Interface module for CDl System 500 Product Usage: The Interface Module for CDl 500 is the interface between the CDl 500 monitor and the system. The module performs the following functions: Communicates primary pump flow data to the CDl 500 Receives blood parameter data from the CDl 500

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    catalog number: 803479 and serial numbers:00009-00122, 00124-00210, 00213-00365, 00367-00458, and 00461-00596.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Terumo Cardiovascular Systems Corporation
Reason for Recall:
Terumo Cardiovascular Systems has received reports of malfunctions involving the loss of data transfer capabilities using the data transfer capabilities using modules on the Terumo Advanced Perfusion System 1. Some reports stated that the odor of smoke was as associated with the failures. The malfunctions occurred during set-up, during cardiopulmonary bypass, and during routine testing of the modu
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Interface module for CDl System 500 Product Usage: The Interface Module for CDl 500 is the interface between the CDl 500 monitor and the system. The module performs the following functions: Communicates primary pump flow data to the CDl 500 Receives blood parameter data from the CDl 500

Product Codes/Lot Numbers:

catalog number: 803479 and serial numbers:00009-00122, 00124-00210, 00213-00365, 00367-00458, and 00461-00596.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2233-2012

Related Recalls

Due to an error in assembly of the thermistor within some units, the affected probes will experience inaccuracy in measured shunt sensor temperature. Additionally, following changes in patient temperature without an in vivo recalibration, affected probes may experience inaccuracy in other measured parameters, including pCO2, pO2, pH, Potassium and calculated parameters, including BE, Bicarbonate, Arterial Oxygen Saturation, Delivered Oxygen, Area Under of the Curve of Delivered Oxygen, Oxygen Consumption, and Oxygen Extraction Ratio, which utilize temperature as an input. Inaccurate data may lead to inappropriate user response.