Arcos Modular Revision Hip System Standard Cone Prox Body, Size B, 70 MM Item Number: 11-301322

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lot Numbers/UDI : 678990 (01)00880304473751(17)310407(10)678990
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Biomet, Inc.
Reason for Recall:
Arcos Proximal Cone Bodies could potentially exhibit chatter in the inner taper fretting corrosion and device failure in vivo leading to surgical intervention
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Arcos Modular Revision Hip System Standard Cone Prox Body, Size B, 70 MM Item Number: 11-301322

Product Codes/Lot Numbers:

Lot Numbers/UDI : 678990 (01)00880304473751(17)310407(10)678990

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2268-2021