Cios Alpha Image-Intensified Fluoroscopic X-Ray System, Model Numbers VA20-10308191 and VA30-10308194 The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial Numbers: 11200 13288 13259 12339 13241 10019 12377 13219 12946 13232 13045 12618 12972 40109 40110 40073 40057 40013 40082 40093 40078 40107 40108 40111 40120 40126 40134 40135 40104 40074 40095 40076 40080 40058 40077 40081 40055 40054 40075 40125 40113 40102 40071 40061 40094 40124 40070 40103 40118 40119 40121 40084 40122 40123 40083 40062 40130 40116
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Siemens Medical Solutions USA, Inc
- Reason for Recall:
- The unplugging of the main cable on the monitor trolley from the X10 connector can cause an unsafe electrical voltage (up to 67,6 V DC) at the X10 connector. This can result in an electrical shock if the user or other persons touch the contact pins of the X10 connector.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Cios Alpha Image-Intensified Fluoroscopic X-Ray System, Model Numbers VA20-10308191 and VA30-10308194 The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications.
Product Codes/Lot Numbers:
Serial Numbers: 11200 13288 13259 12339 13241 10019 12377 13219 12946 13232 13045 12618 12972 40109 40110 40073 40057 40013 40082 40093 40078 40107 40108 40111 40120 40126 40134 40135 40104 40074 40095 40076 40080 40058 40077 40081 40055 40054 40075 40125 40113 40102 40071 40061 40094 40124 40070 40103 40118 40119 40121 40084 40122 40123 40083 40062 40130 40116
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2127-2020
Related Recalls
MAMMOMAT Inspiration;
Siemens Medical Solutions USA
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.
MAMMOMAT Fusion;
Siemens Medical Solutions USA
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.
MAMMOMAT Revelation;
Siemens Medical Solutions USA
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.