Serial interface module RS-485 Product Usage: The System 1 provides perfusion case data transfer through the use of the data transfer module. This module is a general purpose serial interface and is the interface between the TLink Data Management ystem or Sarns CCPro Data Perfusion oftware and the system. The data collected and exported by the data transfer module includes physiological data and device events to record the progress and events during a procedure.
Class I - DangerousWhat Should You Do?
- Check if you have this product: catalog number: 803518 and serial numbers:00007-00038, 00041, and 00043.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Terumo Cardiovascular Systems Corporation
- Reason for Recall:
- Terumo Cardiovascular Systems has received reports of malfunctions involving the loss of data transfer capabilities using the data transfer capabilities using modules on the Terumo Advanced Perfusion System 1. Some reports stated that the odor of smoke was as associated with the failures. The malfunctions occurred during set-up, during cardiopulmonary bypass, and during routine testing of the modu
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Serial interface module RS-485 Product Usage: The System 1 provides perfusion case data transfer through the use of the data transfer module. This module is a general purpose serial interface and is the interface between the TLink Data Management ystem or Sarns CCPro Data Perfusion oftware and the system. The data collected and exported by the data transfer module includes physiological data and device events to record the progress and events during a procedure.
Product Codes/Lot Numbers:
catalog number: 803518 and serial numbers:00007-00038, 00041, and 00043.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2231-2012
Related Recalls
CDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPM Probe Trial Kit
Terumo Cardiovascular Systems
Due to an error in assembly of the thermistor within some units, the affected probes will experience inaccuracy in measured shunt sensor temperature. Additionally, following changes in patient temperature without an in vivo recalibration, affected probes may experience inaccuracy in other measured parameters, including pCO2, pO2, pH, Potassium and calculated parameters, including BE, Bicarbonate, Arterial Oxygen Saturation, Delivered Oxygen, Area Under of the Curve of Delivered Oxygen, Oxygen Consumption, and Oxygen Extraction Ratio, which utilize temperature as an input. Inaccurate data may lead to inappropriate user response.
Terumo CDI OneView Monitoring System Hematocrit / Oxygen Saturation Probe, Catalog Number REF CD1754
Terumo Cardiovascular Systems
During some clinical cases, it has been observed that the SO2 value may periodically appear as dashes on the Touchscreen Display indicating unavailable data. In these cases, a yellow low-limit alarm will also be triggered to the user, due to the dashed-out value.
Pressure relief valve included in certain lots of cardiovascular procedure kits was reportedly opening at approximately 200-300mmHg versus 400mmHg, potential myocardial tissue damage potentially requiring surgical or medical/pharmacological intervention