Medtronic Nexdrive Micropositioning Drive. Models Ml-1000 and Ml-2000. For use in conjunction with the Medtronic Nexframe Stereotactic System for the precise positioning of microelectrodes and implantable leads. A stereotactic guidance system used in conjunction with Medtronic StealthStation Navigation Systems-image-guided surgery (IGS) systems-for Deep Brain Stimulation (DBS) procedures.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lot numbers: 082605614, 082601414, 082602014, 082634313, 082630913, 082629713, 082629613, 082626913, 082623113, 082621413, and 082618213. EXPANDED 10/02/2014 Lot Numbers: 082616214, 082616314, 082618314, and 082618414.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Medtronic Neuromodulation
Reason for Recall:
Potential for misalignment of the Z-stage scale. Using one of these devices for a procedure could result in the microelectrode being inserted to an incorrect target depth.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Medtronic Nexdrive Micropositioning Drive. Models Ml-1000 and Ml-2000. For use in conjunction with the Medtronic Nexframe Stereotactic System for the precise positioning of microelectrodes and implantable leads. A stereotactic guidance system used in conjunction with Medtronic StealthStation Navigation Systems-image-guided surgery (IGS) systems-for Deep Brain Stimulation (DBS) procedures.

Product Codes/Lot Numbers:

Lot numbers: 082605614, 082601414, 082602014, 082634313, 082630913, 082629713, 082629613, 082626913, 082623113, 082621413, and 082618213. EXPANDED 10/02/2014 Lot Numbers: 082616214, 082616314, 082618314, and 082618414.

Distribution:

Distributed in: CA, CO, FL, IL, KS, KY, LA, MN, MT, NC, NE, OR, TX, VA, UT

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2118-2014

Related Recalls

Class I - Dangerous

Evaluation handsets may not be able to communicate with the neurostimulator due to handsets having a faster processing speed than prior handsets. An inability to connect may cause the therapy to stop being delivered and a system error will be displayed. Patient may experience return of underlying disease symptoms (e.g. overactive bladder, urinary retention, or fecal incontinence) based on their response to stimulation during the evaluation period.

Apr 3, 2025 Other Medical Devices Nationwide View Details →