Vital MIS Spinal Fixation system implants CANN POLY EXT TAB 5.5X35MM, Item Number 810M5535 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation. Pedicle screw fixation is indicated for skeletally mature patients and for adolescence patients.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI: (01)00887868324420; Lot Numbers: C19H0007 C19J0004 C19K0001 C19K0003 C20A0002 SBM108403 SBM118985 SBM122395
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Zimmer Biomet, Inc.
Reason for Recall:
There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Vital MIS Spinal Fixation system implants CANN POLY EXT TAB 5.5X35MM, Item Number 810M5535 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation. Pedicle screw fixation is indicated for skeletally mature patients and for adolescence patients.

Product Codes/Lot Numbers:

UDI: (01)00887868324420; Lot Numbers: C19H0007 C19J0004 C19K0001 C19K0003 C20A0002 SBM108403 SBM118985 SBM122395

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2214-2020

Related Recalls

Zimmer Biomet is conducting a medical device correction to update the Instructions for Use (IFU) and the Surgical Techniques for the Polaris Spinal System Translation Screw and Cypher MIS Screw System. Users should be aware that changes were made to the contraindications section indicating that the Translation Screws should not be used with a direct current stimulation, as a high probability of construct failure exists when using Translation Screws containing CoCr with direct current simulation system.

Nov 17, 2020 Other Medical Devices Nationwide View Details →