The ABL software version 6.12 for the ABL800 analyzer. The analyzers are intended for the in Vitro testing of: -Samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cCa2+, cCl, cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF) -Samples of expired air for the parameters pO2 and pCO2 -Pleura samples for the pH parameter. The ABL software version 6.12 for the ABL800 and ABL700 analyzer. The analyzers are intended for the in Vitro testing of: -Samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cCa2+, cCl-, cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF) -Samples of expired air for the parameters pO2 and pCO2 -Pleura samples for the pH parameter.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    ABL800 BASIC: 754R0406N012U ABL800 BASIC with AutoCheck: 754R0742N006 754R0467N002 754R0466N008 754R0467N001 754R0433N011 754R0434N001 754R0437N005 754R0437N006 754R0437N007 754R0436N004 754R0437N001 754R0436N003 754R0436N005 754R0436N006 754R0437N003 754R0437N002 754R0432N003 754R0432N004 754R0432N005 754R0436N002 754R0467N003 754R0276N001 754R1306N002 754R1306N003 754R1308N003 754R1308N005 754R1308N006 754R1309N001 754R1309N005 754R1309N006 754R1309N007 754R1310N001 754R1310N002 754R1310N003 754R1310N004 754R1310N005 754R1310N006 754R1310N007 754R1311N001 754R1311N002 754R1311N003 754R1311N004 754R1311N006 754R1312N001 754R1312N003 754R1312N005 754R0294N008 754R1191N002 754R1191N003 754R1191N004 754R1191N005 754R0400N010 754R0401N001 754R0400N004 754R0400N005 754R0387N012 754R0886N003 754R0886N004 754R0306N001 754R0306N002 754r1098n004 754r1098n006 754R1080N004 754R1080N006 754R1233N004 754R1233N006 754R1211N004 754R1213N001 754R1189N001 754R1189N003 754R1122N002 754R1133N004 754R0999N002 754R0525N011 754R0526N002 754R0526N003 754R0526N004 754R0415N005 754R0415N006 754R0407N006 754R1362N003 754R1363N001 754R1311N005 754R1311N007 754R1239N002 754R1242N008 754R1019N003 754R1019N008 754R0993N009 754R0996N003 754R0967N005 754R0967N007 754R0938N008 754R0939N001 754R0921N004 754R0921N006 754r0809n008 754r0810n001 754R0788N005 754R0787N006 754R0673N005 754R0673N006 754R0673N007 754R0673N008 754R0547N001 754R0547N002 754R0547N003 754R0453N010 754R0454N001 754R0453N011 754R0516N005 754R0516N006 754R0516N007 754R0456N008 754R0457N001 754R0457N002 754R0310N012 754R0259N005 754R0418N010 754R0375N003 754R0375N006 754R0375N004 754R0353N011 754R0353N012 754R1365N003 754R1365N004 754R1365N005 754R1365N007 754R1365N008 754R1233N005 754R1233N007 754R1235N001 754R1235N002 754R1235N004 754R1169N002 754R1169N006 754R1170N001 754R1170N005 754R1170N006 754R1112N006 754R1112N007 754R1116N001 754R1116N002 754R1116N003 754R1050N003 754R1053N004 754R1053N005 754R1053N006 754R1053N007 754R0988N008 754R0988N009 754R0989N002 754R0989N003 754R0989N004 754R0989N005 754R0993N008 754R0921N003 754R0921N005 754R0921N009 754R0922N001 754R0923N003 754R0878N008 754R0878N009 754R0879N001 754R0879N002 754R0879N003 754R0710N001 754R0710N002 754R0710N003 754R0710N004 754R0710N005 754R0703N002 754R0703N005 754R0704N001 754R0704N002 754R0704N003 754R0701N002 754R0701N005 754R0702N003 754R0702N004 754R0702N005 754R0642N002 754R0642N004 754R0642N006 754R0642N007 754R0587N004 754R0586N010 754R0586N008 754R0586N007 754R0498N009 754R0499N009 754R0500N001 754R1242N007 754R0336N004 754R0982N005 754R0458N008 754R0257N006 754R1252N010 754R0454N003 754R0454N004 754R0254N001 754R1166N003 754R1166N004 754R0837N004 754r0569n001 754R0569N002 754R0523N006 754R0401N009 754R0390N002 754R1352N005 754R1353N001 754R1334N002 754R1330N004 754R1319N003 754R1312N002 754R1312N004 754R1284N001 754R1277N007 754R1278N001 754R1258N006 754R1211N005 754R1227N009 754R1205N009 754R1153N003 754R1153N004 754r0375n002 754r0436n011 754R0436N012 754R0439N002 754R1112N005 754R1112N004 754R1099N008 754R1100N002 754R1100N004 754R1100N005 754R0999N005 754R0999N009 754R0971N009 754R0973N008 754R0973N009 754R0951N008 754R0951N009 754R0955N007 754R0947N004 754R0948N001 754R0948N006 754R0948N008 754R0921N001 754R0925N005 754R0854N008 754R0850N009 754R0856N002 754R0850N007 754R0850N008 754R0809N002 754R0809N003 754r0784n006 754R0741N002 754R0741N001 754r0726n009 754R0683N003 754R0683N004 754R0353N010 754R0348N011 754R0348N009 754R0325N012 754R0332N004 754R0332N005 754R0316N005 754R0319N009 754R0320N005 754R0298N002 754R0298N003 754R0303N002 754R0378N007 754R0378N008 754R0378N009 754R0378N010 754R0378N011 754R0378N012 754R0379N001 754R0379N002 754R0484N004 754R0484N005 754R1357N004 754R1357N007 754R1320N005 754R1327N001 754R0521N007 754R0521N008 754R0521N009 754R0521N010 754R0521N011 754R0521N012 754R0522N001 754R0522N002 754R0539N008 754R0539N009 754R0539N010 754R0539N011 754R0539N012 754R0541N008 754R0541N010 754R0541N012 754R1305N006 754R1241N003 754R1242N010 754R1233N008 754R1233N011 754R1213N004 754R1213N006 754R1190N008 754R1191N001 754R1169N004 754R1166N005 754R1162N001 754R1166N001 754R1133N003 754R1133N005 754R1112N001 754R1112N002 754R1082N008 754R1098N002 754R1080N001 754R1080N003 754R1013N007 754R1013N008 754R1013N009 754R0905N006 754R0905N007 754R0905N008 754r0899n007 754R0876N001 754R0876N002 754R0877N002 754R0850N004 754r0841n006 754r0842n001 754r0809n004 754R0781N005 754R0775N006 754R0775N008 754R0760N003 754R0760N007 754R0761N001 754R0761N003 754R0670N001 754R0923N005 754R1259N003 754R1268N004 754R0861N008 754R1034N003 754R1205N010 754R0923N007 754R0923N008 754R1170N003 754R0948N003 754R1025N001 754R1330N006 754R1330N005 754R1327N005 754R1327N002 754R0654N004 754R0642N003 754R1169N005 754R1050N001 754R1050N002 754R1050N005 754R1050N004 754R0999N008 754R0999N006 754R0967N009 754R0948N004 754R0809N007 754R0636N006 754R0603N006 754R0603N007 754R0607N003 754R0574N008 754R0586N002 754R0532N010 754R0532N011 754R0533N011 754R0533N012 754R0477N002 754R0710N006 754R0525N008 754R0602N009 754R0603N001 754R0603N002 754R1169N003 754R1170N002 754R1170N004 754R1170N007 754R1170N008 754R0947N009 754R0948N009 754R0820N001 754R0823N006 754R0731N004 754R0625N003 754R0625N002 754R0521N005 754R0521N006 754R1213N007 754R1166N002 754R0985N002 754R1361N005 754R1365N001 754R1365N002 754R1365N006 754R1327N004 754R1330N001 754R1330N003 754R1330N007 754R1305N005 754R1305N007 754R1306N001 754R1308N007 754R1278N008 754R1281N007 754R1283N003 754R1283N005 754R1283N006 754R1023N009 754R1024N002 754R1024N003 754R1019N004 754R1019N005 754R1019N006 754R1019N007 754R1019N009 754R1024N001 754R1024N006 754R1024N007 754R1024N008 754R1024N009 754R1025N006 754R0968N001 754R0968N002 754R0968N006 754R0968N009 754R0902N009 754R0904N004 754R0904N005 754R0904N007 754R0905N001 754R0880N008 754R0880N009 754R0881N001 754R0878N002 754R0878N005 754R0878N006 754R0856N003 754R0856N005 754R0856N006 754R0856N007 754R0856N008 754R0856N001 754R0856N004 754R0854N006 754R0854N007 754R0828N008 754R0829N001 754R0829N002 754R0829N003 754R0824N008 754R0787N007 754R0788N007 754R0788N008 754R0791N008 754R0788N006 754R0774N004 754R0774N007 754R0714N006 754R0722N009 754R0713N001 754R0711N002 754R0711N006 754R0684N003 754R0654N005 754R0655N006 754R0655N007 754R0643N004 754R0643N002 754R0643N003 754R0643N001 754R0635N001 754R0635N002 754R0635N003 754R0636N007 754R0636N008 754R0637N004 754R0603N003 754R0565N002 754R0565N005 754R0565N006 754R0565N007 754R0566N003 754R0531N012 754R0532N004 754R0532N005 754R0532N008 754R0532N009 754R0532N012 754R0537N010 754R0537N011 754R0537N012 754R0538N001 754R0538N002 754R0538N003 754R0538N006 754R0498N007 754R0499N004 754R0499N006 754R0500N003 754R0500N007 754R0501N002 754R0499N007 754R0448N004 754R0457N004 754R0448N003 754R0407N008 754R0407N009 754R0407N010 754R0407N011 754R0407N012 754R0408N001 754R0402N011 754R0623N001 754R0966N005 754R0565N003 754R0566N005 754R0937N001 754R0938N004 754R0938N005 754R0938N007 754R1270N001 754R1213N002 754R1213N003 754R0569N003 754R0330N004 754R0342N003 754R1171N004 754R1073N005 754R0999N007 754R0999N004 754R0999N003 754R0880N006 754R0854N009 754R0837N001 754R0725N002 754R0586N005 754R0564N008 754R0552N004 754R0537N008 754R0535N002 754R0525N007 754R0499N002 754R0500N009 754R0428N011 754R0431N007 754R0426N002 754R0422N003 754R0418N008 754R0387N011 754R0342N001 754R0298N001 754R0287N001 754R1082N006 754R1239N004 754R1241N002 754R1012N001 754R1000N001 754R0984N009 754R0985N003 754R0967N008 754R0947N007 754R0947N005 754R0934N007 754R0923N006 754r0902n008 754r0866n003 754R0860N004 754R0860N007 754r0859n008 754r0860n002 754r0860n003 754R0836N006 754R0836N007 754R0381N005 754R0864N006 754R0411N012 754R0310N005 754R0246N001 754R0564N009 754R0565N001 754R0545N002 754R0498N008 754R0525N009 754R0304N012 754R0431N008 754R1234N002 754R1234N001 754R1152N005 754R1151N005 754R1082N007 754R1034N002 754R0966N004 754R0711N003 754R0647N008 754R0613N001 754R0353N009 754R0774N003 754R0647N004 754R0565N008 754R0335N001 754R0243N007 754R0804N005 754R0621N005 754R0621N008 754R0621N010 754R0971N008 754R0927N005 754R0927N006 754R0927N007 754R0927N008 754R0927N009 754R0437N004 754R0387N005 754R0387N006 754R0543N001 754R0966N008 754R0966N009 754R0332N006 754R0406N012 754R0668N007 754R0668N006 754R0760N008 754R0682N006 754R0482N008 754R1357N001 754R1362N001 754R1362N002 754R1356N001 754R1356N006 754R1356N007 754R1356N008 754R1357N002 754R1357N003 754R1353N009 754R1353N010 754R1353N011 754R1354N003 754R1346N003 754R1346N001 754R1346N002 754R1339N003 754R1339N004 754R1339N005 754R1339N006 754R1334N003 754R1334N004 754R1334N005 754R1333N007 754R1333N004 754R1333N005 754R1333N006 754R1319N004 754R1320N006 754R1321N001 754R1320N004 754R1319N005 754R1320N001 754R1320N002 754R1320N003 754R1314N008 754R1308N004 754R1298N003 754R1297N005 754R1298N002 754R1298N005 754R1293N001 754R1297N004 754R1298N001 754R1277N001 754R1284N002 754R1277N004 754R1279N001 754R1279N002 754R1283N007 754R1283N008 754R1273N007 754R1277N005 754R1277N006 754R1278N007 754R1272N002 754R1273N001 754R1273N003 754R1274N001 754R1274N002 754R1274N003 754R1256N004 754R1256N005 754R1242N005 754R1242N006 754R1235N008 754R1235N009 754R1238N004 754R1239N001 754R1239N003 754R1233N010 754R1235N003 754R1235N006 754R1235N007 754R1227N008 754R1227N010 754R1227N011 754R1228N001 754R1220N008 754R1220N011 754R1220N009 754R1220N010 754R1221N001 754R1213N005 754R1205N007 754R1206N001 754R1199N008 754R1186N007 754R1188N006 754R1188N007 754R1188N008 754R1189N002 754R1171N003 754R1170N009 754R1169N007 754R1171N002 754R1153N006 754R1142N001 754R1151N004 754R1146N007 754R1147N001 754R1141N002 754R1141N005 754R1141N004 754R1141N006 754R1120N004 754R1120N006 754R1120N007 754R1121N001 754R1117N001 754R1119N002 754R1119N004 754R1119N005 754R1120N005 754R1119N001 754R1118N006 754R1108N007 754R1105N002 754R1100N003 754R1101N002 754R1101N003 754R1104N007 754R1105N001 754R1100N001 754R1101N001 754R1082N002 754R1082N004 754R1082N005 754R1082N003 754R1080N007 754R1080N008 754R1082N001 754R1080N002 754R1080N005 754R1068N008 754R1068N007 754R1068N004 754R1068N006 754R1068N003 754R1068N005 754R1068N002 754R1061N002 754R1036N002 754R1036N003 754R1036N004 754R1036N008 754R1036N007 754R1036N005 754R1036N006 754R1035N008 754R1036N001 754R1034N004 754R1033N008 754R1034N001 754R1028N001 754R1025N003 754R1025N004 754R1023N008 754R1025N002 754R1025N008 754R1025N007 754R1026N002 754R1026N001 754R1027N001 754R1025N009 754R1013N006 754R1014N001 754R1014N002 754R1014N003 754R1014N004 754R1012N002 754R0995N008 754R0995N009 754R0996N001 754R0996N002 754R0989N001 754R0982N006 754R0982N007 754R0982N004 754R0982N002 754R0981N008 754R0981N009 754R0982N001 754R0977N001 754R0976N009 754R0964N008 754R0964N009 754R0965N002 754R0965N006 754R0964N007 754R0965N001 754R0965N003 754R0965N004 754R0965N007 754R0965N005 754R0965N009 754R0966N002 754R0966N003 754R0965N008 754R0966N001 754R0947N008 754R0952N001 754R0949N006 754R0949N003 754R0951N007 754R0949N004 754R0948N002 754R0946N008 754R0946N009 754R0947N006 754R0938N009 754R0939N002 754R0938N006 754R0934N003 754R0934N004 754R0934N005 754R0934N006 754R0934N009 754R0921N007 754R0926N001 754R0926N002 754R0925N007 754R0925N008 754R0920N008 754R0921N008 754R0916N007 754R0916N008 754R0916N009 754R0916N004 754R0916N005 754R0916N006 754R0911N009 754R0912N001 754R0912N005 754R0911N006 754R0911N007 754R0911N008 754R0911N005 754R0906N009 754R0907N001 754R0907N002 754R0905N004 754R0905N009 754R0899N005 754R0899N006 754R0891N009 754R0892N002 754R0886N006 754R0886N005 754R0880N003 754R0881N002 754R0881N006 754R0881N007 754R0881N008 754R0881N009 754R0880N004 754R0880N005 754R0880N007 754R0878N007 754R0879N009 754R0880N002 754R0878N004 754R0880N001 754R0877N008 754R0877N003 754R0877N004 754R0875N008 754R0875N009 754R0875N003 754R0870N001 754R0875N005 754R0875N004 754R0875N006 754R0875N007 754R0869N002 754R0869N005 754R0869N003 754R0869N006 754R0869N009 754R0869N001 754R0859N001 754R0866N001 754R0866N002 754R0864N007 754R0864N008 754R0864N009 754R0860N008 754R0850N006 754R0837N003 754R0841N005 754R0839N005 754R0839N003 754R0837N002 754R0839N004 754R0828N002 754R0828N003 754R0828N005 754R0827N004 754R0828N001 754R0822N004 754R0822N005 754R0824N006 754R0824N007 754R0823N004 754R0824N004 754R0819N004 754R0823N001 754R0823N003 754R0823N005 754R0823N007 754R0823N008 754R0824N001 754R0824N002 754R0824N003 754R0824N005 754R0821N005 754R0821N007 754R0822N001 754R0822N002 754R0822N003 754R0822N006 754R0822N007 754R0822N008 754R0823N002 754R0821N008 754R0821N004 754R0821N006 754R0820N008 754R0821N002 754R0820N003 754R0820N004 754R0820N007 754R0821N001 754R0820N002 754R0820N005 754R0820N006 754R0821N003 754R0815N007 754R0815N008 754R0809N001 754R0811N001 754R0807N001 754R0809N005 754R0809N006 754R0810N002 754R0792N004 754R0792N002 754R0792N003 754R0781N002 754R0784N007 754R0778N002 754R0781N003 754R0774N006 754R0776N008 754R0775N007 754R0776N001 754R0774N008 754R0775N001 754R0775N005 754R0774N005 754R0774N002 754R0770N001 754R0761N002 754R0769N007 754R0769N008 754R0761N004 754R0760N004 754R0760N006 754R0760N005 754R0751N002 754R0751N003 754R0741N004 754R0742N004 754R0741N003 754R0742N005 754R0738N005 754R0738N006 754R0732N001 754R0732N002 754R0732N003 754R0729N009 754R0729N010 754R0726N004 754R0726N010 754R0712N002 754R0712N003 754R0712N004 754R0711N005 754R0712N001 754R0712N005 754R0711N004 754R0684N004 754R0684N005 754R0684N006 754R0682N007 754R0682N008 754R0683N001 754R0683N002 754R0682N001 754R0682N002 754R0682N003 754R0670N002 754R0668N008 754R0663N006 754R0664N005 754R0660N008 754R0663N005 754R0663N008 754R0664N001 754R0664N002 754R0664N003 754R0664N004 754R0664N006 754R0660N005 754R0660N006 754R0661N004 754R0661N005 754R0661N006 754R0654N003 754R0655N001 754R0648N001 754R0647N005 754R0647N007 754R0647N002 754R0647N003 754R0637N001 754R0642N001 754R0637N003 754R0637N005 754R0637N002 754R0625N004 754R0621N009 754R0624N010 754R0625N001 754R0623N002 754R0623N003 754R0623N004 754R0613N003 754R0613N004 754R0613N007 754R0607N001 754R0602N010 754R0603N004 754R0603N005 754R0596N002 754R0596N001 754R0596N003 754R0592N005 754R0586N009 754R0592N002 754R0592N006 754R0588N005 754R0587N005 754R0592N001 754R0587N002 754R0592N003 754R0588N006 754R0587N001 754R0587N003 754R0586N006 754R0574N006 754R0574N007 754R0574N009 754R0569N009 754R0569N006 754R0569N007 754R0569N004 754R0569N005 754R0569N008 754R0544N009 754R0554N001 754R0552N003 754R0553N009 754R0552N006 754R0552N001 754R0552N002 754R0552N005 754R0553N008 754R0548N010 754R0551N002 754R0551N003 754R0548N008 754R0548N009 754R0548N011 754R0541N009 754R0541N011 754R0538N004 754R0539N007 754R0535N001 754R0535N003 754R0532N006 754R0532N007 754R0525N012 754R0525N010 754R0523N005 754R0523N002 754R0523N003 754R0523N004 754R0516N004 754R0516N008 754R0515N006 754R0516N003 754R0515N005 754R0515N004 754R0503N002 754R0503N001 754R0501N001 754R0502N008 754R0500N002 754R0501N003 754R0500N005 754R0500N006 754R0500N008 754R0499N008 754R0500N004 754R0498N006 754R0498N005 754R0482N007 754R0484N003 754R0484N002 754R0482N002 754R0484N001 754R0467N004 754R0474N001 754R0474N003 754R0457N003 754R0459N008 754R0459N005 754R0459N004 754R0459N006 754R0459N007 754R0453N008 754R0453N012 754R0453N009 754R0453N007 754R0448N012 754R0447N004 754R0446N012 754R0447N002 754R0447N003 754R0447N001 754R0447N005 754R0447N007 754R0446N011 754R0445N006 754R0439N006 754R0439N009 754R0439N008 754R0439N010 754R0439N011 754R0439N005 754R0439N004 754R0439N003 754R0439N007 754R0436N001 754R0431N009 754R0431N011 754R0431N012 754R0432N001 754R0428N010 754R0428N009 754R0426N001 754R0426N003 754R0422N004 754R0419N009 754R0419N010 754R0418N009 754R0418N011 754R0418N012 754R0414N002 754R0413N008 754R0413N012 754R0414N001 754R0413N011 754R0413N010 754R0413N009 754R0413N005 754R0413N006 754R0413N007 754R0411N010 754R0411N005 754R0411N006 754R0411N007 754R0411N008 754R0406N005 754R0406N002 754R0406N003 754R0406N004 754R0405N007 754R0405N008 754R0405N009 754R0405N010 754R0398N003 754R0390N003 754R0397N012 754R0398N002 754R0397N011 754R0398N001 754R0390N004 754R0387N008 754R0387N007 754R0387N004 754R0387N009 754R0387N003 754R0384N011 754R0385N001 754R0385N003 754R0384N012 754R0385N002 754R0378N003 754R0375N005 754R0373N001 754R0375N001 754R0361N001 754R0360N011 754R0360N007 754R0353N007 754R0353N008 754r0348N007 754R0348N010 754R0348N008 754R0345N011 754R0344N011 754R0345N012 754R0338N003 754R0338N002 754R0330N005 754R0332N003 754R0330N001 754R0330N006 754R0335N002 754R0330N002 754R0332N002 754R0329N011 754R0330N003 754R0330N007 754R0329N012 754R0316N003 754R0316N004 754R0305N012 754R0310N002 754R0310N003 754R0310N004 754R0310N001 754R0310N007 754R0310N009 754R0310N010 754R0306N003 754R0306N005 754R0303N001 754R0304N011 754R0280N003 754R0306N004 754R0304N010 754R0290N005 754R0297N001 754R0297N002 754R0294N010 754R0294N011 754R0290N003 754R0290N004 754R0290N006 754R0290N002 754R0287N004 754R0287N003 754R0287N002 754R0280N005 754R0280N004 754R0276N002 754R0276N003 754R0259N004 754R0258N009 754R0258N010 754R0258N004 754R0258N001 754R0258N002 754R0257N005 754R0256N010 754R0247N008 754R0247N007 754R0246N005 754R0246N009 754R0247N001 754R0247N002 754R0243N012 754R0244N004 754R0244N008 754R0244N012 754R0245N004 754R0245N009 754R0240N001 754R0240N006 754R0240N010 754R0236N003 754R0238N008 754R0239N004 754R0239N011 754R0238N001 754R0235N011 754R0235N004 754R0236N009 754R0234N009 754R0232N002 754R0232N003 754R0232N009 754R0232N010 754R0233N004 754R0228N001 754R0227N006 754R0226N004 754R0226N011 754R0224N007 754R0225N002 754R0225N009 754R0222N004 754R0222N011 754R0223N006 754R0223N012 754R0220N011 754R0221N010 754R0220N001 754R0220N003 754R0221N005 754R0776N002 754R0574N005 754R0428N012 754R0526N001 754R1283n004 754R1233N009 754R0960N001 754R0781N001 754R0682N004 754R0682N005 754R0663N007 754R0660N007 754R0661N003 754R0661N007 754R0647N001 754R0625N005 754R0613N005 754R0592N004 754R0537N009 754R0560N006 754R0560N007 754R0560N008 754R0547N004 754R0534N001 754R0534N003 754R0499N003 754R0474N002 754R0445N007 754R0432N002 754R1141N007 754r0732n004 754R0711N001 754R0684N007 754R0696N002 754R0696N003 754R0696N004 754R1339N001 754R1339N002 754R1338N005 754R1334N001 754R1327N003 754R1186N006 754R1186N008 754R1171N005 754R1121N002 754R1121N003 754R1108N006 754R1073N004 754R1069N001 754R1054N006 754R0999N001 754R0881N003 754r0081n003 754r0729n008 754r0731n001 754R0661N008 754R0642N005 754R0642N008 754R0613N006 754R0548N012 754R0544N008 754R0545N001 754R0515N003 754R0521N003 754R0521N004 754R0499N001 754R0431N010 754R0327N011 754R0327N012 754R0390N001 754R0381N003 754R0360N008 754R0330N008 754R0348N006 754R1272N001 754R0877N006 754R0804N006 754R0621N006 754R0554N002 754R0538N005 754R0477N001 754R0387N010 754R0342N002 754R0324N012 754R0212N002 754R1364N005 754R1364N006 754R1207N001 754R1153N001 754R1109N006 754R1112N003 754R1061N001 754R1025N005 754R0998N009 754R0982N003 754R0949N005 754R0948N005 754R0850N005 754R1270N003 754R1258N007 754R1259N001 754R1259N002 754R0960N002 754R0949N001 754R0949N002 754R0942N001 754R0942N005 754R0942N006 754R0945N008 754R0934N008 754R0937N002 754R0937N005 754R0936N008 754R0937N004 754R0937N006 754R0936N007 754R0936N009 754R0937N003 754R0936N006 754R0921N002 754R0904N008 754R0904N009 754R0905N002 754R0905N003 754R0905N005 754R1119N003 754R1119N007 754R1364N007 754R0684N002 754R1152N004 754R1153N005 754R1153N007 754R1363N002 754R1363N003 754R1363N004 754R1363N005 754R1363N006 754R1363N007 754R1363N008 754R1364N001 754R1364N002 754R1364N003 754R1364N004 754R1345N004 754R1345N005 754R1345N006 754R1345N007 754R1341N005 754R1342N002 754R1342N004 754R1342N005 754R1343N001 754R1343N002 754R1343N006 754R1343N007 754R1344N001 754R1344N005 754R1344N006 754R1344N007 754R1345N001 754R1345N002 754R1345N003 754R1248N008 754R1251N004 754R1251N005 754R1251N006 754R1251N007 754R1251N008 754R1251N009 754R1251N012 754R1252N002 754R1252N003 754R1252N004 754R1252N005 754R1252N006 754R1252N008 754R1252N009 754R1252N011 754R1252N012 754R1242N009 754R1242N011 754R1243N001 754R1243N002 754R1243N003 754R1243N004 754R1243N005 754R1247N011 754R1247N012 754R1248N001 754R1248N002 754R1248N003 754R1248N004 754R1248N005 754R1248N006 754R1248N007 754R1194N001 754R1194N002 754R1194N003 754R1194N004 754R1194N005 754R1195N001 754R1197N001 754R1197N004 754R1197N005 754R1197N006 754R1197N008 754R1198N001 754R1198N002 754R1199N001 754R1199N007 754R1191N006 754R1191N007 754R1191N008 754R1193N006 754R1193N007 754R1193N008 754R1194N006 754R1194N007 754R1194N008 754R1195N002 754R1195N003 754R1195N004 754R1195N005 754R1195N006 754R1196N005 754R1196N006 754R1196N007 754R1196N008 754R1197N002 754R1197N003 754R1135N002 754R1135N005 754R1135N006 754R1140N002 754R1140N003 754R1140N004 754R1140N005 754R1140N006 754R1140N007 754R1141N003 754R1116N004 754R1116N005 754R1116N006 754R1116N007 754R1117N002 754R1117N003 754R1117N004 754R1117N005 754R1117N006 754R1117N007 754R1118N001 754R1118N002 754R1118N003 754R1118N004 754R1118N005 754R1053N008 754R1054N001 754R1054N002 754R1054N003 754R1054N004 754R1054N005 754R1054N007 754R1054N008 754R1055N001 754R1055N002 754R1055N003 754R1055N004 754R1055N005 754R1055N006 754R1055N007 754R1055N008 754R1056N002 754R1056N003 754R1056N004 754R1056N005 754R1056N006 754R1056N007 754R1056N008 754R1057N001 754R1057N002 754R1057N003 754R1057N004 754R1057N005 754R1057N006 754R1057N007 754R1057N008 754R1058N002 754R1058N003 754R1058N004 754R1058N005 754R1058N006 754R1059N004 754R1059N005 754R1059N006 754R1059N007 754R1019N002 754R1023N001 754R1023N002 754R1023N003 754R1023N004 754R1023N005 754R1023N006 754R1023N007 754R1024N004 754R1024N005 754R0966N007 754R0967N004 754R0967N006 754R0968N003 754R0968N004 754R0968N005 754R0968N007 754R0968N008 754R0969N001 754R0969N002 754R0969N007 754R0969N008 754R0969N009 754R0970N001 754R0970N002 754R0970N003 754R0970N004 754R0971N005 754R0971N006 754R0971N007 754R0904N006 754R0925N002 754R0925N003 754R0925N004 754R0925N006 754R0869N004 754R0923N004 754R0923N009 754R0924N001 754R0924N002 754R0924N003 754R0924N004 754R0924N005 754R0924N006 754R0924N007 754R0924N008 754R0924N009 754R0923N002 754R0886N007 754R0886N008 754R0886N009 754R0890N003 754R0890N004 754R0890N005 754R0890N006 754R0890N007 754R0890N008 754R0890N009 754R0891N001 754R0891N002 754R0891N003 754R0891N004 754R0891N005 754R0891N006 754R0891N007 754R0891N008 754R0892N001 754R0892N003 754R0825N002 754R0825N003 754R0825N004 754R0825N005 754R0825N006 754R0827N001 754R0827N002 754R0827N003 754R0827N005 754R0827N006 754R0827N007 754R0827N008 754R0828N004 754R0828N006 754R0828N007 754R0787N001 754R0787N002 754R0787N003 754R0787N004 754R0787N005 754R0787N008 754R0788N001 754R0788N002 754R0788N003 754R0788N004 754R0725N001 754R0725N003 754R0725N004 754R0725N005 754R0725N006 754R0725N007 754R0725N008 754R0725N009 754R0725N010 754R0726N001 754R0726N002 754R0726N003 754R0726N005 754R0726N006 754R0726N008 754R0713N002 754R0713N003 754R0713N004 754R0713N005 754R0713N006 754R0714N001 754R0714N002 754R0714N003 754R0714N004 754R0714N005 754R0668N005 754R0674N001 754R0674N002 754R0674N003 754R0674N004 754R0678N005 754R0678N006 754R0678N007 754R0678N008 754R0679N001 754R0648N002 754R0648N003 754R0651N001 754R0651N002 754R0651N004 754R0651N005 754R0651N006 754R0651N003 754R0647N006 754R0651N007 754R0651N008 754R0652N001 754R0652N002 754R0652N003 754R0652N004 754R0652N005 754R0652N006 754R0652N007 754R0653N004 754R0652N008 754R0653N001 754R0653N003 754R0653N005 754R0653N002 754R0653N006 754R0653N008 754R0654N001 754R0654N002 754R0653N007 754R0654N006 754R0654N007 754R0654N008 754R0655N002 754R0655N004 754R0655N003 754R0522N003 754R0522N004 754R0522N005 754R0522N006 754R0522N007 754R0522N008 754R0522N009 754R0522N010 754R0522N011 754R0522N012 754R0523N001 754R0479N005 754R0479N006 754R0479N007 754R0479N008 754R0480N001 754R0480N003 754R0480N004 754R0480N005 754R0480N006 754R0480N007 754R0480N008 754R0481N003 754R0481N004 754R0481N005 754R0481N006 754R0481N007 754R0481N008 754R0480N002 754R0482N001 754R0482N003 754R0482N004 754R0482N005 754R0482N006 754R0447N006 754R0447N009 754R0447N010 754R0447N011 754R0447N012 754R0448N007 754R0448N008 754R0448N009 754R0448N010 754R0448N011 754R0406N008 754R0408N002 754R0408N003 754R0408N004 754R0408N005 754R0408N006 754R0408N007 754R0408N009 754R0408N010 754R0411N004 754R0405N006 754R0406N006 754R0406N009 754R0406N010 754R0406N011 754R0373N002 754R0371N001 754R0371N002 754R0371N003 754R0373N003 754R0373N004 754R0342N004 754R0344N001 754R0344N002 754R0344N010 754R0338N004 754R0339N001 754R0339N002 754R0339N003 754R0327N009 754R0327N010 754R0328N001 754R0328N002 754R0324N010 754R0324N011 754R0327N007 754R0327N008 754R0311N001 754R0311N002 754R0316N001 754R0316N002 754R0309N012 754R0310N006 754R0310N008 754R0310N011 754R0290N007 754R0294N009 754R1232N001 754R1235N005 754R1061N003 754R0625N006 754R0625N007 754R0454N002 754R0411N009 754R0411N011 754R0378N004 754R1341N003 754R1012N003 754R0877N005 754R0877N007 754R0877N009 754R0878N001 754R0878N003 754R0856N009 754R0859N009 754R0860N001 754R0860N005 754R0860N006 754R0841N007 754R0842N002 754R0842N003 754R0842N004 754R0842N005 754R0837N005 754R0837N006 754R0837N007 754R0839N001 754R0839N002 754R0544N007 754R0499N005 754R0447N008 754R0390N005 754R0407N007 754R0739N003 754R0739N002 754R0739N001 754R0738N004 754R0738N003 754R0607N002 754R0560N009 754R0535N004 754R0534N002 754R0459N003 754R0459N002 754R0459N001 754R0458N007 754R0458N006 754R0458N005 754R0381N002 754R0381N004 754R0381N001 754R0371N004 754R0621N007 754R0289N009 754R1098N003 754R1098N001 754R0920N009 754R0948N007 754R0415N007 754R1153N002 754R1364N008 754R1211N006 754R1251N010 754R1252N007 754R1253N001 754R1253N002 754R1253N003 754R1253N004 754R1253N005 754R1253N008 754R1253N011 754R1133N006 754R1133N007 754R1134N001 754R1134N002 754R1134N003 754R1134N004 754R1134N007 754R1135N001 754R1135N003 754R1135N004 754R1098N005 754R1014N005 754R0966N006 754R0792N007 754R0797N001 754R0792N006 754R0712N006 754R0655N005 754R0613N002 754R0565N004 754R0565N009 754R0566N001 754R0566N002 754R0566N004 754R0419N011 754R0422N001 754R0419N012 754R0422N002 And 754R0415N008 ABL800 FLEX: 754R0560N001 754R0058N003 754R0448N001 754R0024N010 754R0122N004 754R0206N003 754R0544N006 754R0547N008 754R0547N009 754R0560N001 754R0578N003 754R0579N003 754R0617N007 754R0706N006 754R0766N002 754R0768N007 754R0953N004 754R1011N008 754R0102N002U 754R0253N009U 754R0112N001U 754R0175N003U 754R0103n004u 754R0006N010U 754R0100N006U 754R0175N009u 754R0094N001u 754R0220N012U 754R0100N007U 754R0026N003U 754R0300N004U 754R0403N022U 754R0156N003U 754R0128N006U 754R0127N002U 754R0131N004U 754R0416N007U 754R0425N012U 754R0017N010U 754R0076N004U 754R0193N005U 754R0128N005U 754R0128N004U 754R0028N007U 754R0323N011U 754R0093N006u 754R0331N004U 754R0385N012U 754R0384N004U 754R0458N008U 754R0193N007U 754R0184N007U 754R0008N003U 754R0006N008U 754R0425N011U 754R0300N009U 754R0116N001U 754R0075N008U 754R0099N005U And 754R0255N007U ABL800 FLEX with AutoCheck: 754R0445N008 754R0445N009 754R0445N010 754R0445N011 754R0445N012 754R0781N006 754R0792N001 754R0792N005 754R0792N008 754R0915N009 754R0925N001 754R0588N010 754R1366N004 754R1343N005 754R1343N004 754R1337N004 754R1273N004 754R1256N003 754R1236N002 754R1192N004 754R1158N001 754R1158N002 754R1155N001 754R1146N005 754R1083N006 754R1083N007 754R0715N008 754R1014N006 754R1014N007 754R0237N012 754R1134N006 754R0349N002 754r0357n003 754R0757N003 754R0757N004 754R0761N005 754R0761N006 754R0761N007 754R0762N001 754R0762N002 754R0724N006 754R0722N007 754R0722N008 754R0722N010 754R0724N001 754R0724N002 754R0724N003 754R0724N007 754R0472N006 754R0715N001 754R0715N002 754R0715N003 754R0715N004 754R0715N005 754R0715N006 754R0715N007 754R0399N010 754R0397N002 754R0397N004 754R0397N008 754R0399N001 754R0399N002 754R0399N003 754R0399N004 754R0399N005 754R0399N006 754R0399N007 754R0399N008 754R0399N009 754R0394N002 754R0394N003 754R0390N006 754R0397N005 754R0397N006 754R0397N003 754R0390N011 754R0348N012 754R0342N007 754R0339N007 754R0342N005 754R0342N006 754R0298N010 754R1237N005 754R1258N005 754R1192N005 754R1166N007 754R1166N006 754R1211N009 754R1192N003 754R1161N007 754R0826N001 754R0588N007 754R0391N001 754R0406N001 754R0405N012 754R0403N037 754R0360N009 754R0588N008 754R0588N009 754R0942N002 754R0942N004 754R0942N003 754R0616N001 754R0616N007 754R0616N004 754R0616N005 754R0616N006 754R0321N011 754R0321N010 754R1366N001 754R1366N003 754R1349N003 754R1347N007 754R1348N001 754R1348N002 754R1348N004 754R1348N005 754R1348N003 754R1331N001 754R1335N001 754R1335N002 754R1336N002 754R1336N003 754R1331N006 754R1331N007 754R1318N005 754R1316N004 754R1318N004 754R1306N007 754R1303N001 754R1293N003 754R1293N004 754R1286N007 754R1293N002 754R1286N001 754R1286N002 754R1286N003 754R1286N004 754R1287N003 754R1287N006 754R1287N007 754R1287N008 754R1288N001 754R1287N004 754R1287N005 754R1284N004 754R1286N005 754R1286N006 754R1286N008 754R1287N002 754R1273N005 754R1258N003 754R1258N004 754R1256N001 754R1256N002 754R1258N001 754R1258N002 754R1227N007 754R1236N009 754R1237N002 754R1237N011 754R1238N001 754R1227N005 754R1214N006 754R1214N002 754R1214N004 754R1219N008 754R1219N009 754R1220N001 754R1214N003 754R1214N005 754R1219N007 754R1212N002 754R1214N001 754R1205N008 754R1203N001 754R1203N002 754R1200N003 754R1200N007 754R1201N002 754R1198N003 754R1200N002 754R1201N003 754R1200N005 754R1201N001 754R1180N001 754R1184N001 754R1184N002 754R1184N003 754R1179N006 754R1175N005 754R1179N001 754R1179N002 754R1179N004 754R1175N002 754R1175N003 754R1175N004 754R1175N006 754R1175N007 754R1176N001 754R1176N002 754R1143N005 754R1155N002 754R1145N001 754R1145N007 754R1145N002 754R1145N003 754R1145N004 754R1145N005 754R1146N001 754R1144N007 754R1123N005 754R1122N003 754R1109N003 754R1109N004 754R1119N006 754R1120N001 754R1109N001 754R1107N001 754R1109N002 754R1083N008 754R1101N004 754R1101N005 754R1101N006 754R1076N005 754R1076N007 754R1083N005 754R1073N003 754R1073N001 754R1073N002 754R1062N002 754R1062N001 754R1062N006 754R1062N007 754R1062N003 754R1062N004 754R1062N005 754R1062N008 754R1060N005 754R1060N007 754R1058N007 75
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Radiometer America Inc
Reason for Recall:
A customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. If the instrument transmits results to a HIS system and the connection is broken the results are queued for later transmission. If the ABL and HIS system are configured with HL7 2.5, any results in the queue will be mixed up when the instrument transmits t
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

The ABL software version 6.12 for the ABL800 analyzer. The analyzers are intended for the in Vitro testing of: -Samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cCa2+, cCl, cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF) -Samples of expired air for the parameters pO2 and pCO2 -Pleura samples for the pH parameter. The ABL software version 6.12 for the ABL800 and ABL700 analyzer. The analyzers are intended for the in Vitro testing of: -Samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cCa2+, cCl-, cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF) -Samples of expired air for the parameters pO2 and pCO2 -Pleura samples for the pH parameter.

Product Codes/Lot Numbers:

ABL800 BASIC: 754R0406N012U ABL800 BASIC with AutoCheck: 754R0742N006 754R0467N002 754R0466N008 754R0467N001 754R0433N011 754R0434N001 754R0437N005 754R0437N006 754R0437N007 754R0436N004 754R0437N001 754R0436N003 754R0436N005 754R0436N006 754R0437N003 754R0437N002 754R0432N003 754R0432N004 754R0432N005 754R0436N002 754R0467N003 754R0276N001 754R1306N002 754R1306N003 754R1308N003 754R1308N005 754R1308N006 754R1309N001 754R1309N005 754R1309N006 754R1309N007 754R1310N001 754R1310N002 754R1310N003 754R1310N004 754R1310N005 754R1310N006 754R1310N007 754R1311N001 754R1311N002 754R1311N003 754R1311N004 754R1311N006 754R1312N001 754R1312N003 754R1312N005 754R0294N008 754R1191N002 754R1191N003 754R1191N004 754R1191N005 754R0400N010 754R0401N001 754R0400N004 754R0400N005 754R0387N012 754R0886N003 754R0886N004 754R0306N001 754R0306N002 754r1098n004 754r1098n006 754R1080N004 754R1080N006 754R1233N004 754R1233N006 754R1211N004 754R1213N001 754R1189N001 754R1189N003 754R1122N002 754R1133N004 754R0999N002 754R0525N011 754R0526N002 754R0526N003 754R0526N004 754R0415N005 754R0415N006 754R0407N006 754R1362N003 754R1363N001 754R1311N005 754R1311N007 754R1239N002 754R1242N008 754R1019N003 754R1019N008 754R0993N009 754R0996N003 754R0967N005 754R0967N007 754R0938N008 754R0939N001 754R0921N004 754R0921N006 754r0809n008 754r0810n001 754R0788N005 754R0787N006 754R0673N005 754R0673N006 754R0673N007 754R0673N008 754R0547N001 754R0547N002 754R0547N003 754R0453N010 754R0454N001 754R0453N011 754R0516N005 754R0516N006 754R0516N007 754R0456N008 754R0457N001 754R0457N002 754R0310N012 754R0259N005 754R0418N010 754R0375N003 754R0375N006 754R0375N004 754R0353N011 754R0353N012 754R1365N003 754R1365N004 754R1365N005 754R1365N007 754R1365N008 754R1233N005 754R1233N007 754R1235N001 754R1235N002 754R1235N004 754R1169N002 754R1169N006 754R1170N001 754R1170N005 754R1170N006 754R1112N006 754R1112N007 754R1116N001 754R1116N002 754R1116N003 754R1050N003 754R1053N004 754R1053N005 754R1053N006 754R1053N007 754R0988N008 754R0988N009 754R0989N002 754R0989N003 754R0989N004 754R0989N005 754R0993N008 754R0921N003 754R0921N005 754R0921N009 754R0922N001 754R0923N003 754R0878N008 754R0878N009 754R0879N001 754R0879N002 754R0879N003 754R0710N001 754R0710N002 754R0710N003 754R0710N004 754R0710N005 754R0703N002 754R0703N005 754R0704N001 754R0704N002 754R0704N003 754R0701N002 754R0701N005 754R0702N003 754R0702N004 754R0702N005 754R0642N002 754R0642N004 754R0642N006 754R0642N007 754R0587N004 754R0586N010 754R0586N008 754R0586N007 754R0498N009 754R0499N009 754R0500N001 754R1242N007 754R0336N004 754R0982N005 754R0458N008 754R0257N006 754R1252N010 754R0454N003 754R0454N004 754R0254N001 754R1166N003 754R1166N004 754R0837N004 754r0569n001 754R0569N002 754R0523N006 754R0401N009 754R0390N002 754R1352N005 754R1353N001 754R1334N002 754R1330N004 754R1319N003 754R1312N002 754R1312N004 754R1284N001 754R1277N007 754R1278N001 754R1258N006 754R1211N005 754R1227N009 754R1205N009 754R1153N003 754R1153N004 754r0375n002 754r0436n011 754R0436N012 754R0439N002 754R1112N005 754R1112N004 754R1099N008 754R1100N002 754R1100N004 754R1100N005 754R0999N005 754R0999N009 754R0971N009 754R0973N008 754R0973N009 754R0951N008 754R0951N009 754R0955N007 754R0947N004 754R0948N001 754R0948N006 754R0948N008 754R0921N001 754R0925N005 754R0854N008 754R0850N009 754R0856N002 754R0850N007 754R0850N008 754R0809N002 754R0809N003 754r0784n006 754R0741N002 754R0741N001 754r0726n009 754R0683N003 754R0683N004 754R0353N010 754R0348N011 754R0348N009 754R0325N012 754R0332N004 754R0332N005 754R0316N005 754R0319N009 754R0320N005 754R0298N002 754R0298N003 754R0303N002 754R0378N007 754R0378N008 754R0378N009 754R0378N010 754R0378N011 754R0378N012 754R0379N001 754R0379N002 754R0484N004 754R0484N005 754R1357N004 754R1357N007 754R1320N005 754R1327N001 754R0521N007 754R0521N008 754R0521N009 754R0521N010 754R0521N011 754R0521N012 754R0522N001 754R0522N002 754R0539N008 754R0539N009 754R0539N010 754R0539N011 754R0539N012 754R0541N008 754R0541N010 754R0541N012 754R1305N006 754R1241N003 754R1242N010 754R1233N008 754R1233N011 754R1213N004 754R1213N006 754R1190N008 754R1191N001 754R1169N004 754R1166N005 754R1162N001 754R1166N001 754R1133N003 754R1133N005 754R1112N001 754R1112N002 754R1082N008 754R1098N002 754R1080N001 754R1080N003 754R1013N007 754R1013N008 754R1013N009 754R0905N006 754R0905N007 754R0905N008 754r0899n007 754R0876N001 754R0876N002 754R0877N002 754R0850N004 754r0841n006 754r0842n001 754r0809n004 754R0781N005 754R0775N006 754R0775N008 754R0760N003 754R0760N007 754R0761N001 754R0761N003 754R0670N001 754R0923N005 754R1259N003 754R1268N004 754R0861N008 754R1034N003 754R1205N010 754R0923N007 754R0923N008 754R1170N003 754R0948N003 754R1025N001 754R1330N006 754R1330N005 754R1327N005 754R1327N002 754R0654N004 754R0642N003 754R1169N005 754R1050N001 754R1050N002 754R1050N005 754R1050N004 754R0999N008 754R0999N006 754R0967N009 754R0948N004 754R0809N007 754R0636N006 754R0603N006 754R0603N007 754R0607N003 754R0574N008 754R0586N002 754R0532N010 754R0532N011 754R0533N011 754R0533N012 754R0477N002 754R0710N006 754R0525N008 754R0602N009 754R0603N001 754R0603N002 754R1169N003 754R1170N002 754R1170N004 754R1170N007 754R1170N008 754R0947N009 754R0948N009 754R0820N001 754R0823N006 754R0731N004 754R0625N003 754R0625N002 754R0521N005 754R0521N006 754R1213N007 754R1166N002 754R0985N002 754R1361N005 754R1365N001 754R1365N002 754R1365N006 754R1327N004 754R1330N001 754R1330N003 754R1330N007 754R1305N005 754R1305N007 754R1306N001 754R1308N007 754R1278N008 754R1281N007 754R1283N003 754R1283N005 754R1283N006 754R1023N009 754R1024N002 754R1024N003 754R1019N004 754R1019N005 754R1019N006 754R1019N007 754R1019N009 754R1024N001 754R1024N006 754R1024N007 754R1024N008 754R1024N009 754R1025N006 754R0968N001 754R0968N002 754R0968N006 754R0968N009 754R0902N009 754R0904N004 754R0904N005 754R0904N007 754R0905N001 754R0880N008 754R0880N009 754R0881N001 754R0878N002 754R0878N005 754R0878N006 754R0856N003 754R0856N005 754R0856N006 754R0856N007 754R0856N008 754R0856N001 754R0856N004 754R0854N006 754R0854N007 754R0828N008 754R0829N001 754R0829N002 754R0829N003 754R0824N008 754R0787N007 754R0788N007 754R0788N008 754R0791N008 754R0788N006 754R0774N004 754R0774N007 754R0714N006 754R0722N009 754R0713N001 754R0711N002 754R0711N006 754R0684N003 754R0654N005 754R0655N006 754R0655N007 754R0643N004 754R0643N002 754R0643N003 754R0643N001 754R0635N001 754R0635N002 754R0635N003 754R0636N007 754R0636N008 754R0637N004 754R0603N003 754R0565N002 754R0565N005 754R0565N006 754R0565N007 754R0566N003 754R0531N012 754R0532N004 754R0532N005 754R0532N008 754R0532N009 754R0532N012 754R0537N010 754R0537N011 754R0537N012 754R0538N001 754R0538N002 754R0538N003 754R0538N006 754R0498N007 754R0499N004 754R0499N006 754R0500N003 754R0500N007 754R0501N002 754R0499N007 754R0448N004 754R0457N004 754R0448N003 754R0407N008 754R0407N009 754R0407N010 754R0407N011 754R0407N012 754R0408N001 754R0402N011 754R0623N001 754R0966N005 754R0565N003 754R0566N005 754R0937N001 754R0938N004 754R0938N005 754R0938N007 754R1270N001 754R1213N002 754R1213N003 754R0569N003 754R0330N004 754R0342N003 754R1171N004 754R1073N005 754R0999N007 754R0999N004 754R0999N003 754R0880N006 754R0854N009 754R0837N001 754R0725N002 754R0586N005 754R0564N008 754R0552N004 754R0537N008 754R0535N002 754R0525N007 754R0499N002 754R0500N009 754R0428N011 754R0431N007 754R0426N002 754R0422N003 754R0418N008 754R0387N011 754R0342N001 754R0298N001 754R0287N001 754R1082N006 754R1239N004 754R1241N002 754R1012N001 754R1000N001 754R0984N009 754R0985N003 754R0967N008 754R0947N007 754R0947N005 754R0934N007 754R0923N006 754r0902n008 754r0866n003 754R0860N004 754R0860N007 754r0859n008 754r0860n002 754r0860n003 754R0836N006 754R0836N007 754R0381N005 754R0864N006 754R0411N012 754R0310N005 754R0246N001 754R0564N009 754R0565N001 754R0545N002 754R0498N008 754R0525N009 754R0304N012 754R0431N008 754R1234N002 754R1234N001 754R1152N005 754R1151N005 754R1082N007 754R1034N002 754R0966N004 754R0711N003 754R0647N008 754R0613N001 754R0353N009 754R0774N003 754R0647N004 754R0565N008 754R0335N001 754R0243N007 754R0804N005 754R0621N005 754R0621N008 754R0621N010 754R0971N008 754R0927N005 754R0927N006 754R0927N007 754R0927N008 754R0927N009 754R0437N004 754R0387N005 754R0387N006 754R0543N001 754R0966N008 754R0966N009 754R0332N006 754R0406N012 754R0668N007 754R0668N006 754R0760N008 754R0682N006 754R0482N008 754R1357N001 754R1362N001 754R1362N002 754R1356N001 754R1356N006 754R1356N007 754R1356N008 754R1357N002 754R1357N003 754R1353N009 754R1353N010 754R1353N011 754R1354N003 754R1346N003 754R1346N001 754R1346N002 754R1339N003 754R1339N004 754R1339N005 754R1339N006 754R1334N003 754R1334N004 754R1334N005 754R1333N007 754R1333N004 754R1333N005 754R1333N006 754R1319N004 754R1320N006 754R1321N001 754R1320N004 754R1319N005 754R1320N001 754R1320N002 754R1320N003 754R1314N008 754R1308N004 754R1298N003 754R1297N005 754R1298N002 754R1298N005 754R1293N001 754R1297N004 754R1298N001 754R1277N001 754R1284N002 754R1277N004 754R1279N001 754R1279N002 754R1283N007 754R1283N008 754R1273N007 754R1277N005 754R1277N006 754R1278N007 754R1272N002 754R1273N001 754R1273N003 754R1274N001 754R1274N002 754R1274N003 754R1256N004 754R1256N005 754R1242N005 754R1242N006 754R1235N008 754R1235N009 754R1238N004 754R1239N001 754R1239N003 754R1233N010 754R1235N003 754R1235N006 754R1235N007 754R1227N008 754R1227N010 754R1227N011 754R1228N001 754R1220N008 754R1220N011 754R1220N009 754R1220N010 754R1221N001 754R1213N005 754R1205N007 754R1206N001 754R1199N008 754R1186N007 754R1188N006 754R1188N007 754R1188N008 754R1189N002 754R1171N003 754R1170N009 754R1169N007 754R1171N002 754R1153N006 754R1142N001 754R1151N004 754R1146N007 754R1147N001 754R1141N002 754R1141N005 754R1141N004 754R1141N006 754R1120N004 754R1120N006 754R1120N007 754R1121N001 754R1117N001 754R1119N002 754R1119N004 754R1119N005 754R1120N005 754R1119N001 754R1118N006 754R1108N007 754R1105N002 754R1100N003 754R1101N002 754R1101N003 754R1104N007 754R1105N001 754R1100N001 754R1101N001 754R1082N002 754R1082N004 754R1082N005 754R1082N003 754R1080N007 754R1080N008 754R1082N001 754R1080N002 754R1080N005 754R1068N008 754R1068N007 754R1068N004 754R1068N006 754R1068N003 754R1068N005 754R1068N002 754R1061N002 754R1036N002 754R1036N003 754R1036N004 754R1036N008 754R1036N007 754R1036N005 754R1036N006 754R1035N008 754R1036N001 754R1034N004 754R1033N008 754R1034N001 754R1028N001 754R1025N003 754R1025N004 754R1023N008 754R1025N002 754R1025N008 754R1025N007 754R1026N002 754R1026N001 754R1027N001 754R1025N009 754R1013N006 754R1014N001 754R1014N002 754R1014N003 754R1014N004 754R1012N002 754R0995N008 754R0995N009 754R0996N001 754R0996N002 754R0989N001 754R0982N006 754R0982N007 754R0982N004 754R0982N002 754R0981N008 754R0981N009 754R0982N001 754R0977N001 754R0976N009 754R0964N008 754R0964N009 754R0965N002 754R0965N006 754R0964N007 754R0965N001 754R0965N003 754R0965N004 754R0965N007 754R0965N005 754R0965N009 754R0966N002 754R0966N003 754R0965N008 754R0966N001 754R0947N008 754R0952N001 754R0949N006 754R0949N003 754R0951N007 754R0949N004 754R0948N002 754R0946N008 754R0946N009 754R0947N006 754R0938N009 754R0939N002 754R0938N006 754R0934N003 754R0934N004 754R0934N005 754R0934N006 754R0934N009 754R0921N007 754R0926N001 754R0926N002 754R0925N007 754R0925N008 754R0920N008 754R0921N008 754R0916N007 754R0916N008 754R0916N009 754R0916N004 754R0916N005 754R0916N006 754R0911N009 754R0912N001 754R0912N005 754R0911N006 754R0911N007 754R0911N008 754R0911N005 754R0906N009 754R0907N001 754R0907N002 754R0905N004 754R0905N009 754R0899N005 754R0899N006 754R0891N009 754R0892N002 754R0886N006 754R0886N005 754R0880N003 754R0881N002 754R0881N006 754R0881N007 754R0881N008 754R0881N009 754R0880N004 754R0880N005 754R0880N007 754R0878N007 754R0879N009 754R0880N002 754R0878N004 754R0880N001 754R0877N008 754R0877N003 754R0877N004 754R0875N008 754R0875N009 754R0875N003 754R0870N001 754R0875N005 754R0875N004 754R0875N006 754R0875N007 754R0869N002 754R0869N005 754R0869N003 754R0869N006 754R0869N009 754R0869N001 754R0859N001 754R0866N001 754R0866N002 754R0864N007 754R0864N008 754R0864N009 754R0860N008 754R0850N006 754R0837N003 754R0841N005 754R0839N005 754R0839N003 754R0837N002 754R0839N004 754R0828N002 754R0828N003 754R0828N005 754R0827N004 754R0828N001 754R0822N004 754R0822N005 754R0824N006 754R0824N007 754R0823N004 754R0824N004 754R0819N004 754R0823N001 754R0823N003 754R0823N005 754R0823N007 754R0823N008 754R0824N001 754R0824N002 754R0824N003 754R0824N005 754R0821N005 754R0821N007 754R0822N001 754R0822N002 754R0822N003 754R0822N006 754R0822N007 754R0822N008 754R0823N002 754R0821N008 754R0821N004 754R0821N006 754R0820N008 754R0821N002 754R0820N003 754R0820N004 754R0820N007 754R0821N001 754R0820N002 754R0820N005 754R0820N006 754R0821N003 754R0815N007 754R0815N008 754R0809N001 754R0811N001 754R0807N001 754R0809N005 754R0809N006 754R0810N002 754R0792N004 754R0792N002 754R0792N003 754R0781N002 754R0784N007 754R0778N002 754R0781N003 754R0774N006 754R0776N008 754R0775N007 754R0776N001 754R0774N008 754R0775N001 754R0775N005 754R0774N005 754R0774N002 754R0770N001 754R0761N002 754R0769N007 754R0769N008 754R0761N004 754R0760N004 754R0760N006 754R0760N005 754R0751N002 754R0751N003 754R0741N004 754R0742N004 754R0741N003 754R0742N005 754R0738N005 754R0738N006 754R0732N001 754R0732N002 754R0732N003 754R0729N009 754R0729N010 754R0726N004 754R0726N010 754R0712N002 754R0712N003 754R0712N004 754R0711N005 754R0712N001 754R0712N005 754R0711N004 754R0684N004 754R0684N005 754R0684N006 754R0682N007 754R0682N008 754R0683N001 754R0683N002 754R0682N001 754R0682N002 754R0682N003 754R0670N002 754R0668N008 754R0663N006 754R0664N005 754R0660N008 754R0663N005 754R0663N008 754R0664N001 754R0664N002 754R0664N003 754R0664N004 754R0664N006 754R0660N005 754R0660N006 754R0661N004 754R0661N005 754R0661N006 754R0654N003 754R0655N001 754R0648N001 754R0647N005 754R0647N007 754R0647N002 754R0647N003 754R0637N001 754R0642N001 754R0637N003 754R0637N005 754R0637N002 754R0625N004 754R0621N009 754R0624N010 754R0625N001 754R0623N002 754R0623N003 754R0623N004 754R0613N003 754R0613N004 754R0613N007 754R0607N001 754R0602N010 754R0603N004 754R0603N005 754R0596N002 754R0596N001 754R0596N003 754R0592N005 754R0586N009 754R0592N002 754R0592N006 754R0588N005 754R0587N005 754R0592N001 754R0587N002 754R0592N003 754R0588N006 754R0587N001 754R0587N003 754R0586N006 754R0574N006 754R0574N007 754R0574N009 754R0569N009 754R0569N006 754R0569N007 754R0569N004 754R0569N005 754R0569N008 754R0544N009 754R0554N001 754R0552N003 754R0553N009 754R0552N006 754R0552N001 754R0552N002 754R0552N005 754R0553N008 754R0548N010 754R0551N002 754R0551N003 754R0548N008 754R0548N009 754R0548N011 754R0541N009 754R0541N011 754R0538N004 754R0539N007 754R0535N001 754R0535N003 754R0532N006 754R0532N007 754R0525N012 754R0525N010 754R0523N005 754R0523N002 754R0523N003 754R0523N004 754R0516N004 754R0516N008 754R0515N006 754R0516N003 754R0515N005 754R0515N004 754R0503N002 754R0503N001 754R0501N001 754R0502N008 754R0500N002 754R0501N003 754R0500N005 754R0500N006 754R0500N008 754R0499N008 754R0500N004 754R0498N006 754R0498N005 754R0482N007 754R0484N003 754R0484N002 754R0482N002 754R0484N001 754R0467N004 754R0474N001 754R0474N003 754R0457N003 754R0459N008 754R0459N005 754R0459N004 754R0459N006 754R0459N007 754R0453N008 754R0453N012 754R0453N009 754R0453N007 754R0448N012 754R0447N004 754R0446N012 754R0447N002 754R0447N003 754R0447N001 754R0447N005 754R0447N007 754R0446N011 754R0445N006 754R0439N006 754R0439N009 754R0439N008 754R0439N010 754R0439N011 754R0439N005 754R0439N004 754R0439N003 754R0439N007 754R0436N001 754R0431N009 754R0431N011 754R0431N012 754R0432N001 754R0428N010 754R0428N009 754R0426N001 754R0426N003 754R0422N004 754R0419N009 754R0419N010 754R0418N009 754R0418N011 754R0418N012 754R0414N002 754R0413N008 754R0413N012 754R0414N001 754R0413N011 754R0413N010 754R0413N009 754R0413N005 754R0413N006 754R0413N007 754R0411N010 754R0411N005 754R0411N006 754R0411N007 754R0411N008 754R0406N005 754R0406N002 754R0406N003 754R0406N004 754R0405N007 754R0405N008 754R0405N009 754R0405N010 754R0398N003 754R0390N003 754R0397N012 754R0398N002 754R0397N011 754R0398N001 754R0390N004 754R0387N008 754R0387N007 754R0387N004 754R0387N009 754R0387N003 754R0384N011 754R0385N001 754R0385N003 754R0384N012 754R0385N002 754R0378N003 754R0375N005 754R0373N001 754R0375N001 754R0361N001 754R0360N011 754R0360N007 754R0353N007 754R0353N008 754r0348N007 754R0348N010 754R0348N008 754R0345N011 754R0344N011 754R0345N012 754R0338N003 754R0338N002 754R0330N005 754R0332N003 754R0330N001 754R0330N006 754R0335N002 754R0330N002 754R0332N002 754R0329N011 754R0330N003 754R0330N007 754R0329N012 754R0316N003 754R0316N004 754R0305N012 754R0310N002 754R0310N003 754R0310N004 754R0310N001 754R0310N007 754R0310N009 754R0310N010 754R0306N003 754R0306N005 754R0303N001 754R0304N011 754R0280N003 754R0306N004 754R0304N010 754R0290N005 754R0297N001 754R0297N002 754R0294N010 754R0294N011 754R0290N003 754R0290N004 754R0290N006 754R0290N002 754R0287N004 754R0287N003 754R0287N002 754R0280N005 754R0280N004 754R0276N002 754R0276N003 754R0259N004 754R0258N009 754R0258N010 754R0258N004 754R0258N001 754R0258N002 754R0257N005 754R0256N010 754R0247N008 754R0247N007 754R0246N005 754R0246N009 754R0247N001 754R0247N002 754R0243N012 754R0244N004 754R0244N008 754R0244N012 754R0245N004 754R0245N009 754R0240N001 754R0240N006 754R0240N010 754R0236N003 754R0238N008 754R0239N004 754R0239N011 754R0238N001 754R0235N011 754R0235N004 754R0236N009 754R0234N009 754R0232N002 754R0232N003 754R0232N009 754R0232N010 754R0233N004 754R0228N001 754R0227N006 754R0226N004 754R0226N011 754R0224N007 754R0225N002 754R0225N009 754R0222N004 754R0222N011 754R0223N006 754R0223N012 754R0220N011 754R0221N010 754R0220N001 754R0220N003 754R0221N005 754R0776N002 754R0574N005 754R0428N012 754R0526N001 754R1283n004 754R1233N009 754R0960N001 754R0781N001 754R0682N004 754R0682N005 754R0663N007 754R0660N007 754R0661N003 754R0661N007 754R0647N001 754R0625N005 754R0613N005 754R0592N004 754R0537N009 754R0560N006 754R0560N007 754R0560N008 754R0547N004 754R0534N001 754R0534N003 754R0499N003 754R0474N002 754R0445N007 754R0432N002 754R1141N007 754r0732n004 754R0711N001 754R0684N007 754R0696N002 754R0696N003 754R0696N004 754R1339N001 754R1339N002 754R1338N005 754R1334N001 754R1327N003 754R1186N006 754R1186N008 754R1171N005 754R1121N002 754R1121N003 754R1108N006 754R1073N004 754R1069N001 754R1054N006 754R0999N001 754R0881N003 754r0081n003 754r0729n008 754r0731n001 754R0661N008 754R0642N005 754R0642N008 754R0613N006 754R0548N012 754R0544N008 754R0545N001 754R0515N003 754R0521N003 754R0521N004 754R0499N001 754R0431N010 754R0327N011 754R0327N012 754R0390N001 754R0381N003 754R0360N008 754R0330N008 754R0348N006 754R1272N001 754R0877N006 754R0804N006 754R0621N006 754R0554N002 754R0538N005 754R0477N001 754R0387N010 754R0342N002 754R0324N012 754R0212N002 754R1364N005 754R1364N006 754R1207N001 754R1153N001 754R1109N006 754R1112N003 754R1061N001 754R1025N005 754R0998N009 754R0982N003 754R0949N005 754R0948N005 754R0850N005 754R1270N003 754R1258N007 754R1259N001 754R1259N002 754R0960N002 754R0949N001 754R0949N002 754R0942N001 754R0942N005 754R0942N006 754R0945N008 754R0934N008 754R0937N002 754R0937N005 754R0936N008 754R0937N004 754R0937N006 754R0936N007 754R0936N009 754R0937N003 754R0936N006 754R0921N002 754R0904N008 754R0904N009 754R0905N002 754R0905N003 754R0905N005 754R1119N003 754R1119N007 754R1364N007 754R0684N002 754R1152N004 754R1153N005 754R1153N007 754R1363N002 754R1363N003 754R1363N004 754R1363N005 754R1363N006 754R1363N007 754R1363N008 754R1364N001 754R1364N002 754R1364N003 754R1364N004 754R1345N004 754R1345N005 754R1345N006 754R1345N007 754R1341N005 754R1342N002 754R1342N004 754R1342N005 754R1343N001 754R1343N002 754R1343N006 754R1343N007 754R1344N001 754R1344N005 754R1344N006 754R1344N007 754R1345N001 754R1345N002 754R1345N003 754R1248N008 754R1251N004 754R1251N005 754R1251N006 754R1251N007 754R1251N008 754R1251N009 754R1251N012 754R1252N002 754R1252N003 754R1252N004 754R1252N005 754R1252N006 754R1252N008 754R1252N009 754R1252N011 754R1252N012 754R1242N009 754R1242N011 754R1243N001 754R1243N002 754R1243N003 754R1243N004 754R1243N005 754R1247N011 754R1247N012 754R1248N001 754R1248N002 754R1248N003 754R1248N004 754R1248N005 754R1248N006 754R1248N007 754R1194N001 754R1194N002 754R1194N003 754R1194N004 754R1194N005 754R1195N001 754R1197N001 754R1197N004 754R1197N005 754R1197N006 754R1197N008 754R1198N001 754R1198N002 754R1199N001 754R1199N007 754R1191N006 754R1191N007 754R1191N008 754R1193N006 754R1193N007 754R1193N008 754R1194N006 754R1194N007 754R1194N008 754R1195N002 754R1195N003 754R1195N004 754R1195N005 754R1195N006 754R1196N005 754R1196N006 754R1196N007 754R1196N008 754R1197N002 754R1197N003 754R1135N002 754R1135N005 754R1135N006 754R1140N002 754R1140N003 754R1140N004 754R1140N005 754R1140N006 754R1140N007 754R1141N003 754R1116N004 754R1116N005 754R1116N006 754R1116N007 754R1117N002 754R1117N003 754R1117N004 754R1117N005 754R1117N006 754R1117N007 754R1118N001 754R1118N002 754R1118N003 754R1118N004 754R1118N005 754R1053N008 754R1054N001 754R1054N002 754R1054N003 754R1054N004 754R1054N005 754R1054N007 754R1054N008 754R1055N001 754R1055N002 754R1055N003 754R1055N004 754R1055N005 754R1055N006 754R1055N007 754R1055N008 754R1056N002 754R1056N003 754R1056N004 754R1056N005 754R1056N006 754R1056N007 754R1056N008 754R1057N001 754R1057N002 754R1057N003 754R1057N004 754R1057N005 754R1057N006 754R1057N007 754R1057N008 754R1058N002 754R1058N003 754R1058N004 754R1058N005 754R1058N006 754R1059N004 754R1059N005 754R1059N006 754R1059N007 754R1019N002 754R1023N001 754R1023N002 754R1023N003 754R1023N004 754R1023N005 754R1023N006 754R1023N007 754R1024N004 754R1024N005 754R0966N007 754R0967N004 754R0967N006 754R0968N003 754R0968N004 754R0968N005 754R0968N007 754R0968N008 754R0969N001 754R0969N002 754R0969N007 754R0969N008 754R0969N009 754R0970N001 754R0970N002 754R0970N003 754R0970N004 754R0971N005 754R0971N006 754R0971N007 754R0904N006 754R0925N002 754R0925N003 754R0925N004 754R0925N006 754R0869N004 754R0923N004 754R0923N009 754R0924N001 754R0924N002 754R0924N003 754R0924N004 754R0924N005 754R0924N006 754R0924N007 754R0924N008 754R0924N009 754R0923N002 754R0886N007 754R0886N008 754R0886N009 754R0890N003 754R0890N004 754R0890N005 754R0890N006 754R0890N007 754R0890N008 754R0890N009 754R0891N001 754R0891N002 754R0891N003 754R0891N004 754R0891N005 754R0891N006 754R0891N007 754R0891N008 754R0892N001 754R0892N003 754R0825N002 754R0825N003 754R0825N004 754R0825N005 754R0825N006 754R0827N001 754R0827N002 754R0827N003 754R0827N005 754R0827N006 754R0827N007 754R0827N008 754R0828N004 754R0828N006 754R0828N007 754R0787N001 754R0787N002 754R0787N003 754R0787N004 754R0787N005 754R0787N008 754R0788N001 754R0788N002 754R0788N003 754R0788N004 754R0725N001 754R0725N003 754R0725N004 754R0725N005 754R0725N006 754R0725N007 754R0725N008 754R0725N009 754R0725N010 754R0726N001 754R0726N002 754R0726N003 754R0726N005 754R0726N006 754R0726N008 754R0713N002 754R0713N003 754R0713N004 754R0713N005 754R0713N006 754R0714N001 754R0714N002 754R0714N003 754R0714N004 754R0714N005 754R0668N005 754R0674N001 754R0674N002 754R0674N003 754R0674N004 754R0678N005 754R0678N006 754R0678N007 754R0678N008 754R0679N001 754R0648N002 754R0648N003 754R0651N001 754R0651N002 754R0651N004 754R0651N005 754R0651N006 754R0651N003 754R0647N006 754R0651N007 754R0651N008 754R0652N001 754R0652N002 754R0652N003 754R0652N004 754R0652N005 754R0652N006 754R0652N007 754R0653N004 754R0652N008 754R0653N001 754R0653N003 754R0653N005 754R0653N002 754R0653N006 754R0653N008 754R0654N001 754R0654N002 754R0653N007 754R0654N006 754R0654N007 754R0654N008 754R0655N002 754R0655N004 754R0655N003 754R0522N003 754R0522N004 754R0522N005 754R0522N006 754R0522N007 754R0522N008 754R0522N009 754R0522N010 754R0522N011 754R0522N012 754R0523N001 754R0479N005 754R0479N006 754R0479N007 754R0479N008 754R0480N001 754R0480N003 754R0480N004 754R0480N005 754R0480N006 754R0480N007 754R0480N008 754R0481N003 754R0481N004 754R0481N005 754R0481N006 754R0481N007 754R0481N008 754R0480N002 754R0482N001 754R0482N003 754R0482N004 754R0482N005 754R0482N006 754R0447N006 754R0447N009 754R0447N010 754R0447N011 754R0447N012 754R0448N007 754R0448N008 754R0448N009 754R0448N010 754R0448N011 754R0406N008 754R0408N002 754R0408N003 754R0408N004 754R0408N005 754R0408N006 754R0408N007 754R0408N009 754R0408N010 754R0411N004 754R0405N006 754R0406N006 754R0406N009 754R0406N010 754R0406N011 754R0373N002 754R0371N001 754R0371N002 754R0371N003 754R0373N003 754R0373N004 754R0342N004 754R0344N001 754R0344N002 754R0344N010 754R0338N004 754R0339N001 754R0339N002 754R0339N003 754R0327N009 754R0327N010 754R0328N001 754R0328N002 754R0324N010 754R0324N011 754R0327N007 754R0327N008 754R0311N001 754R0311N002 754R0316N001 754R0316N002 754R0309N012 754R0310N006 754R0310N008 754R0310N011 754R0290N007 754R0294N009 754R1232N001 754R1235N005 754R1061N003 754R0625N006 754R0625N007 754R0454N002 754R0411N009 754R0411N011 754R0378N004 754R1341N003 754R1012N003 754R0877N005 754R0877N007 754R0877N009 754R0878N001 754R0878N003 754R0856N009 754R0859N009 754R0860N001 754R0860N005 754R0860N006 754R0841N007 754R0842N002 754R0842N003 754R0842N004 754R0842N005 754R0837N005 754R0837N006 754R0837N007 754R0839N001 754R0839N002 754R0544N007 754R0499N005 754R0447N008 754R0390N005 754R0407N007 754R0739N003 754R0739N002 754R0739N001 754R0738N004 754R0738N003 754R0607N002 754R0560N009 754R0535N004 754R0534N002 754R0459N003 754R0459N002 754R0459N001 754R0458N007 754R0458N006 754R0458N005 754R0381N002 754R0381N004 754R0381N001 754R0371N004 754R0621N007 754R0289N009 754R1098N003 754R1098N001 754R0920N009 754R0948N007 754R0415N007 754R1153N002 754R1364N008 754R1211N006 754R1251N010 754R1252N007 754R1253N001 754R1253N002 754R1253N003 754R1253N004 754R1253N005 754R1253N008 754R1253N011 754R1133N006 754R1133N007 754R1134N001 754R1134N002 754R1134N003 754R1134N004 754R1134N007 754R1135N001 754R1135N003 754R1135N004 754R1098N005 754R1014N005 754R0966N006 754R0792N007 754R0797N001 754R0792N006 754R0712N006 754R0655N005 754R0613N002 754R0565N004 754R0565N009 754R0566N001 754R0566N002 754R0566N004 754R0419N011 754R0422N001 754R0419N012 754R0422N002 And 754R0415N008 ABL800 FLEX: 754R0560N001 754R0058N003 754R0448N001 754R0024N010 754R0122N004 754R0206N003 754R0544N006 754R0547N008 754R0547N009 754R0560N001 754R0578N003 754R0579N003 754R0617N007 754R0706N006 754R0766N002 754R0768N007 754R0953N004 754R1011N008 754R0102N002U 754R0253N009U 754R0112N001U 754R0175N003U 754R0103n004u 754R0006N010U 754R0100N006U 754R0175N009u 754R0094N001u 754R0220N012U 754R0100N007U 754R0026N003U 754R0300N004U 754R0403N022U 754R0156N003U 754R0128N006U 754R0127N002U 754R0131N004U 754R0416N007U 754R0425N012U 754R0017N010U 754R0076N004U 754R0193N005U 754R0128N005U 754R0128N004U 754R0028N007U 754R0323N011U 754R0093N006u 754R0331N004U 754R0385N012U 754R0384N004U 754R0458N008U 754R0193N007U 754R0184N007U 754R0008N003U 754R0006N008U 754R0425N011U 754R0300N009U 754R0116N001U 754R0075N008U 754R0099N005U And 754R0255N007U ABL800 FLEX with AutoCheck: 754R0445N008 754R0445N009 754R0445N010 754R0445N011 754R0445N012 754R0781N006 754R0792N001 754R0792N005 754R0792N008 754R0915N009 754R0925N001 754R0588N010 754R1366N004 754R1343N005 754R1343N004 754R1337N004 754R1273N004 754R1256N003 754R1236N002 754R1192N004 754R1158N001 754R1158N002 754R1155N001 754R1146N005 754R1083N006 754R1083N007 754R0715N008 754R1014N006 754R1014N007 754R0237N012 754R1134N006 754R0349N002 754r0357n003 754R0757N003 754R0757N004 754R0761N005 754R0761N006 754R0761N007 754R0762N001 754R0762N002 754R0724N006 754R0722N007 754R0722N008 754R0722N010 754R0724N001 754R0724N002 754R0724N003 754R0724N007 754R0472N006 754R0715N001 754R0715N002 754R0715N003 754R0715N004 754R0715N005 754R0715N006 754R0715N007 754R0399N010 754R0397N002 754R0397N004 754R0397N008 754R0399N001 754R0399N002 754R0399N003 754R0399N004 754R0399N005 754R0399N006 754R0399N007 754R0399N008 754R0399N009 754R0394N002 754R0394N003 754R0390N006 754R0397N005 754R0397N006 754R0397N003 754R0390N011 754R0348N012 754R0342N007 754R0339N007 754R0342N005 754R0342N006 754R0298N010 754R1237N005 754R1258N005 754R1192N005 754R1166N007 754R1166N006 754R1211N009 754R1192N003 754R1161N007 754R0826N001 754R0588N007 754R0391N001 754R0406N001 754R0405N012 754R0403N037 754R0360N009 754R0588N008 754R0588N009 754R0942N002 754R0942N004 754R0942N003 754R0616N001 754R0616N007 754R0616N004 754R0616N005 754R0616N006 754R0321N011 754R0321N010 754R1366N001 754R1366N003 754R1349N003 754R1347N007 754R1348N001 754R1348N002 754R1348N004 754R1348N005 754R1348N003 754R1331N001 754R1335N001 754R1335N002 754R1336N002 754R1336N003 754R1331N006 754R1331N007 754R1318N005 754R1316N004 754R1318N004 754R1306N007 754R1303N001 754R1293N003 754R1293N004 754R1286N007 754R1293N002 754R1286N001 754R1286N002 754R1286N003 754R1286N004 754R1287N003 754R1287N006 754R1287N007 754R1287N008 754R1288N001 754R1287N004 754R1287N005 754R1284N004 754R1286N005 754R1286N006 754R1286N008 754R1287N002 754R1273N005 754R1258N003 754R1258N004 754R1256N001 754R1256N002 754R1258N001 754R1258N002 754R1227N007 754R1236N009 754R1237N002 754R1237N011 754R1238N001 754R1227N005 754R1214N006 754R1214N002 754R1214N004 754R1219N008 754R1219N009 754R1220N001 754R1214N003 754R1214N005 754R1219N007 754R1212N002 754R1214N001 754R1205N008 754R1203N001 754R1203N002 754R1200N003 754R1200N007 754R1201N002 754R1198N003 754R1200N002 754R1201N003 754R1200N005 754R1201N001 754R1180N001 754R1184N001 754R1184N002 754R1184N003 754R1179N006 754R1175N005 754R1179N001 754R1179N002 754R1179N004 754R1175N002 754R1175N003 754R1175N004 754R1175N006 754R1175N007 754R1176N001 754R1176N002 754R1143N005 754R1155N002 754R1145N001 754R1145N007 754R1145N002 754R1145N003 754R1145N004 754R1145N005 754R1146N001 754R1144N007 754R1123N005 754R1122N003 754R1109N003 754R1109N004 754R1119N006 754R1120N001 754R1109N001 754R1107N001 754R1109N002 754R1083N008 754R1101N004 754R1101N005 754R1101N006 754R1076N005 754R1076N007 754R1083N005 754R1073N003 754R1073N001 754R1073N002 754R1062N002 754R1062N001 754R1062N006 754R1062N007 754R1062N003 754R1062N004 754R1062N005 754R1062N008 754R1060N005 754R1060N007 754R1058N007 75

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2162-2012

Related Recalls

Class I - Dangerous

Software Security; The action is being initiated because of software security vulnerabilities with the firm's analyzer operating system, which may cause the device to shutdown or reboot resulting in delayed medical treatment.

Jul 10, 2019 Other Medical Devices Nationwide View Details →

From three incidents in Denmark and Sweden, comparison studies on patient blood samples carried out for quality assurance demonstrated that creatinine measurements as determined on analyzers could be underestimated with up to 28% as compared with measurements from COBAS. The underestimated creatinine occurs specifically, when measurements on Autocheck 6+ level 1 (also known as S7835 Autocheck module and abbreviated AC6+L1) are less than or equal to 230 micro-meter of a specified insert interval for AC6+L1 with insert limits 210 290 micro-meter. Ultimately, a lower read-out on patient samples is observed Customers are concerned about underestimation of creatinine results in patients, which are to undergo radiocontrast x-ray investigations. If these patients are not healthy, the exposure of radiocontrast agents may induce kidney insufficiency (contrast-induced nephropathy). Hazardous situation: Underestimation of the concentration of creatinine.

Jul 2, 2018 Other Medical Devices Nationwide View Details →