Pinnacle3 Software Version 9.0, 9.2, 9.4 and 9.6, Model Numbers 453560446041, 459800091001, 459800220161, 459800232931, 459800235871, 459800338451. UPDATED: Pinnacle3 Software Version 8.0h, 8.0k, 8.0m, 8.0n.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    453560446041, 459800091001, 459800220161, 459800232931, 459800235871, 459800338451 UPDATED: 453560446051 [A] 459800096361 [Al 453560446061 IA] 459800096541 [A] 453560446091 [A] 459800096741 [A] 453560446101 [A] 459800096921 [A] 453560446111 [A] 459800097111 [A] 453560446131 [A] 459800097291 [A] 453560446141 [A] 459800097471 [A] 453560446161 [A] 459800097651 [B] 453560446181 [A] 459800097831 [B] 453560446201 [A] 459800098171 [B] 459800089221 [A] 459800098351 [A] 459800091451 [A] 459800109261 [A] 459800138101 [BJ 459800189431 [A) 459800220131 [B] 459800239171 [A] 459800233271 [8] 459800233281 [A] 459800233291 [B] 459800233301 [A] 459800233311 [8] 459800233321 [B] 459800233331 [B] 459800233341 [A] 459800233351 [A] 459800233361 [A] 459800234101 [A] 459800234111 [A] 459800234121 [A] 459800318391 [Al 459800405751 [A] 459800417961 [A] 459800249651 [8] 459800249721 [A] 459800249851 [A] 459800249931 [A] 459800250001 [Al 459800250081 [A] 459800250151 [A] 459800250221 [A] 459800250291 [A] 459800250361 [A] 459800250431 [A] 459800250501 [A] 459800250571 [A] 459800338961 [B] 459800340181 [A] 459800340201 [A] 459800341171 [A] 459800341191 [A] 459800341481 [A] 459800341501 [A] 459800345001 [A] 459800345021 [A] 459800345041 [A] 459800345061 [A] 459800345081 [A] 459800345101 [A] 459800454251 [A] 459800455341 [A] 459800493591 [A] 459800494011 [A] 459800554831 [A] 459800554881 [A] 453560435761 453560434881 453560454791 453560461331 453560461341
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Philips Electronics North America Corporation
Reason for Recall:
Philips, Pinnacle Radiation Treatment Planning System version 8.0h, 8.0k, 8.0m, 8.0n, 9 0, 9 2 9 4, 9 6 is being recalled because the dose may be inconsistent with the density of a density-overridden ROl.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Pinnacle3 Software Version 9.0, 9.2, 9.4 and 9.6, Model Numbers 453560446041, 459800091001, 459800220161, 459800232931, 459800235871, 459800338451. UPDATED: Pinnacle3 Software Version 8.0h, 8.0k, 8.0m, 8.0n.

Product Codes/Lot Numbers:

453560446041, 459800091001, 459800220161, 459800232931, 459800235871, 459800338451 UPDATED: 453560446051 [A] 459800096361 [Al 453560446061 IA] 459800096541 [A] 453560446091 [A] 459800096741 [A] 453560446101 [A] 459800096921 [A] 453560446111 [A] 459800097111 [A] 453560446131 [A] 459800097291 [A] 453560446141 [A] 459800097471 [A] 453560446161 [A] 459800097651 [B] 453560446181 [A] 459800097831 [B] 453560446201 [A] 459800098171 [B] 459800089221 [A] 459800098351 [A] 459800091451 [A] 459800109261 [A] 459800138101 [BJ 459800189431 [A) 459800220131 [B] 459800239171 [A] 459800233271 [8] 459800233281 [A] 459800233291 [B] 459800233301 [A] 459800233311 [8] 459800233321 [B] 459800233331 [B] 459800233341 [A] 459800233351 [A] 459800233361 [A] 459800234101 [A] 459800234111 [A] 459800234121 [A] 459800318391 [Al 459800405751 [A] 459800417961 [A] 459800249651 [8] 459800249721 [A] 459800249851 [A] 459800249931 [A] 459800250001 [Al 459800250081 [A] 459800250151 [A] 459800250221 [A] 459800250291 [A] 459800250361 [A] 459800250431 [A] 459800250501 [A] 459800250571 [A] 459800338961 [B] 459800340181 [A] 459800340201 [A] 459800341171 [A] 459800341191 [A] 459800341481 [A] 459800341501 [A] 459800345001 [A] 459800345021 [A] 459800345041 [A] 459800345061 [A] 459800345081 [A] 459800345101 [A] 459800454251 [A] 459800455341 [A] 459800493591 [A] 459800494011 [A] 459800554831 [A] 459800554881 [A] 453560435761 453560434881 453560454791 453560461331 453560461341

Distribution:

Distributed in: US, DC, PR

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2200-2015

Related Recalls

There are 3 issues: 1) Under certain conditions, real time waveforms may not be accurately plotted or displayed. 2) The oxygen saturation (SpO2) numeric value displayed on the device may freeze. 3) In auto cycle mode, if the NIBP communication is lost, the pump would not cycle and the non-invasive blood pressure (NIBP) numeric value displayed on the device would not update.

Nov 15, 2018 Diagnostic Equipment View Details →

A problem has been detected in the Philips IntelliVue MX4O that, if it were to occur, could affect the performance of the equipment. The volume of the MX4O speaker in these devices may be diminished or not audible above 4,500 feet when operating in Monitor Mode. These devices may intermittently exhibit Speaker Malfunct INOP messages after the Power On Self-Test. The issue is only apparent when the device is being used in Monitor Mode at an altitude of greater than 4,500 feet.

Oct 10, 2018 Diagnostic Equipment Nationwide View Details →

Automated external defibrillators may not fully meet IPx5 water ingress specification. The device may fail to function should water intrusion occur.

Oct 10, 2018 Implants & Prosthetics Nationwide View Details →