Achieva 3.0T; Model Numbers (REF): (1) 781177, (2) 781277, (3) 781278, (4) 781344, (5) 781345;
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model Numbers (REF): (1) 781177, (2) 781277, (3) 781278, (4) 781344, (5) 781345; UDI-DI: (1) None, (2) None, (3) 00884838004139, (4) None, (5) None; Serial Numbers: (1) 17008, 17011, 17017, 17030, 17035, 17038, 17039, 17043, 17049, 17051, 17056, 17061, 17072, 17075, 17079, 17093, 17094, 17099, 17114, 17130, 17134, 17139, 17142, 17144, 17146, 17147, 17149, 17161, 17167, 17168, 17170, 17171, 17178, 17179, 17182, 17183, 17185, 17187, 17189, 17216, 17217, 17226, 17233, 17234, 17235, 17243, 17251, 17256, 17270, 17279, 17295, 17296, 17304, 17316, 17319, 17324, 17334, 17347, 17349, 17358, 17362, 17365, 17374, 17379, 17384, 17392, 17393, 17393, 17402, 17404, 17414, 17418, 17426, 17427, 17431, 17432, 17443, 17445, 17446, 17448, 17450, 17452, 17453, 17456, 17469, 17471, 17494, 17497, 17500, 17523, 17535, 17558, 17561, 17596, 17599, 17607, 24007, 24009, 24010, 24019, 24037, 24038, 24049, 24050, 24053, 24056, 24059, 24060, 24066, 24068, 24070, 24078, 24079, 24081, 24089, 24091, 24096, 24097, 24099, 24100, 24107, 24114, 24116, 24120, 24125, 24130, 24131, 24138, 24154, 34148, (2) 24115, 34002, 34011, 34020, 34028, 34029, 34030, 34033, 34034, 34038, 34043, 34065, 34066, 34071, 34073, 34074, 34077, 34082, 34086, 34095, 34106, 34109, 34113, 34115, 34118, 34120, 34123, 34135, 34138, 34139, 34140, 34144, 34159, 34160, 34165, 34200, 34201, 34203, 34207, 34213, 34220, 34225, 34230, 34231, 34232, 34234, 34237, 34239, 34241, 34242, 34247, 34253, 34254, 34255, 34256, 34258, 34263, 34264, 34268, 34270, 34271, 34272, 34273, 34274, 34275, 34276, 34281, 34282, 34285, 34286, 34287, 34288, 34289, 34290, 34291, 34293, 34294, 34295, 34297, 34298, 34299, 34300, 34301, 34305, 34306, 34309, 34311, 73012, (3) 38002, 38003, 38007, 38012, 38017, 38018, 38021, 38022, 38033, 38034, 38043, 38044, 38047, 38054, 38055, 38056, 38057, 38059, 38060, 38061, 38064, 38069, 38072, 38079, 38085, 38086, 38087, 38090, 38091, 38093, 38095, 38098, 38099, 38103, 38104, 38108, 38112, 38115, 38130, 38133, 38137, 38139, 38140, 38141, 38143, 38144, 38145, 38150, 38151, 38152, 38157, 38158, 38160, 38167, 38169, 38171, 38173, 38174, 38181, 38185, 38186, 38187, 38188, 38190, 38191, 38194, 38196, 38200, 38206, 38207, 38209, 38210, 38212, 38215, 38216, 38220, 38223, 38227, 38228, 38229, 38231, 38232, 38233, 38236, 38238, 38240, 38241, 38245, 38246, 38247, 38248, 38249, 38251, 38252, 38253, 38254, 38255, 38256, 38257, 38258, 38260, 38261, 38262, 38264, 38265, 38266, 38267, 38268, 38269, 38270, 38271, 38272, 38273, 38274, 38276, 38278, 38279, 38280, 38282, 38283, 38284, 38287, 38288, 38291, 38292, 38293, 38295, 38296, 38297, 38298, 38299, 38302, 38303, 38304, 38307, 38312, 38314, 38315, 38318, 38319, 38320, 38321, 38322, 38323, 38324, 38326, 38327, 38328, 38329, 38330, 38331, 38332, 38335, 38337, 38338, 38339, 38340, 38341, 38343, 38344, 38345, 38346, 38347, 38349, 38351, 38355, 38358, 38359, 38360, 38361, 38363, 38367, 38368, 38370, 38372, 38401, 38402, 38405, 38407, 38408, 38410, 38412, 38414, 38415, 38420, 38421, 38423, 38424, 38425, 38426, 38427, 38429, 38430, 38431, 38432, 38433, 38436, 38437, 38438, 38439, 38441, 38442, 38443, 38445, 38447, 38448, 38450, 38453, 38454, 38456, 38457, 38458, 382320, (4) 73003, 73005, 73006, 73008, 73009, 73010, (5) 74001, 74005, 74007, 74009, 74010, 74020, 74021, 74026, 74027, 74029, 74030, 74032, 74033, 74034, 74035, 74037, 74039, 74041, 74044, 74046, 74047, 74048, 74052;
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Philips North America
- Reason for Recall:
- The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Achieva 3.0T; Model Numbers (REF): (1) 781177, (2) 781277, (3) 781278, (4) 781344, (5) 781345;
Product Codes/Lot Numbers:
Model Numbers (REF): (1) 781177, (2) 781277, (3) 781278, (4) 781344, (5) 781345; UDI-DI: (1) None, (2) None, (3) 00884838004139, (4) None, (5) None; Serial Numbers: (1) 17008, 17011, 17017, 17030, 17035, 17038, 17039, 17043, 17049, 17051, 17056, 17061, 17072, 17075, 17079, 17093, 17094, 17099, 17114, 17130, 17134, 17139, 17142, 17144, 17146, 17147, 17149, 17161, 17167, 17168, 17170, 17171, 17178, 17179, 17182, 17183, 17185, 17187, 17189, 17216, 17217, 17226, 17233, 17234, 17235, 17243, 17251, 17256, 17270, 17279, 17295, 17296, 17304, 17316, 17319, 17324, 17334, 17347, 17349, 17358, 17362, 17365, 17374, 17379, 17384, 17392, 17393, 17393, 17402, 17404, 17414, 17418, 17426, 17427, 17431, 17432, 17443, 17445, 17446, 17448, 17450, 17452, 17453, 17456, 17469, 17471, 17494, 17497, 17500, 17523, 17535, 17558, 17561, 17596, 17599, 17607, 24007, 24009, 24010, 24019, 24037, 24038, 24049, 24050, 24053, 24056, 24059, 24060, 24066, 24068, 24070, 24078, 24079, 24081, 24089, 24091, 24096, 24097, 24099, 24100, 24107, 24114, 24116, 24120, 24125, 24130, 24131, 24138, 24154, 34148, (2) 24115, 34002, 34011, 34020, 34028, 34029, 34030, 34033, 34034, 34038, 34043, 34065, 34066, 34071, 34073, 34074, 34077, 34082, 34086, 34095, 34106, 34109, 34113, 34115, 34118, 34120, 34123, 34135, 34138, 34139, 34140, 34144, 34159, 34160, 34165, 34200, 34201, 34203, 34207, 34213, 34220, 34225, 34230, 34231, 34232, 34234, 34237, 34239, 34241, 34242, 34247, 34253, 34254, 34255, 34256, 34258, 34263, 34264, 34268, 34270, 34271, 34272, 34273, 34274, 34275, 34276, 34281, 34282, 34285, 34286, 34287, 34288, 34289, 34290, 34291, 34293, 34294, 34295, 34297, 34298, 34299, 34300, 34301, 34305, 34306, 34309, 34311, 73012, (3) 38002, 38003, 38007, 38012, 38017, 38018, 38021, 38022, 38033, 38034, 38043, 38044, 38047, 38054, 38055, 38056, 38057, 38059, 38060, 38061, 38064, 38069, 38072, 38079, 38085, 38086, 38087, 38090, 38091, 38093, 38095, 38098, 38099, 38103, 38104, 38108, 38112, 38115, 38130, 38133, 38137, 38139, 38140, 38141, 38143, 38144, 38145, 38150, 38151, 38152, 38157, 38158, 38160, 38167, 38169, 38171, 38173, 38174, 38181, 38185, 38186, 38187, 38188, 38190, 38191, 38194, 38196, 38200, 38206, 38207, 38209, 38210, 38212, 38215, 38216, 38220, 38223, 38227, 38228, 38229, 38231, 38232, 38233, 38236, 38238, 38240, 38241, 38245, 38246, 38247, 38248, 38249, 38251, 38252, 38253, 38254, 38255, 38256, 38257, 38258, 38260, 38261, 38262, 38264, 38265, 38266, 38267, 38268, 38269, 38270, 38271, 38272, 38273, 38274, 38276, 38278, 38279, 38280, 38282, 38283, 38284, 38287, 38288, 38291, 38292, 38293, 38295, 38296, 38297, 38298, 38299, 38302, 38303, 38304, 38307, 38312, 38314, 38315, 38318, 38319, 38320, 38321, 38322, 38323, 38324, 38326, 38327, 38328, 38329, 38330, 38331, 38332, 38335, 38337, 38338, 38339, 38340, 38341, 38343, 38344, 38345, 38346, 38347, 38349, 38351, 38355, 38358, 38359, 38360, 38361, 38363, 38367, 38368, 38370, 38372, 38401, 38402, 38405, 38407, 38408, 38410, 38412, 38414, 38415, 38420, 38421, 38423, 38424, 38425, 38426, 38427, 38429, 38430, 38431, 38432, 38433, 38436, 38437, 38438, 38439, 38441, 38442, 38443, 38445, 38447, 38448, 38450, 38453, 38454, 38456, 38457, 38458, 382320, (4) 73003, 73005, 73006, 73008, 73009, 73010, (5) 74001, 74005, 74007, 74009, 74010, 74020, 74021, 74026, 74027, 74029, 74030, 74032, 74033, 74034, 74035, 74037, 74039, 74041, 74044, 74046, 74047, 74048, 74052;
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2200-2025
Related Recalls
Philips SmartPath to dStream for 3.0T with MR Elastography (MRE). 1. Model Number (REF): 782145.
Philips North America
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Philips SmartPath to Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 782118. 2. Model Number (REF): 782144. 3. Model Number (REF): 782163.
Philips North America
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.