Modular Hybrid Glenoid Base Large for the Comprehensive Shoulder System, The product is placed into an AlOx coated PET pouch and vacuum sealed with an Argon flush. The vacuum sealed product is then placed into a Tyvek/film pouch and that pouch is heat sealed. The fully pouched product is then labeled and placed into a carton with an IFU. The carton is then labeled and shrink wrapped. The labels also include a sterilization indication dot. This system is intended for primary and revision total shoulder arthroplasty, as well as hemi-arthroplasty
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model Number: 113956, Modular Hybrid Glenoid Base Large, Item Number: 113956, Lot Number: 270260, UDI Number: (01) 00880304462618 (17) 211204 (10) 270260
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Zimmer Biomet, Inc.
- Reason for Recall:
- Zimmer Biomet is conducting a voluntary recall for a single lot of the Modular Hybrid Glenoid Base Large for the Comprehensive Shoulder System after it was determined that two pieces from the lot were moved to finished goods and shipped despite having an open Non-Conformance Report (NCR). The NCR indicated that one piece from the lot had scratches and the other non-confirming threads.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Modular Hybrid Glenoid Base Large for the Comprehensive Shoulder System, The product is placed into an AlOx coated PET pouch and vacuum sealed with an Argon flush. The vacuum sealed product is then placed into a Tyvek/film pouch and that pouch is heat sealed. The fully pouched product is then labeled and placed into a carton with an IFU. The carton is then labeled and shrink wrapped. The labels also include a sterilization indication dot. This system is intended for primary and revision total shoulder arthroplasty, as well as hemi-arthroplasty
Product Codes/Lot Numbers:
Model Number: 113956, Modular Hybrid Glenoid Base Large, Item Number: 113956, Lot Number: 270260, UDI Number: (01) 00880304462618 (17) 211204 (10) 270260
Distribution:
Distributed in: US, OH, AZ, ID, PA, AR, MN, IL, CA, MD, NY
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2216-2017
Related Recalls
Device has the potential for fracture during use.
Polaris 5.5 Spinal System and Polaris 5.5/Cypher Spinal System, Thoracolumbosacral pedicle screw system
Zimmer Biomet
Zimmer Biomet is conducting a medical device correction to update the Instructions for Use (IFU) and the Surgical Techniques for the Polaris Spinal System Translation Screw and Cypher MIS Screw System. Users should be aware that changes were made to the contraindications section indicating that the Translation Screws should not be used with a direct current stimulation, as a high probability of construct failure exists when using Translation Screws containing CoCr with direct current simulation system.
Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.