Medtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D00U003; 2. Percepta Serena Solara application, Software Model Number D00U004; 3. Cobalt Crome application, Software Model Number D00U005; 4. Micra VR application, Software Model Number D00U006; 5. Micra AV application, Software Model Number D00U007; 6. Viva Consulta Syncra Advisa Ensura application, Software Model Number D00U008; 7. Claria Amplia Compia application, Software Model Number D00U009; 8. Evera MRI application, Software Model Number D00U010; 9. Visia AF application, Software Model Number D00U011; 10. Viva Brava Evera application, Software Model Number D00U012; 11. Micra VR2 AV2 application, Software Model Number D00U022.
Class I - DangerousWhat Should You Do?
- Check if you have this product: 1. Software Model Number, D00U003, GTIN 00763000002039, Updated Software Version 6.5.5; 2. Software Model Number, D00U004, GTIN 00763000002046, Updated Software Version 6.5.5; 3. Software Model Number, D00U005, GTIN 00763000002053, Updated Software Version 9.5.2; 4. Software Model Number, D00U006, GTIN 00763000397852, Updated Software Version 4.4.6; 5. Software Model Number, D00U007, GTIN 00763000397869, Updated Software Version 4.4.6; 6. Software Model Number, D00U008, GTIN 00763000397876, Updated Software Version 3.5.5; 7. Software Model Number, D00U009, GTIN 00763000397883, Updated Software Version 3.4.2; 8. Software Model Number, D00U010, GTIN 00763000397890, Updated Software Version 3.4.2; 9. Software Model Number, D00U011, GTIN 00763000397906, Updated Software Version 3.4.2; 10. Software Model Number, D00U012, GTIN 00763000397913, Updated Software Version 3.4.2; 11. Software Model Number, D00U022, GTIN 00763000544300, Updated Software Version 3.7.5.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Medtronic, Inc.
- Reason for Recall:
- A software update for Medtronic CareLink SmartSync Device Manager (SmartSync) is now available. The update eliminates the potential display of an erroneous electrical reset (pop-up) message upon interrogation with a SmartSync Device Manager. The erroneous message indicates an electrical reset has occurred. However, no reset has actually occurred and there are no parameter changes or loss of diagnostics. Prior to this software update, if an erroneous SmartSync message was displayed, confusion regarding the message could have caused a clinician to consider an unnecessary system revision. Medtronic has received 39 reports of this display error with two instances of unnecessary device explant through 30 April 2025.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Medtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D00U003; 2. Percepta Serena Solara application, Software Model Number D00U004; 3. Cobalt Crome application, Software Model Number D00U005; 4. Micra VR application, Software Model Number D00U006; 5. Micra AV application, Software Model Number D00U007; 6. Viva Consulta Syncra Advisa Ensura application, Software Model Number D00U008; 7. Claria Amplia Compia application, Software Model Number D00U009; 8. Evera MRI application, Software Model Number D00U010; 9. Visia AF application, Software Model Number D00U011; 10. Viva Brava Evera application, Software Model Number D00U012; 11. Micra VR2 AV2 application, Software Model Number D00U022.
Product Codes/Lot Numbers:
1. Software Model Number, D00U003, GTIN 00763000002039, Updated Software Version 6.5.5; 2. Software Model Number, D00U004, GTIN 00763000002046, Updated Software Version 6.5.5; 3. Software Model Number, D00U005, GTIN 00763000002053, Updated Software Version 9.5.2; 4. Software Model Number, D00U006, GTIN 00763000397852, Updated Software Version 4.4.6; 5. Software Model Number, D00U007, GTIN 00763000397869, Updated Software Version 4.4.6; 6. Software Model Number, D00U008, GTIN 00763000397876, Updated Software Version 3.5.5; 7. Software Model Number, D00U009, GTIN 00763000397883, Updated Software Version 3.4.2; 8. Software Model Number, D00U010, GTIN 00763000397890, Updated Software Version 3.4.2; 9. Software Model Number, D00U011, GTIN 00763000397906, Updated Software Version 3.4.2; 10. Software Model Number, D00U012, GTIN 00763000397913, Updated Software Version 3.4.2; 11. Software Model Number, D00U022, GTIN 00763000544300, Updated Software Version 3.7.5.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2127-2025
Related Recalls
ventricular fibrillation (VF) during radiofrequency (RF) therapy with the Medtronic Sphere-9 catheter in patients who have a Biotronik Implantable Cardioverter Defibrillator (ICD) or Biotronik Cardiac Resynchronization Therapy-Defibrillator (CRT-D) implanted. The observation is caused by an interaction between the Sphere-9 catheter current flow and a safety feature built into Biotronik ICDs and CRT Ds. Medtronic recommends physicians use increased caution when using the Sphere-9 catheter for RF ablation therapy in patients implanted with a Biotronik ICD/CRT-D system, and to ensure external defibrillation systems are readily available.
There is a potential for delayed time to high-voltage (HV) therapy should a rare sequence of events occur.
In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of time for the user to abort a therapy, thereby limiting the user's ability to cancel a high voltage therapy delivery.