Oxygen Masks, Elongated (under chin), Adult Med Conc., No Tubing, Product Code 104-E;
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot Number 102153, 102339, 102941, 103176, 103453, 103718, 103919, 104471, 105034, 105282, 105427, 105567, 105996, 106960, 108658, 112024, 112530
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- ConvaTec, Inc
- Reason for Recall:
- Reports of tubing detaching from oxygen masks either prior to use (in packaging) or during use were made by different end users and for different batches. The failures were caused by an improper solvent bond between the oxygen tubing and the oxygen mask connector. While the defect affects less than 2% of recalled product, there is a potential for harm when oxygen supply to the user is disrupted.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Oxygen Masks, Elongated (under chin), Adult Med Conc., No Tubing, Product Code 104-E;
Product Codes/Lot Numbers:
Lot Number 102153, 102339, 102941, 103176, 103453, 103718, 103919, 104471, 105034, 105282, 105427, 105567, 105996, 106960, 108658, 112024, 112530
Distribution:
Distributed in: CA, CT, FL, GA, IL, MO, NJ, NY, NC, OH, PA, RI, VA
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2228-2017
Related Recalls
Wound dressing may have foreign matter on the product.
Ostomy skin barrier package contains a pre-cut wafer of 28mm instead of 16mm and/or 35mm pre-cut wafers, as stated package labeling. If pre-cut wafers are too big the device cannot be used and if the pre-cut wafer is too small and the device user tries to apply, this could cause abrasion, stomal cut, and bleeding
The product does not meet sterility assurance level.