SENSATION 7 Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES (CHINA) D684-00-0469-01C SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES (INDIA) D684-00-0469-07 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (CHINA) D684-00-0470-01C SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (INDIA) D684-00-0470-07 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (TURKEY) D684-00-0470-09 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES D684-00-0470-01 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES, US ONLY D684-00-0470-01U SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES, US ONLY D684-00-0469-01U SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES D684-00-0469-01 SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES, TURKEY D684-00-0469-09 PACKAGED INSERTION KIT - SENSATION 7 Fr. 34/40 cc IABs D884-00-0019-16

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Product Description UDI SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES (CHINA) 10607567109367 SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES (INDIA) 10607567109732 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (CHINA) 10607567109374 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (INDIA) 10607567109749 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (TURKEY) 10607567113135 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES 10607567106779 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES, US ONLY 10607567109558 SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES, US ONLY 10607567109541 SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES 10607567106755 SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES, TURKEY 10607567113128 PACKAGED INSERTION KIT - SENSATION 7 Fr. 34/40 cc IABs 10607567106724
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Datascope Corp.
Reason for Recall:
During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

SENSATION 7 Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES (CHINA) D684-00-0469-01C SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES (INDIA) D684-00-0469-07 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (CHINA) D684-00-0470-01C SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (INDIA) D684-00-0470-07 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (TURKEY) D684-00-0470-09 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES D684-00-0470-01 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES, US ONLY D684-00-0470-01U SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES, US ONLY D684-00-0469-01U SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES D684-00-0469-01 SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES, TURKEY D684-00-0469-09 PACKAGED INSERTION KIT - SENSATION 7 Fr. 34/40 cc IABs D884-00-0019-16

Product Codes/Lot Numbers:

Product Description UDI SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES (CHINA) 10607567109367 SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES (INDIA) 10607567109732 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (CHINA) 10607567109374 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (INDIA) 10607567109749 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (TURKEY) 10607567113135 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES 10607567106779 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES, US ONLY 10607567109558 SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES, US ONLY 10607567109541 SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES 10607567106755 SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES, TURKEY 10607567113128 PACKAGED INSERTION KIT - SENSATION 7 Fr. 34/40 cc IABs 10607567106724

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2217-2023

Related Recalls