Magellan Diagnostics LeadCare Ultra and LeadCare Plus test Kits. Product Usage: The LeadCare Blood Lead Test Kit is for in vitro diagnostic use only. The Test Kit is for the quantitative measurement of lead in fresh whole blood. This product is for professional use only. Sensors and Treatment Reagent of the Test Kit are specific for lead only. Contents of this LeadCare Blood Lead Test Kit must only be used with a LeadCare Analyzer and Blood Lead Testing System.
Class I - DangerousWhat Should You Do?
- Check if you have this product: All lot codes
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
โ ๏ธ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Magellan Diagnostics, Inc.
- Reason for Recall:
- Underestimates the lead concentration of venous blood samples when the sample is analyzed with the LeadCare systems
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Magellan Diagnostics LeadCare Ultra and LeadCare Plus test Kits. Product Usage: The LeadCare Blood Lead Test Kit is for in vitro diagnostic use only. The Test Kit is for the quantitative measurement of lead in fresh whole blood. This product is for professional use only. Sensors and Treatment Reagent of the Test Kit are specific for lead only. Contents of this LeadCare Blood Lead Test Kit must only be used with a LeadCare Analyzer and Blood Lead Testing System.
Product Codes/Lot Numbers:
All lot codes
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2226-2017
Related Recalls
Calibration Buttons being used in the kitting process were incorrectly labeled as Lot 2323M for lot 2333M. The use of mislabeled calibration Button with Lot #2333 can lead to delayed or erroneous high or erroneous low results.
LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only. The test kit is for the quantitative measurement of lead in capillary whole blood.
Magellan Diagnostics
During an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes were found to not meet the minimum dimensional lip requirement and may have inadequate cap-to-tube seals. Manufacturing with these caps may result in cap-to-tube seals that allow leaking of HCl or may result in cap-to-tube seals that allow evaporation of treatment reagent. The consequence of this would be a change in the volume of treatment reagent and a change in the ratios of either added patient sample or control material to treatment reagent. The result of this might be a falsely elevated result (e.g., a patient false positive [PATFP] result or a Controls Out of Range-High [COOR-Hi] result).
LeadCare Ultra Blood Lead Test Kit Catalog Number 70-8098
Magellan Diagnostics
Magellan has received reports that control tests of either the Low-Control and/or the High-Control generated a low result. The issue may cause the underestimation of blood lead levels in patient samples.