ACV01655 ACV01658 ACV01668 ACV01679 ACV01681 ACV01682 ACV01691 ACV01692 ACV01696 ACV01704 ACV01708 ACV01709 ACV01711 ACV01716 ACV01717 ACV01719 ACV01720 ACV01726 ACV01739 ACV01744 ACV01751 ACV01752 ACV01757 ACV01759 ADV01001 ADV01006 ADV01025 ADV01031 ADV01037 ADV01044 ADV01048 ADV01050 ADV01052 ADV01055 ADV01057 ADV01059 ADV01061 ADV01186 ADV01192 ADV01207 ADV01226 ADV01241 ADV01251 ADV01257 ADV01259 ADV01263 ADV01265 ADV01282 ADV01283 ADV01289 ADV01292 ADV01293 ADV01297 ADV01299 ADV01304 ADV01318 ADV01319 ADV01326 ADV01327 ADV01328 ADV01329 ADV01331 ADV01336 ADV01338 ADV01342 ADV01349 ADV01351 ADV01354 ADV01369 ADV01372 ADV01390 ADV01414 ADV01422 ADV01435 ADV01436 ADV01437 ADV01451 ADV01501 ADV01505 ADV01513 ADV01516 ADV01522 ADV01530 ADV01531 ADV01534 ADV01535 ADV01537 ADV01538 ADV01539 ADV01540 ADV01544 ADV01546 ADV01552 ADV01553 ADV01554 ADV01555 ADV01556 ADV01557 ADV01559 ADV01563 ADV01564 ADV01565 ADV01568 ADV01570 ADV01571 ADV01572 ADV01573 ADV01575 ADV01578 ADV01581 ADV01583 ADV01585 ADV01587 ADV01589 ADV01592 ADV01595 ADV01597 ADV01603 ADV01615 ADV01616 ADV01620 ADV01622 ADV01635 ADV01637 ADV01639 ADV01640 ADV01641 ADV01645 ADV01647 ADV01651 ADV01655 ADV01657 ADV01660 ADV01661 ADV01664 ADV01666 ADV01667 ADV01668 ADV01670 ADV01672 ADV01674 ADV01675 ADV01676 ADV01677 ADV01680 ADV01681 ADV01682 ADV01685 ADV01687 ADV01689 ADV01692 ADV01696 ADV01698 ADV01699 ADV01702 ADV01703 ADV01706 ADV01708 ADV01715 ADV01719 ADV01720 ADV01729 ADV01730 ADV01738 ADV01744 ADV01746 ADV01748 ADV01758 ADV01760 ADV01765 ADV01781 ADV01790 ADV01794 ADV01796 ADV01803 ADV01807 ADV01808 ADV01809 ADV01811 ADV01813 ADV01822 ADV01833 ADV01846 ADV01856 ADV01860 ADV01865 ADV01867 ADV01872 ADV01873 ADV01875 ADV01876 ADV01878 ADV01881 ADV01884 ADV01886 ADV01888 ADV02002 ADV02004 ADV02006 ADV02007 ADV02011 ADV02012 ADV02015 ADV02019 ADV02020 ADV02021 ADV02023 ADV02025 ADV02030 ADV02032 ADV02033 ADV02037 ADV02042 ADV02049 ADV02050 ADV02051 ADV02055 ADV02057 ADV02058 ADV02059 ADV02062 ADV02065 ADV02066 ADV02068 ADV02069 ADV02070 ADV02071 ADV02073 ADV02081 ADV02083 ADV02092 ADV02093 ADV02096 ADV02097 ADV02098 ADV02100 ADV02101 ADV02104 ADV02107 ADV02109 ADV02113 ADV02116 ADV02119 ADV02120 ADV02121 ADV02123 ADV02127 ADV02128 ADV02129 ADV02130 ADV02131 ADV02133 ADV02134 ADV02135 ADV02138 ADV02142 ADV02148 ADV02151 ADV02153 ADV02154 ADV02160 ADV02161 ADV02163 ADV02180 ADV02185 ADV02207 ADV02220 ADV03000 ADV03001 ADV03002 ADV03003 ADV03005 ADV03006 ADV03007 ADV03008 ADV03009 ADV03010 ADV03018 ADV03019 ADV03021 ADV03022 ADV03024 ADV03025 ADV03026 ADV03027 ADV03031 ADV03033 ADV03034 ADV03035 ADV03036 ADV03037 ADV03039 ADV03042 ADV03045 ADV03047 ADV03048 ADV03049 ADV03050 ADV03051 ADV03052 ADV03053 ADV03055 ADV03056 ADV03059 ADV03065 ADV03067 ADV03069 ADV03099 ADV03100 ADV03103 ADV03104 ADV03108 ADV03110 ADV03111 ADV03112 ADV03113 ADV03114 ADV03115 ADV04001 ADV04002 ADV04004 ADV04006 ADV04008 ADV04009 ADV04010 ADV04013 ADV04015 ADV04017 ADV04018 ADV04020 ADV04021 ADV04029 ADV04031 ADV04033 ADV04041 ADV04050 ADV04051 ADV04054 ADV04055 ADV04056 ADV04057 ADV04058 ADV04098 ADV04100 ADV04101 ADV04102 ADV04103 ADV04105 ADV04107 ADV04110 ADV04111 ADV04114 ADV04115 ADV04116 ADV04117 ADV04120 ADV04121 ADV04123 ADV04124 ADV04125 ADV04126 ADV04127 ADV04129 ADV04130 ADV04132 ADV04136 ADV04137 ADV04457 ADV04458 ADV04459 ADV04460 ADV04461 ADV05000 ADV05001 ADV05002 ADV05003 ADV05004 ADV05006 ADV05007 ADV05008 ADV05015 ADV05017 ADV05020 ADV05022 ADV05023 ADV05024 ADV05026 ADV05027 ADV05028 ADV05029 ADV05031 ADV05033 ADV05035 ADV05037 ADV05045 ADV05047 ADV05147 ADV05157 ADV05158 ADV05161 ADV05162 ADV05168 ADV05172 ADV05173 ADV05174 ADV05177 ADV05178 ADV05179 ADV05180 ADV05181 ADV05182 ADV05183 ADV05184 ADV05185 ADV05188 ADV05189 ADV05192 ADV05194 ADV05196 ADV05200 ADV06000 ADV06002 ADV06003 ADV06005 ADV06006 ADV06008 ADV06010 ADV06017 ADV06029 ADV06032 ADV06033 ADV06034 ADV06035 ADV06038 ADV06041 ADV06047 ADV06048 ADV06050 ADV06052 ADV06054 ADV06065 ADV06066 ADV06070 ADV06072 ADV06073 ADV06077 ADV06079 ADV06081 ADV06082 ADV06084 ADV06088 ADV06091 ADV06093 ADV06095 ADV06096 ADV06098 ADV06101 ADV06102 ADV06103 ADV06105 ADV06109 ADV06111 ADV06115 ADV06119 ADV06123 ADV06129 ADV06132 ADV06133 ADV06140 ADV06144 ADV06145 ADV06146 ADV06147 ADV06152 ADV06155 ADV06162 ADV06164 ADV06165 ADV06168 ADV06169 ADV06173 ADV06174 ADV06177 ADV06179 ADV06180 ADV06181 ADV06183 ADV06184 ADV06186 ADV06187 ADV06189 ADV06190 ADV06195 ADV06196 ADV06199 ADV06201 ADV06204 ADV06205 ADV06207 ADV06209 ADV06210 ADV06212 ADV06214 ADV06216 ADV06219 ADV06221 ADV06223 ADV06224 ADV06226 ADV06227 ADV06230 ADV06231 ADV06232 ADV06234 ADV06235 ADV06237 ADV06238 ADV06239 ADV06240 ADV06241 ADV06242 ADV06243 ADV06244 ADV06246 ADV06247 ADV06248 ADV06249 ADV06252 ADV06258 ADV06261 ADV06263 ADV06267 ADV06268 ADV06270 ADV06275 ADV06277 ADV06278 ADV06281 ADV06282 ADV06283 ADV06284 ADV06287 ADV06290 ADV06291 ADV06292 ADV06293 ADV06294 ADV06298 ADV06299 ADV06300 ADV06301 ADV06304 ADV06305 ADV06306 ADV06308 ADV06309 ADV06310 ADV06313 ADV06314 ADV06315 ADV06321 ADV06323 ADV06326 ADV06327 ADV06328 ADV06329 ADV06330 ADV06332 ADV06333 ADV06336 ADV06337 ADV06338 ADV06342 ADV06343 ADV06347 ADV06349 ADV06350 ADV06353 ADV06354 ADV06355 ADV06356 ADV06358 ADV06359 ADV06361 ADV06362 ADV06363 ADV06365 ADV06367 ADV06370 ADV06371 ADV06379 ADV06381 ADV06393 ADV06394 ADV06395 ADV06396 ADV06398 ADV06399 ADV06405 ADV06406 ADV06407 ADV06408 ADV06409 ADV06420 AEV01002 AEV01009 AEV01011 AEV01020 AEV01032 AEV01036 AEV01037 AEV01038 AEV01039 AEV01040 AEV01046 AEV01049 AEV01052 AEV01057 AEV01500 AEV01503 AEV01505 AEV01508 AEV01509 AEV01512 AEV01515 AEV01518 AEV01520 AEV01521 AEV01527 AEV01528 AEV01530 AEV01532 AEV01533 AEV01534 AEV01535 AEV01536 AEV01537 AEV01538 AEV01539 AEV01540 AEV01541 AEV01542 AEV01543 AEV01544 AEV01545 AEV01549 AEV01551 AEV01552 AEV01553 AEV01558 AEV01561 AEV01562 AEV01563 AEV01564 AEV01566 AEV01568 AEV01569 AEV01570 AEV01572 AEV01574 AEV01575 AEV01576 AEV01577 AEV01578 AEV01579 AEV01580 AEV01583 AEV01584 AEV01586 AEV01587 AEV01589 AEV01590 AEV01591 AEV01592 AEV01593 AEV01594 AEV01595 AEV01597 AEV01600 AEV01602 AEV01603 AEV01604 AEV01605 AEV01606 AEV01607 AEV01608 AEV01609 AEV01610 AEV01611 AEV01612 AEV01613 AEV01614 AEV01615 AEV01616 AEV01617 AEV01618 AEV01619 AEV01620 AEV01621 AEV01622 AEV01624 AEV01625 AEV01626 AEV01627 AEV01628 AEV01629 AEV01631 AEV01632 AEV01633 AEV01638 AEV01639 AEV01640 AEV01641 AEV01642 AEV01645 AEV01646 AEV01647 AEV01648 AEV01651 AEV01652 AEV01653 AEV01654 AEV01656 AEV01657 AEV01658 AEV01660 AEV01661 AEV01662 AEV01663 AEV01664 AEV01665 AEV01670 AEV01674 AEV01675 AEV01677 AEV01679 AEV01682 AEV01683 AEV01684 AEV01685 AEV01687 AEV01688 AEV01689 AEV01690 AEV01691 AEV01692 AEV01694 AEV01695 AEV01696 AEV01697 AEV01701 AEV01703 AEV01704 AEV01708 AEV01709 AEV01710 AEV01711 AEV01712 AEV01713 AEV01714 AEV01715 AEV01716 AEV01718 AEV01719 AEV01720 AEV01721 AEV01723 AEV01724 AEV01725 AEV01726 AEV01727 AEV01728 AEV01729 AEV01730 AEV01731 AEV01733 AEV01741 AEV01742 AEV01744 AEV01745 AEV01749 AEV01753 AEV01756 AEV01757 AEV01761 AEV01762 AEV01763 AEV01764 AEV01765 AEV01766 AEV01768 AEV01770 AEV01771 AEV01772 AEV01773 AEV01774 AEV01775 AEV01776 AEV01777 AEV01779 AEV01780 AEV01781 AEV01782 AEV01783 AEV01785 AEV01788 AEV01791 AEV01792 AEV01793 AEV01796 AEV01797 AEV01798 AEV01799 AEV01801 AEV01802 AEV01806 AEV01807 AEV01809 AEV01810 AEV01812 AEV01813 AEV01814 AEV01815 AEV01816 AEV01817 AEV01818 AEV01821 AEV01822 AEV01825 AEV01829 AEV01831 AEV01834 AEV01835 AEV01837 AEV01840 AEV01841 AEV01843 AEV01844 AEV01845 AEV01846 AEV01847 AEV01851 AEV01854 AEV01855 AEV01859 AEV01865 AEV01866 AEV01870 AEV01876 AEV01883 AEV01884 AEV01886 AEV01887 AEV01888 AEV01894 AEV01896 AEV01897 AEV01903 AEV01906 AEV01908 AEV01909 AEV01911 AEV01912 AEV01914 AEV01919 AEV01921 AEV01922 AEV01923 AEV01926 AEV01928 AEV01929 AEV01930 AEV01931 AEV01932 AEV01933 AEV01934 AEV01936 AEV01937 AEV01939 AEV01942 AEV01943 AEV01944 AEV01945 AEV01946 AEV01947 AEV01948 AEV01951 AEV01952 AEV01953 AEV01955 AEV01958 AEV01961 AEV01962 AEV01964 AEV01968 AEV01969 AEV01972 AEV01974 AEV01978 AEV01979 AEV01981 AEV01984 AEV01986 AEV01987 AEV01993 AEV01995 AEV01996 AEV01997 AEV02002 AEV02003 AEV02011 AEV02018 AEV02019 AEV02021 AEV02023 AEV02026 AEV02029 AEV02030 AEV02032 AEV02033 AEV02034 AEV02037 AEV02041 AEV02043 AEV02046 AEV02050 AEV02051 AEV02052 AEV02056 AEV02060 AEV02062 AEV02063 AEV02064 AEV02066 AEV02070 AEV02071 AEV02072 AEV02076 AEV02078 AEV02079 AEV02080 AEV02082 AEV02087 AEV02088 AEV02089 AEV02091 AEV02092 AEV02094 AEV02095 AEV02099 AEV02102 AEV02106 AEV02108 AEV02110 AEV02114 AEV02117 AEV02118 AEV02119 AEV02120 AEV02122 AEV02123 AEV02124 AEV02125 AEV02130 AEV02136 AEV02138 AEV02141 AEV02142 AEV02144 AEV02145 AEV02148 AEV02151 AEV02154 AEV02158 AEV02160 AEV02161 AEV02162 AEV02164 AEV02165 AEV02166 AEV02168 AEV02169 AEV02170 AEV02171 AEV02172 AEV02174 AEV02175 AEV02177 AEV02178 AEV02179 AEV02180 AEV02181 AEV02182 AEV02183 AEV02184 AEV02185 AEV02186 AEV02187 AEV02188 AEV02192 AEV02197 AEV02202 AEV02204 AEV02205 AEV02206 AEV02207 AEV02208 AEV02209 AEV02211 AEV02213 AEV02214 AEV02221 AEV02222 AEV02225 AEV02226 AEV02228 AEV02229 AEV02230 AEV02233 AEV02234 AEV02236 AEV02237 AEV02239 AEV02243 AEV02244 AEV02248 AEV02249 AEV02252 AEV02253 AEV02254 AEV02255 AEV02258 AEV02259 AEV02260 AEV02261 AEV02262 AEV02264 AEV02265 AEV02266 AEV02268 AEV02269 AEV02272 AEV02274 AEV02275 AEV02276 AEV02277 AEV02279 AEV02283 AEV02285 AEV02287 AEV02288 AEV02292 AEV02293 AEV02296 AEV02298 AEV02299 AEV02300 AEV02301 AEV02303 AEV02306 AEV02309 AEV02310 AEV02311 AEV02313 AEV02318 AEV02320 AEV02321 AEV02323 AEV02325 AEV02326 AEV02327 AEV02328 AEV02329 AEV02331 AEV02332 AEV02333 AEV02334 AEV02337 AEV02339 AEV02340 AEV02341 AEV02342 AEV02347 AEV02350 AEV02351 AEV02355 AEV02356 AEV02357 AEV02358 AEV02359 AEV02360 AEV02361 AEV02362 AEV02363 AEV02364 AEV02365 AEV02366 AEV02367 AEV02369 AEV02370 AEV02371 AEV02372 AEV02373 AEV02375 AEV02376 AEV02378 AEV02380 AEV02381 AEV02383 AEV02384 AEV02385 AEV02386 AEV02387 AEV02390 AEV02391 AEV02394 AEV02395 AEV02406 AEV02407 AEV02409 AEV02411 AEV02412 AEV02414 AEV02415 AEV02417 AEV02419 AEV02420 AEV02421 AEV02422 AEV02423 AEV02424 AEV02425 AEV02426 AEV02427 AEV02428 AEV02430 AEV02432 AEV02435 AEV02440 AEV02441 AEV02442 AEV02443 AEV02445 AEV02446 AEV02447 AEV02448 AEV02449 AEV02450 AEV02455 AEV02456 AEV02461 AEV02466 AEV02471 AEV02472 AEV02473 AEV02474 AEV02475 AEV02476 AEV02477 AEV02478 AEV02480 AEV02482 AEV02484 AEV02488 AEV02491 AEV02495 AEV02497 AEV02499 AEV02500 AEV02504 AEV02505 AEV02506 AEV02507 AEV02513 AEV02515 AEV02518 AEV02521 AEV02523 AEV02524 AEV02525 AEV02526 AEV02537 AEV02544 AEV02545 AEV02547 AEV02548 AEV02549 AEV02551 AEV02552 AEV02554 AEV02556 AEV02557 AEV02561 AEV02562 AEV02565 AEV02566 AEV02571 AEV02574 AEV02575 AEV02576 AEV02577 AEV02580 AEV02581 AEV02585 AEV02586 AEV02587 AEV02588 AEV02589 AEV02590 AEV02591 AEV02594 AEV02601 AEV02602 AEV02608 AEV02612 AEV02613 AEV02614 AEV02617 AEV02620 AEV02627 AEV02629 AEV02631 AEV02635 AEV02636 AEV02643 AEV02644 AEV02648 AEV02653 AEV02654 AEV02656 AEV02657 AEV02662 AEV02665 AEV02666 AEV02667 AEV02670 AEV02674 AEV02675 AEV02679 AEV02681 AEV02682 AEV02684 AEV02685 AEV02691 AEV02693 AEV02694 AEV02695 AEV02696 AEV02697 AEV02699 AEV02700 AEV02704 AEV02706 AEV02707 AEV02708 AEV02709 AEV02710 AEV02712 AEV02713 AEV02717 AEV02718 AEV02719 AEV02720 AEV02721 AEV02723 AEV02725 AEV02726 AEV02727 AEV02728 AEV02729 AEV02730 AEV02736 AEV02738 AEV02739 AEV02743 AEV02744 AEV02745 AEV02746 AEV02747 AEV02750 AEV02752 AEV02753 AEV02755 AEV02757 AEV02760 AEV02761 AEV02763 AEV02766 AEV02768 AFV01001 AFV01002 AFV01003 AFV01005 AFV01006 AFV01008 AFV01010 AFV01013 AFV01017 AFV01022 AFV01028 AFV01029 AFV01030 AFV01034 AFV01037 AFV01040 AFV01042 AFV01043 AFV01047 AFV01048 AFV01053 AFV01054 AFV01055 AFV01057 AFV01062 AFV01064 AFV01065 AFV01068 AFV01072 AFV01075 AFV01076 AFV01078 AFV01079 AFV01081 AFV01090 AFV01092 AFV01097 AFV01105 AFV01114 AFV01117 AFV01142 AFV01150 AFV01151 AFV01157 AFV01158 AFV01159 AFV01163 AFV01166 AFV01169 AFV01175 AFV01176 AFV01177 AFV01179 AFV01181 AFV01184 AFV01185 AFV01189 AFV01190 AFV01191 AFV01192 AFV01193 AFV01194 AFV01195 AFV01200 AFV01202 AFV01203 AFV01209 AFV01211 AFV01212 AFV01214 AFV01216 AFV01217 AFV01220 AFV01221 AFV01222 AFV01225 AFV01230 AFV01233 AFV01237 AFV01238 AFV01241 AFV01242 AFV01245 AFV01251 AFV01255 AFV01256 AFV01257 AFV01258 AFV01259 AFV01260 AFV01262 AFV01264 AFV01265 AFV01267 AFV01268 AFV01271 AFV01272 AFV01275 AFV01277 AFV01280 AFV01284 AFV01285 AFV01286 AFV01287 AFV01289 AFV01291 AFV01292 AFV01293 AFV01294 AFV01295 AFV01296 AFV01298 AFV01302 AFV01309 AFV01311 AFV01313 AFV01315 AFV01320 AFV01321 AFV01322 AFV01323 AFV01324 AFV01325 AFV01327 AFV01331 AFV01333 AFV01334 AFV01335 AFV01336 AFV01337 AFV01338 AFV01339 AFV01341 AFV01343 AFV01346 AFV01352 AFV01355 AFV01359 AFV01360 AFV01361 AFV01362 AFV01363 AFV01365 AFV01367 AFV01368 AFV01369 AFV01370 AFV01371 AFV01372 AFV01373 AFV01374 AFV01375 AFV01379 AFV01380 AFV01381 AFV01382 AFV01383 AFV01385 AFV01386 AFV01387 AFV01388 AFV01389 AFV01390 AFV01392 AFV01393 AFV01394 AFV01395 AFV01396 AFV01398 AFV01400 AFV01401 AFV01402 AFV01403 AFV01404 AFV01406 AFV01407 AFV01408 AFV01409 AFV01413 AFV01416 AFV01417 AFV01418 AFV01419 AFV01421 AFV01422 AFV01423 AFV01424 AFV01425 AFV01427 AFV01428 AFV01429 AFV01430 AFV01431 AFV01433 AFV01434 AFV01435 AFV01436 AFV01437 AFV01438 AFV01440 AFV01441 AFV01442 AFV01443 AFV01444 AFV01447 AFV01449 AFV01451 AFV01452 AFV01454 AFV01456 AFV01458 AFV01460 AFV01462 AFV01463 AFV01464 AFV01466 AFV01467 AFV01468 AFV01471 AFV01472 AFV01473 AFV01475 AFV01476 AFV01477 AFV01478 AFV01479 AFV01480 AFV01481 AFV01482 AFV01484 AFV01485 AFV01487 AFV01488 AFV01489 AFV01490 AFV01491 AFV01494 AFV01495 AFV01496 AFV01499 AFV01500 AFV01503 AFV01504 AFV01505 AFV01510 AFV01511 AFV01514 AFV01521 AFV01523 AFV01524 AFV01528 AFV01530 AFV01532 AFV01533 AFV01534 AFV01536 AFV01537 AFV01540 AFV01541 AFV01543 AFV01544 AFV01545 AFV01546 AFV01549 AFV01550 AFV01551 AFV01552 AFV01553 AFV01554 AFV01556 AFV01557 AFV01558 AFV01559 AFV01560 AFV01563 AFV01565 AFV01566 AFV01567 AFV01568 AFV01569 AFV01570 AFV01571 AFV01572 AFV01573 AFV01574 AFV01575 AFV01576 AFV01577 AFV01578 AFV01579 AFV01580 AFV01582 AFV01584 AFV01585 AFV01586 AFV01587 AFV01588 AFV01589 AFV01591 AFV01599 AFV01600 AFV01601 AFV01602 AFV01603 AFV01610 AFV01612 AFV01616 AFV01618 AFV01619 AFV01620 AFV01622 AFV01623 AFV01626 AFV01627 AFV01628 AFV01629 AFV01630 AFV01631 AFV01632 AFV01633 AFV01634 AFV01635 AFV01637 AFV01640 AFV01641 AFV01643 AFV01644 AFV01645 AFV01647 AFV01649 AFV01652 AFV01653 AFV01654 AFV01655 AFV01656 AFV01657 AFV01658 AFV01661 AFV01662 AFV01665 AFV01666 AFV01668 AFV01677 AFV01685 AFV01686 AFV01688 AFV01690 AFV01695 AFV01697 AFV01704 AFV01705 AFV01706 AFV01717 AFV01719 AFV01720 AFV01721 AFV01722 AFV01729 AFV01732 AFV01736 AFV01742 AFV01744 AFV01745 AFV01746 AFV01747 AFV01748 AFV01756 AFV01757 AFV01758 AFV01761 AFV01762 AFV01767 AFV01769 AFV01776 AFV01780 AFV01781 AFV01787 AFV01789 AFV01790 AFV01791 AFV01792 AFV01793 AFV01794 AFV01795 AFV01798 AFV01799 AFV01800 AFV01801 AFV01802 AFV01803 AFV01804 AFV01806 AFV01808 AFV01810 AFV01811 AFV01815 AFV01818 AFV01819 AFV01822 AFV01824 AFV01828 AFV01830 AFV01831 AFV01834 AFV01835 AFV01836 AFV01837 AFV01838 AFV01840 AFV01842 AFV01844 AFV01845 AFV01846 AFV01848 AFV01849 AFV01851 AFV01852 AFV01854 AFV01856 AFV01857 AFV01858 AFV01860 AFV01862 AFV01863 AFV01864 AFV01865 AFV01866 AFV01867 AFV01868 AFV01869 AFV01870 AFV01871 AFV01873 AFV01874 AFV01875 AFV01876 AFV01877 AFV01878 AFV01879 AFV01880 AFV01885 AFV01887 AFV01888 AFV01889 AFV01891 AFV01892 AFV01893 AFV01895 AFV01898 AFV01899 AFV01902 AFV01905 AFV01907 AFV01908 AFV01909 AFV01911 AFV01917 AFV01919 AFV01922 AFV01928 AFV01932 AFV01933 AFV01934 AFV01941 AFV01944 AFV01946 AFV01947 AFV01954 AFV01958 AFV01960 AFV01962 AFV01963 AFV01964 AFV01981 AFV01985 AFV01986 AFV01988 AFV01990 AFV01991 AFV01992 AFV01994 AFV01995 AFV02000 AFV02003 AFV02004 AFV02008 AFV02011 AFV02012 AFV02013 AFV02014 AFV02015 AFV02016 AFV02017 AFV02019 AFV02020 AFV02021 AFV02022 AFV02023 AFV02025 AFV02026 AFV02028 AFV02030 AFV02031 AFV02034 AFV02035 AFV02036 AFV02037 AFV02038 AFV02039 AFV02040 AFV02041 AFV02043 AFV02044 AFV02045 AFV02047 AFV02048 AFV02052 AFV02053 AFV02054 AFV02055 AFV02056 AFV02058 AFV02059 AFV02060 AFV02061 AFV02062 AFV02063 AFV02065 AFV02067 AFV02070 AFV02072 AFV02074 AFV02075 AFV02076 AFV02078 AFV02079 AFV02080 AFV02082 AFV02083 AFV02087 AFV02088 AFV02090 AFV02092 AFV02093 AFV02095 AFV02096 AFV02097 AFV02098 AFV02099 AFV02101 AFV02102 AFV02104 AFV02105 AFV02106 AFV02109 AFV02113 AFV02114 AFV02115 AFV02117 AFV02119 AFV02120 AFV02122 AFV02123 AFV02126 AFV02127 AFV02129 AFV02130 AFV02132 AFV02133 AFV02134 AFV02136 AFV02137 AFV02141 AFV02142 AFV02143 AFV02144 AFV02145 AFV02146 AFV02147 AFV02150 AFV02152 AFV02153 AFV02155 AFV02156 AFV02157 AFV02159 AFV02162 AFV02164 AFV02165 AFV02168 AFV02170 AFV02171 AFV02172 AFV02173 AFV02174 AFV02176 AFV02177 AFV02178 AFV02179 AFV02182 AFV02183 AFV02187 AFV02188 AFV02190 AFV02191 AFV02192 AFV02193 AFV02196 AFV02197 AFV02201 AFV02205 AFV02206 AFV02207 AFV02208 AFV02209 AFV02210 AFV02211 AFV02214 AFV02215 AFV02217 AFV02225 AFV02226 AFV02227 AFV02228 AFV02229 AFV02231 AFV02232 AFV02233 AFV02234 AFV02235 AFV02236 AFV02237 AFV02238 AFV02239 AFV02240 AFV02242 AFV02244 AFV02245 AFV02246 AFV02247 AFV02248 AFV02250 AFV02251 AFV02253 AFV02254 AFV02257 AFV02260 AFV02261 AFV02262 AFV02263 AFV02264 AFV02266 AFV02268 AFV02269 AFV02270 AFV02273 AFV02274 AFV02275 AFV02277 AFV02279 AFV02280 AFV02281 AFV02282 AFV02283 AFV02284 AFV02287 AFV02288 AFV02289 AFV02291 AFV02292 AFV02293 AFV02294 AFV02295 AFV02296 AFV02297 AFV02298 AFV02299 AFV02301 AFV02302 AFV02303 AFV02306 AFV02307 AFV02308 AFV02310 AFV02311 AFV02317 AFV02318 AFV02319 AFV02322 AFV02323 AFV02324 AFV02325 AFV02326 AFV02328 AFV02329 AFV02332 AFV02334 AFV02341 AFV02344 AFV02346 AFV02347 AFV02352 AFV02353 AFV02354 AFV02355 AFV02356 AFV02357 AFV02359 AFV02363 AFV02365 AFV02366 AFV02368 AFV02369 AFV02371 AFV02372 AFV02376 AFV02378 AFV02379 AFV02380 AFV02381 AFV02382 AFV02385 AFV02387 AFV02388 AFV02394 AFV02395 AFV02397 AFV02398 AFV02399 AFV02401 AFV02402 AFV02404 AFV02405 AFV02407 AFV02412 AFV02413 AFV02415 AFV02416 AFV02417 AFV02418 AFV02420 AFV02421 AFV02422 AFV02423 AFV02424 AFV02427 AFV02429 AFV02433 AFV02434 AFV02435 AFV02438 AFV02439 AFV02441 AFV02443 AFV02444 AFV02446 AFV02447 AFV02448 AFV02449 AFV02450 AFV02451 AFV02453 AFV02455 AFV02458 AFV02460 AFV02461 AFV02462 AFV02463 AFV02464 AFV02466 AFV02473 AFV02479 AFV02480 AFV02481 AFV02483 AFV02484 AFV02485 AFV02486 AFV02487 AFV02488 AFV02490 AFV02491 AFV02492 AFV02493 AFV02494 AFV02495 AFV02497 AFV02499 AFV02500 AFV02501 AFV02502 AFV02504 AFV02505 AFV02506 AFV02507 AFV02508 AFV02511 AFV02512 AFV02513 AFV02514 AFV02516 AFV02517 AFV02518 AFV02519 AFV02520 AFV02521 AFV02522 AFV02523 AFV02524 AFV02525 AFV02526 AFV02527 AFV02528 AFV02529 AFV02530 AFV02531 AFV02532 AFV02533 AFV02534 AFV02535 AFV02536 AFV02537 AFV02538 AFV02539 AFV02540 AFV02542 AFV02543 AFV02545 AFV02547 AFV02548 AFV02549 AFV02550 AFV02551 AFV02552 AFV02553 AFV02554 AFV02555 AFV02556 AFV02557 AFV02558 AFV02559 AFV02560 AFV02561 AFV02562 AFV02563 AFV02564 AFV02565 AFV02566 AFV02567 AFV02568 AFV02569 AFV02570 AFV02571 AFV02572 AFV02573 AFV02574 AFV02575 AFV02576 AFV02577 AFV02578 AFV02579 AFV02580 AFV02581 AFV02582 AFV02583 AFV02584 AFV02585 AFV02586 AFV02587 AFV02588 AFV02589 AFV02590 AFV02591 AFV02592 AFV02593 AFV02594 AFV02595 AFV02596 AFV02597 AFV02598 AFV02599 AFV02600 AFV02601 AFV02602 AFV02603 AFV02604 AFV02605 AFV02606 AFV02608 AFV02609 AFV02610 AFV02611 AFV02612 AFV02613 AFV02614 AFV02615 AFV02616 AFV02617 AFV02619 AFV02621 AFV02623 AFV02624 AFV02626 AFV02770 AGV01000 AGV01001 AGV01002 AGV01003 AGV01004 AGV01005 AGV01006 AGV01007 AGV01009 AGV01010 AGV01014 AGV01020 AGV01022 AGV01023 AGV01024 AGV01025 AGV01027 AGV01031 AGV01034 AGV01037 AGV01038 AGV01040 AGV01041 AGV01042 AGV01043 AGV01044 AGV01045 AGV01046 AGV01047 AGV01048 AGV01049 AGV01050 AGV01051 AGV01052 AGV01053 AGV01055 AGV01056 AGV01061 AGV01064 AGV01068 AGV01071 AGV01074 AGV01082 AGV01087 AGV01095 AGV01098 AGV01100 AGV01104 AGV01106 AGV01109 AGV01113 AGV01116 AGV01119 AGV01120 AGV01121 AGV01126 AGV01127 AGV01128 AGV01129 AGV01131 AGV01132 AGV01136 AGV01138 AGV01140 AGV01146 AGV01149 AGV01151 AGV01152 AGV01155 AGV01156 AGV01157 AGV01160 AGV01161 AGV01162 AGV01163 AGV01165 AGV01166 AGV01169 AGV01170 AGV01173 AGV01175 AGV01177 AGV01178 AGV01179 AGV01180 AGV01183 AGV01186 AGV01188 AGV01193 AGV01195 AGV01196 AGV01197 AGV01198 AGV01199 AGV01200 AGV01201 AGV01202 AGV01203 AGV01204 AGV01205 AGV01206 AGV01207 AGV01209 AGV01210 AGV01212 AGV01214 AGV01215 AGV01216 AGV01217 AGV01219 AGV01220 AGV01221 AGV01222 AGV01225 AGV01226 AGV01227 AGV01228 AGV01229 AGV01230 AGV01233 AGV01235 AGV01237 AGV01238 AGV01239 AGV01241 AGV01242 AGV01243 AGV01244 AGV01248 AGV01249 AGV01250 AGV01251 AGV01252 AGV01254 AGV01255 AGV01256 AGV01257 AGV01258 AGV01259 AGV01261 AGV01262 AGV01263 AGV01268 AGV01269 AGV01270 AGV01271 AGV01275 AGV01277 AGV01278 AGV01279 AGV01280 AGV01281 AGV01284 AGV01286 AGV01287 AGV01288 AGV01293 AGV01295 AGV01299 AGV01300 AGV01301 AGV01304 AGV01305 AGV01312 AGV01314 AGV01317 AGV01318 AGV01321 AGV01322 AGV01323 AGV01327 AGV01328 AGV01329 AGV01333 AGV01334 AGV01335 AGV01338 AGV01349 AGV01352 AGV01353 AGV01358 AGV01362 AGV01379 AGV01382 AGV01385 AGV01390 AGV01393 AGV01402 AGV01405 AGV01410 AGV01412 AGV01413 AGV01414 AGV01415 AGV01426 AGV01429 AGV01441 AGV01443 AGV01445 AGV01446 AGV01447 AGV01449 AGV01452 AGV01453 AGV01454 AGV01455 AGV01457 AGV01458 AGV01459 AGV01460 AGV01461 AGV01462 AGV01463 AGV01464 AGV01465 AGV01468 AGV01469 AGV01470 AGV01477 AGV01478 AGV01480 AGV01486 AGV01487 AGV01488 AGV01491 AGV01492 AGV01493 AGV01494 AGV01495 AGV01497 AGV01498 AGV01502 AGV01503 AGV01505 AGV01506 AGV01509 AGV01511 AGV01512 AGV01513 AGV01514 AGV01517 AGV01519 AGV01520 AGV01521 AGV01522 AGV01523 AGV01524 AGV01525 AGV01526 AGV01527 AGV01528 AGV01530 AGV01531 AGV01532 AGV01533 AGV01534 AGV01535 AGV01536 AGV01537 AGV01538 AGV01539 AGV01541 AGV01543 AGV01544 AGV01546 AGV01547 AGV01550 AGV01552 AGV01553 AGV01554 AGV01557 AGV01561 AGV01563 AGV01569 AGV01570 AGV01571 AGV01576 AGV01577 AGV01580 AGV01581 AGV01584 AGV01586 AGV01587 AGV01589 AGV01590 AGV01592 AGV01593 AGV01594 AGV01595 AGV01596 AGV01597 AGV01605 AGV01616 AGV01621 AGV01629 AGV01632 AGV01633 AGV01636 AGV01642 AGV01653 AGV01655 AGV01656 AGV01657 AGV01659 AGV01667 AGV01668 AGV01672 AGV01675 AGV01676 AGV01677 AGV01681 AGV01682 AGV01684 AGV01685 AGV01686 AGV01689 AGV01704 AGV01707 AGV01708 AGV01709 AGV01710 AGV01711 AGV01712 AGV01714 AGV01719 AGV01720 AGV01721 AGV01722 AGV01725 AGV01726 AGV01727 AGV01728 AGV01729 AGV01730 AGV01731 AGV01732 AGV01733 AGV01734 AGV01736 AGV01737 AGV01738 AGV01739 AGV01740 AGV01741 AGV01742 AGV01743 AGV01744 AGV01745 AGV01746 AGV01748 AGV01749 AGV01750 AGV01751 AGV01752 AGV01753 AGV01754 AGV01755 AGV01756 AGV01757 AGV01758 AGV01759 AGV01760 AGV01761 AGV01762 AGV01763 AGV01764 AGV01765 AGV01766 AGV01767 AGV01768 AGV01769 AGV01770 AGV01771 AGV01772 AGV01773 AGV01775 AGV01776 AGV01777 AGV01780 AGV01781 AGV01782 AGV01784 AGV01786 AGV01787 AGV01789 AGV01791 AGV01792 AGV01793 AGV01794 AGV01795 AGV01796 AGV01797 AGV01798 AGV01799 AGV01800 AGV01801 AGV01802 AGV01803 AGV01804 AGV01806 AGV01809 AGV01810 AGV01811 AGV01812 AGV01834 AGV01835 AGV01837 AGV01838 AGV01842 AGV01846 AGV01847 AGV01848 AGV01849 AGV01850 AGV01851 AGV01852 AGV01853 AGV01854 AGV01855 AGV01856 AGV01857 AGV01858 AGV01859 AGV01860 AGV01861 AGV01862 AGV01863 AGV01864 AGV01865 AGV01866 AGV01867 AGV01868 AGV01869 AGV01870 AGV01871 AGV01872 AGV01873 AGV01875 AGV01876 AGV01877 AGV01878 AGV01880 AGV01881 AGV01882 AGV01883 AGV01884 AGV01885 AGV01886 AGV01887 AGV01890 AGV01891 AGV01892 AGV01893 AGV01894 AGV01898 AGV01899 AGV01900 AGV01901 AGV01902 AGV01903 AGV01904 AGV01909 AGV01911 AGV01912 AGV01918 AGV01920 AGV01922 AGV01928 AGV01929 AGV01930 AGV01931 AGV01932 AGV01933 AGV01938 AGV01940 AGV01942 AGV01953 AGV01960 AGV01964 AGV01968 AGV01971 AGV01972 AGV01978 AGV01979 AGV01980 AGV01984 AGV01985 AGV01986 AGV01987 AGV01988 AGV01991 AGV01992 AGV01993 AGV01994 AGV01995 AGV01996 AGV01997 AGV01998 AGV02000 AGV02001 AGV02002 AGV02003 AGV02004 AGV02008 AGV02009 AGV02010 AGV02011 AGV02012 AGV02013 AGV02014 AGV02015 AGV02016 AGV02017 AGV02018 AGV02019 AGV02020 AGV02021 AGV02022 AGV02024 AGV02025 AGV02026 AGV02027 AGV02028 AGV02029 AGV02031 AGV02034 AGV02035 AGV02036 AGV02038 AGV02040 AGV02041 AGV02043 AGV02044 AGV02046 AGV02048 AGV02049 AGV02057 AGV02060 AGV02062 AGV02065 AGV02066 AGV02067 AGV02068 AGV02070 AGV02071 AGV02072 AGV02073 AGV02075 AGV02077 AGV02078 AGV02081 AGV02082 AGV02086 AGV02087 AGV02088 AGV02089 AGV02090 AGV02093 AGV02094 AGV02096 AGV02097 AGV02098 AGV02101 AGV02104 AGV02106 AGV02108 AGV02111 AGV02113 AGV02114 AGV02115 AGV02116 AGV02117 AGV02119 AGV02121 AGV02122 AGV02124 AGV02126 AGV02127 AGV02128 AGV02130 AGV02131 AGV02134 AGV02136 AGV02139 AGV02142 AGV02144 AGV02147 AGV02149 AGV02150 AGV02152 AGV02153 AGV02155 AGV02156 AGV02158 AGV02161 AGV02162 AGV02163 AGV02170 AGV02171 AGV02177 AGV02179 AGV02183 AGV02185 AGV02186 AGV02187 AGV02188 AGV02190 AGV02192 AGV02193 AGV02194 AGV02195 AGV02197 AGV02198 AGV02200 AGV02201 AGV02203 AGV02204 AGV02205 AGV02206 AGV02207 AGV02209 AGV02210 AGV02211 AGV02213 AGV02216 AGV02217 AGV02218 AGV02222 AGV02223 AGV02224 AGV02226 AGV02231 AGV02232 AGV02234 AGV02237 AGV02239 AGV02241 AGV02243 AGV02244 AGV02247 AGV02249 AGV02252 AGV02260 AGV02266 AGV02272 AGV02278 AGV02282 AGV02284 AGV02286 AGV02289 AGV02291 AGV02293 AGV02297 AGV02300 AGV02302 AGV02304 AGV02310 AGV02316 AGV02321 AGV02324 AGV02325 AGV02330 AGV02331 AGV02332 AGV02333 AGV02336 AGV02338 AGV02345 AGV02349 AGV02351 AGV02354 AGV02355 AGV02357 AGV02358 AGV02359 AGV02362 AGV02364 AGV02366 AGV02368 AGV02369 AGV02372 AGV02373 AGV02375 AGV02377 AGV02378 AGV02383 AGV02384 AGV02387 AGV02388 AGV02391 AGV02394 AGV02397 AGV02400 AGV02402 AGV02404 AGV02405 AGV02406 AGV02407 AGV02409 AGV02410 AGV02412 AGV02413 AGV02416 AGV02417 AGV02419 AGV02420 AGV02423 AGV02425 AGV02426 AGV02427 AGV02429 AGV02432 AGV02435 AGV02436 AGV02440 AGV02441 AGV02443 AGV02444 AGV02446 AGV02447 AGV02450 AGV02451 AGV02452 AGV02454 AGV02455 AGV02459 AGV02466 AGV02471 AGV02476 AGV02480 AGV02481 AGV02482 AGV02485 AGV02486 AGV02487 AGV02488 AGV02489 AGV02490 AGV02491 AGV02492 AGV02493 AGV02494 AGV02495 AGV02496 AGV02497 AGV02499 AGV02500 AGV02503 AGV02508 AGV02509 AGV02510 AGV02511 AGV02513 AGV02514 AGV02515 AGV02516 AGV02518 AGV02519 AGV02520 AGV02521 AGV02522 AGV02523 AGV02525 AGV02527 AGV02528 AGV02530 AGV02531 AGV02532 AGV02533 AGV02534 AGV02535 AGV02536 AGV02538 AGV02540 AGV02541 AGV02542 AGV02545 AGV02546 AGV02549 AGV02550 AGV02551 AGV02552 AGV02553 AGV02555 AGV02556 AGV02557 AGV02558 AGV02559 AGV02560 AGV02561 AGV02562 AGV02563 AGV02564 AGV02565 AGV02566
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2273-2013
Related Recalls
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).