The ABL software version 6.12 for the ABL800 and ABL700 analyzer. The analyzers are intended for the in Vitro testing of: -Samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cCa2+, cCl-, cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF) -Samples of expired air for the parameters pO2 and pCO2 -Pleura samples for the pH parameter. The ABL software version 6.12 for the ABL800 and ABL700 analyzer. The analyzers are intended for the in Vitro testing of: -Samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cCa2+, cCl-, cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF) -Samples of expired air for the parameters pO2 and pCO2 -Pleura samples for the pH parameter.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    ABL725: 441R0086N008, 441R0292N010, and 441R0520N009U ABL 735: 441R0050N001, 441R0231N0009, 441R0234N0010, 441R0292N0009, 441R0303N010, 441R0340N004, 441R0343N010, and 441R0522N008
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Radiometer America Inc
Reason for Recall:
A customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. If the instrument transmits results to a HIS system and the connection is broken the results are queued for later transmission. If the ABL and HIS system are configured with HL7 2.5, any results in the queue will be mixed up when the instrument transmits t
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

The ABL software version 6.12 for the ABL800 and ABL700 analyzer. The analyzers are intended for the in Vitro testing of: -Samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cCa2+, cCl-, cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF) -Samples of expired air for the parameters pO2 and pCO2 -Pleura samples for the pH parameter. The ABL software version 6.12 for the ABL800 and ABL700 analyzer. The analyzers are intended for the in Vitro testing of: -Samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cCa2+, cCl-, cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF) -Samples of expired air for the parameters pO2 and pCO2 -Pleura samples for the pH parameter.

Product Codes/Lot Numbers:

ABL725: 441R0086N008, 441R0292N010, and 441R0520N009U ABL 735: 441R0050N001, 441R0231N0009, 441R0234N0010, 441R0292N0009, 441R0303N010, 441R0340N004, 441R0343N010, and 441R0522N008

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2163-2012

Related Recalls

Class I - Dangerous

Software Security; The action is being initiated because of software security vulnerabilities with the firm's analyzer operating system, which may cause the device to shutdown or reboot resulting in delayed medical treatment.

Jul 10, 2019 Other Medical Devices Nationwide View Details →

From three incidents in Denmark and Sweden, comparison studies on patient blood samples carried out for quality assurance demonstrated that creatinine measurements as determined on analyzers could be underestimated with up to 28% as compared with measurements from COBAS. The underestimated creatinine occurs specifically, when measurements on Autocheck 6+ level 1 (also known as S7835 Autocheck module and abbreviated AC6+L1) are less than or equal to 230 micro-meter of a specified insert interval for AC6+L1 with insert limits 210 290 micro-meter. Ultimately, a lower read-out on patient samples is observed Customers are concerned about underestimation of creatinine results in patients, which are to undergo radiocontrast x-ray investigations. If these patients are not healthy, the exposure of radiocontrast agents may induce kidney insufficiency (contrast-induced nephropathy). Hazardous situation: Underestimation of the concentration of creatinine.

Jul 2, 2018 Other Medical Devices Nationwide View Details →