lmmunalysis EDDP Urine Control Set 1 The EDDP Urine Controls (High 125 ng/ml and Low 75 ng/ml) are used as control materials in the ammonolysis EDDP Specific Urine Enzyme Immunoassay, which has a cutoff of 100 ng/ml. This test provides only a preliminary analytical test result and a more specific alternate chemical method, such as Gas Chromatography/Mass Spectrometry or Liquid Chromatography/Tandem Mass Spectrometry, must be used to obtain a confirmed analytical result.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Part Number: 3006-5, Lot Number: EK16135
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Alere San Diego Inc. DBA Immunalysis Corporation
- Reason for Recall:
- The High (125 ng/ml) Control does not meet specification - it has been confirmed to produce values below 100 ng/mL when run on validated clinical chemistry analyzers, which is below the 100 ng/mL cutoff of the EDDP Specific Urine Enzyme Immunoassay with which this control is utilized.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
lmmunalysis EDDP Urine Control Set 1 The EDDP Urine Controls (High 125 ng/ml and Low 75 ng/ml) are used as control materials in the ammonolysis EDDP Specific Urine Enzyme Immunoassay, which has a cutoff of 100 ng/ml. This test provides only a preliminary analytical test result and a more specific alternate chemical method, such as Gas Chromatography/Mass Spectrometry or Liquid Chromatography/Tandem Mass Spectrometry, must be used to obtain a confirmed analytical result.
Product Codes/Lot Numbers:
Part Number: 3006-5, Lot Number: EK16135
Distribution:
Distributed in: US, CA, CO, CT, FL, LA, MD, NC, NJ, OH, OK, TN, TX, WI, WV
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2140-2018