SHBG IMMULITE 2000/IMMULITE 2000 XPi For the in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers-For the quantitative measurement of SHBG in serum, as aid in the differential diagnosis of hirsutism.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot Numbers: 1) 357, 2) 358, 3) 360, 4) 361 & 5) 363 Expiration Dates: 1) 2017-01-30, 2) 2017-03-31, 3) 2017-04-30, 4) 2017-05-31 & 5) 2017-06-30
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Siemens Healthcare Diagnostics, Inc.
- Reason for Recall:
- Certain lots of Sex Hormone Binding Globulin (SHBG) do not meet the thirty (30) day open vial adjustor stability claim when reconstituted and stored at 2-8¿C, as published in the Instructions for Use (IFU). Adjustors stored at 2-8C after 14 days may show a bias in patient samples and quality control falling outside of published ranges. Lookback not recommended.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
SHBG IMMULITE 2000/IMMULITE 2000 XPi For the in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers-For the quantitative measurement of SHBG in serum, as aid in the differential diagnosis of hirsutism.
Product Codes/Lot Numbers:
Lot Numbers: 1) 357, 2) 358, 3) 360, 4) 361 & 5) 363 Expiration Dates: 1) 2017-01-30, 2) 2017-03-31, 3) 2017-04-30, 4) 2017-05-31 & 5) 2017-06-30
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2166-2017
Related Recalls
Food Panel 3 Specific Allergen; Siemens Material Number (SMN): 10385891; Catalog Number: FP3L4;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
3gAllergy Specific IgE Universal Kit; Kit Component: 3gAllergy Specific IgE Adjustor Antibody (L2UNS1), 3gAllergy Specific IgE Control Antibody (L2UNS2); Siemens Material Number (SMN): 10380875; Catalog Number: L2KUN6;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
IMMULITE 2000 Oak Mix Specific Allergen; Siemens Material Number (SMN): 10386105; Catalog Number: L2KUN6;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.