i-Stat BNP cartridges Abbott Point of Care Inc. The i-Stat BNP test is an in vitro diagnostic test for the quantitative measurement of B-type Natriuretic Peptide (BNP) in whole blood or plasma samples using EDTA as the anticoagulant. BNP measurements can be used as an aid in the diagnosis and assessment of the severity of congestive heart failure.
Class I - DangerousWhat Should You Do?
- Check if you have this product: List number 06F30-01, 06F30-02, 03P93-25, 600-9010-25 All lots
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Abbott Point Of Care Inc.
- Reason for Recall:
- The frequency of suppressed results for i-Stat cTnl, BNP and CK-MP cartridges is affected by atmospheric pressure.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
i-Stat BNP cartridges Abbott Point of Care Inc. The i-Stat BNP test is an in vitro diagnostic test for the quantitative measurement of B-type Natriuretic Peptide (BNP) in whole blood or plasma samples using EDTA as the anticoagulant. BNP measurements can be used as an aid in the diagnosis and assessment of the severity of congestive heart failure.
Product Codes/Lot Numbers:
List number 06F30-01, 06F30-02, 03P93-25, 600-9010-25 All lots
Distribution:
Distributed in: AZ, CO
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2170-2012
Related Recalls
i-STAT EG6+ cartridge. List Number: 03P77-25.
Abbott Point Of Care
Lack of a 510(k) premarket clearance for the i-STAT CG8+, EG7+ and EG6+ cartridges to reflect updated sample type for potassium and ionized calcium and measurement ranges for pH and PO2.
i-STAT CG4+ cartridge (white). List Number: 03P85-25.
Abbott Point Of Care
Lack of a 510(k) premarket clearance for the i-STAT CG4+ cartridge to reflect updated sample type for lactate and measurement ranges for pH and PO2.
i-STAT CG8+ cartridge. List Number: 03P88-25. 510(k) numbers: K894914, K940918, K223710, K230275, K230285, K230300
Abbott Point Of Care
Lack of a 510(k) premarket clearance for the i-STAT CG8+, EG7+ and EG6+ cartridges to reflect updated sample type for potassium and ionized calcium and measurement ranges for pH and PO2.