BEAD TIP GD WIRE 3.0MM X 60CM; 29402 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended for drilling bone.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    563080 563090 407870 408160 408140 412690 412750 412720 412740 668840 202750 389030 410170 410180 395170 513920 201790 232930 639560 800410 980930 134560 271820 351760 504870 578400 605700 750960 750970 913730 921150 058010 631570 969080 101380 944200 969090 203000 449730 449740 449750 449720 665510 820130 850170 978340 221240 419010 848820 848810 278720 387620 437340 502200 623450 038560 186020 212630 183170 183180 183190 434630 183200 930950 693480
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Zimmer Biomet, Inc.
Reason for Recall:
The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the potential for sterile barrier breach, leading to possible risk for infection.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

BEAD TIP GD WIRE 3.0MM X 60CM; 29402 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended for drilling bone.

Product Codes/Lot Numbers:

563080 563090 407870 408160 408140 412690 412750 412720 412740 668840 202750 389030 410170 410180 395170 513920 201790 232930 639560 800410 980930 134560 271820 351760 504870 578400 605700 750960 750970 913730 921150 058010 631570 969080 101380 944200 969090 203000 449730 449740 449750 449720 665510 820130 850170 978340 221240 419010 848820 848810 278720 387620 437340 502200 623450 038560 186020 212630 183170 183180 183190 434630 183200 930950 693480

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2239-2018

Related Recalls

Zimmer Biomet is conducting a medical device correction to update the Instructions for Use (IFU) and the Surgical Techniques for the Polaris Spinal System Translation Screw and Cypher MIS Screw System. Users should be aware that changes were made to the contraindications section indicating that the Translation Screws should not be used with a direct current stimulation, as a high probability of construct failure exists when using Translation Screws containing CoCr with direct current simulation system.