REF 8L92-21, Abbott Activated Alanine Aminotransferase, for use with Architect, IVD, CE, UDI: (01) 00380740017866
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot Numbers / Expiration Date 15014UN20 20OCT2021 09153UN21 09MAR2022 *Not distributed within the US.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Abbott Laboratories
- Reason for Recall:
- There is a potential for >10% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in patient results with samples containing bilirubin concentrations greater that 31 mg/dL. This could lead to incorrect patient results
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
REF 8L92-21, Abbott Activated Alanine Aminotransferase, for use with Architect, IVD, CE, UDI: (01) 00380740017866
Product Codes/Lot Numbers:
Lot Numbers / Expiration Date 15014UN20 20OCT2021 09153UN21 09MAR2022 *Not distributed within the US.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2186-2021
Related Recalls
Software issue for hq analyzer results in system not visibly applying appropriate flagging to results when saturation is present which may result in incorrect results.
Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-01 Software Version: All software versions Product Description: CELL-DYN Ruby Component: No
Abbott Laboratories
When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.
Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-10 Software Version: All software versions Product Description: CELL-DYN Ruby Component: No
Abbott Laboratories
When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.