SOMATOM Definition Flash, Model Number 10430603 - Product Usage: Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning. These CT systems can be used for low dose lunch cancer screening in high risk populations (as defined by professional medical societies).

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Serial Numbers: 73008 98642
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Siemens Medical Solutions USA, Inc
Reason for Recall:
Potential for four safety-related software issues impacting systems utilizing syngo.CT software version VB10A which may result in a scan abort.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

SOMATOM Definition Flash, Model Number 10430603 - Product Usage: Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning. These CT systems can be used for low dose lunch cancer screening in high risk populations (as defined by professional medical societies).

Product Codes/Lot Numbers:

Serial Numbers: 73008 98642

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2270-2019

Related Recalls

MAMMOMAT Inspiration;

Siemens Medical Solutions USA

Class I - Dangerous

There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.

Oct 24, 2025 Other Medical Devices Nationwide View Details →

MAMMOMAT Fusion;

Siemens Medical Solutions USA

Class I - Dangerous

There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.

Oct 24, 2025 Other Medical Devices Nationwide View Details →

MAMMOMAT Revelation;

Siemens Medical Solutions USA

Class I - Dangerous

There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.

Oct 24, 2025 Other Medical Devices Nationwide View Details →