Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Monitor UDI-DI: 00858481006017. Lot (C20206): 231220
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Capnia Inc
Reason for Recall:
Sampling set nasal cannula used with End-Tidal Carbon Monoxide (ETCO) monitor may have residual carbon monoxide and volatile organic compounds present due to insufficient aeration of the cannula's carbon filter prior to sealing, which may lead to higher than expected ETOCc scores during clinical evaluations.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor

Product Codes/Lot Numbers:

Monitor UDI-DI: 00858481006017. Lot (C20206): 231220

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2198-2025