Assayed Bovine Multi-Sera Level 1, Model No. AL1027
Class I - DangerousWhat Should You Do?
- Check if you have this product: 205SL
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Randox Laboratories, Limited
- Reason for Recall:
- The firm is re-assigning the Mean of all Instruments target and range for Bile Acids (5th Generation Colorimetric) in the Assayed Bovine Multi-Sera Level 1 Control Lot 205SL.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Assayed Bovine Multi-Sera Level 1, Model No. AL1027
Product Codes/Lot Numbers:
205SL
Distribution:
Distributed in: UK
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2207-2019
Related Recalls
Randox Lipoprotein (a) - In vitro diagnostic (IVD) use as Randox Lipoprotein (a) assay on automated analysers. Catalog Number: LP2757
Randox Laboratories, Limited
Lipoprotein (a) on Roche Cobas c501 with nmol/l claims does not have a 510(k) or PMA.
Randox Lipoprotein (a) - In vitro diagnostic (IVD) use as Randox Lipoprotein (a) assay on automated analysers. Catalog Number: LP3403
Randox Laboratories, Limited
Lipoprotein (a) on Roche Cobas c501 with nmol/l claims does not have a 510(k) or PMA.
RX Daytona Plus (with ISE/without ISE); Models Nos. RX4040 (w/ISE) and RX4041 (without ISE) - Product Usage: A fully automated, random access, clinical chemistry analyzer complete with dedicated analyzer software. For use in clinical laboratories.
Randox Laboratories, Limited
Software version UI2550642107 for the RX Daytona + instrument released to correct the following issues: 1) The order of samples displayed is not sequential according to SID (sample identification number) when searching the run/results screen; 2) Incorrect color flagging of IQC as "out of range."