GE Healthcare Vivid S70N, ultrasound device
Class I - DangerousWhat Should You Do?
- Check if you have this product: Software versions: v203, v204, v205, v206
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- GE Vingmed Ultrasound As
- Reason for Recall:
- GE HealthCare has become aware that certain Vivid ultrasound systems cannot boot up in a timely fashion. If this occurs, it can delay availability of the device in time-critical emergency situations.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
GE Healthcare Vivid S70N, ultrasound device
Product Codes/Lot Numbers:
Software versions: v203, v204, v205, v206
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2127-2023
Related Recalls
GE Vivid Ultrasound products, Model Numbers: a) Vivid S70, H45041SW
GE Vingmed Ultrasound As
GE HealthCare has become aware of an issue with certain Vivid ultrasound systems where, due to a slow DICOM Worklist response time, incorrect patient name and demographics can be displayed on the ultrasound scanner screen rather than the intended patient selected by the user.
GE Vivid Ultrasound products, Model Numbers: a) Vivid E90, H45581LB; b) Vivid E95, H45581DA
GE Vingmed Ultrasound As
GE HealthCare has become aware of an issue with certain Vivid ultrasound systems where, due to a slow DICOM Worklist response time, incorrect patient name and demographics can be displayed on the ultrasound scanner screen rather than the intended patient selected by the user.
GE Healthcare Vivid S60N, ultrasound device
GE Vingmed Ultrasound As
GE HealthCare has become aware that certain Vivid ultrasound systems cannot boot up in a timely fashion. If this occurs, it can delay availability of the device in time-critical emergency situations.