Giraffe OmniBed Carestation CS1, Model Numbers: a. 2082844-001-XXX, b. 2074920-001-XXXXXX, c. M1050205, d. M1135822, e. M1145475, f. M1145476, g. M1235957, h. M1149051, i. M1146098; Labeled as the following: GIRAFFE OMNIBED CARESTATION CS1, GIRAFFE OMNIBED CARESTATION, GIRAFFE OMNIBED, GIRAFFE OMNIBED CS1, OMNI G 115/230V SERVO2 CARESTATION, ASSY-LAU, OMNIBED CHINA EBASE S HUMID, Manufacturing assembly - Make, OBSOLETE-Omnibed Czech Servo O2 Nist/Scale/Re-usable Probe/SW w/port hole/CE Plug, ASSY-LAU, Omnibed Czech S-PH w/Nist O2 w/Reus Probe CE Power Cord, Finished good - Make, ASSY-LAU, Omnibed Czech S-PH w/Reusable Probe CE Power Cord, Finished good - Make, Omnibed Czech Servo O2 Nist/Scale/Re-usable Probe/SW w/port hole/Ce Plug, ASSY-LAU, Omnibed Indonesian S-GR W/Reuseable Probe CE Plug, Finished good - Make, ASSY-LAU, Omnibed ENG S-Gr w/REUProbe CE Power Cord, Finished good - Make, ASSY-LAU, Omnibed Portuguese S-PH w/NIST O2w/Reusable Probe CE Power Cord, Finished good - Make. Incubator, Neonatal, Warmer, Infant radiant

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    a. Model Number 2082844-001-XXX; GTIN Numbers: 010084068211686221, 010084068211686221; All Lot/Serial Numbers. b. Model Number 2074920-001-XXXXXX; No GTIN Number; Serial Numbers: HDGR52137, HDGL50099, HDGT60164, HDGT60143, HDGT60142, HDGT60139, HDGT60138, HDGT60137, HDGU60753, HDGT60652, HDGU60069, HDGT60141, HDGU60156, HDGU60120, HDGU60119, HDGT60163, HDGT60703, HDGT60245, HDGT60160, HDGT60459, HDGT60246, HDGT60235, HDGU67645, HDGU60065, HDGU60063, HDGU60061, HDGU63122, HDGU63121, HDGU60062, HDGU57534, HDGU67533, HDGU67534, HDGT60460, HDGU67236, HDGT60249, HDGU60201, HDGU60570, HDGU60620, HDGU60023, HDGU60020, HDGT60612, HDGU60022, HDGU60571, HDGU60686, HDGU60202, HDGV60190, HDGW60244, HDGU67677, HDGU67676, HDGU67675, HDGU67280, HDGU67416, HDGU67415, HDGU67382, HDGU67377, HDGU67369, HDGU67366, HDGU67362, HDGU67289, HDGU67284, HDGU67282, HDGU63165, HDGU63167, HDGU63166, HDGV61084, HDGU67436, HDGU67423, HDGU67422, HDGU67421, HDGU67420, HDGU67419, HDGU67418, HDGU67414, HDGU67413, HDGU67381, HDGU67380, HDGU67379, HDGU67378, HDGU67370, HDGU67368, HDGU67367, HDGU67365, HDGU67363, HDGU67361, HDGU67360, HDGU67359, HDGU67358, HDGU67357, HDGU67304, HDGU67303, HDGU67302, HDGU67301, HDGU67300, HDGU67299, HDGU67298, HDGU67297, HDGU67296, HDGU67295, HDGU67294, HDGU67293, HDGU67292, HDGU67291, HDGU67290, HDGU67288, HDGU67287, HDGU67286, HDGU67285, HDGU67283, HDGU67281, HDGJ51159, HDGT60613, HDGT60614, HDGU60700, HDGU60703. c. Model Number M1050205; No GTIN Number; Serial Numbers: HDGV60148, HDGU60656, HDGU60658, HDGT60481, HDGS50047, HDGV60146. d. Model Number M1135822; No GTIN Number; Serial Numbers: HDGL51737, HDGL51738, HDGL51734, HDGM50736, HDGM50841, HDGL51735. e. Model Number M1145475; No GTIN Number; Serial Numbers: HDGM51813, HDGM52120, HDGM52119, HDGN50474, HDGN50472, HDGN50471, HDGN50469, HDGN50466, HDGN50464, HDGN50848, HDGN50847, HDGN50846, HDGN50845, HDGN50844, HDGQ50049, HDGR53203, HDGS51654, HDGS51653, HDGS51652, HDGS51651, HDGS51650, HDGS51649, HDGS51648, HDGS51647, HDGS51646, HDGS51645, HDGM51179, HDGQ50425, HDGN50465, HDGN50849, HDGN50467, HDGN50470, HDGN50475, HDGR52324, HDGR52323, HDGR52322, HDGR52321, HDGN50468. f. Model Number M1145476; No GTIN Number; Serial Numbers: HDGR52896. g. Model Number M1235957; No GTIN Number; Serial Numbers: HDGT60222. h. Model Number M1149051; No GTIN Number; Serial Numbers: HDGR53018, HDGR53085, HDGR52927, HDGR53086, HDGR53021. i. Model Number M1146098; No GTIN Number; Serial Numbers: HDGS52233, HDGS52234.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
DATEX--OHMEDA, INC.
Reason for Recall:
GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Giraffe OmniBed Carestation CS1, Model Numbers: a. 2082844-001-XXX, b. 2074920-001-XXXXXX, c. M1050205, d. M1135822, e. M1145475, f. M1145476, g. M1235957, h. M1149051, i. M1146098; Labeled as the following: GIRAFFE OMNIBED CARESTATION CS1, GIRAFFE OMNIBED CARESTATION, GIRAFFE OMNIBED, GIRAFFE OMNIBED CS1, OMNI G 115/230V SERVO2 CARESTATION, ASSY-LAU, OMNIBED CHINA EBASE S HUMID, Manufacturing assembly - Make, OBSOLETE-Omnibed Czech Servo O2 Nist/Scale/Re-usable Probe/SW w/port hole/CE Plug, ASSY-LAU, Omnibed Czech S-PH w/Nist O2 w/Reus Probe CE Power Cord, Finished good - Make, ASSY-LAU, Omnibed Czech S-PH w/Reusable Probe CE Power Cord, Finished good - Make, Omnibed Czech Servo O2 Nist/Scale/Re-usable Probe/SW w/port hole/Ce Plug, ASSY-LAU, Omnibed Indonesian S-GR W/Reuseable Probe CE Plug, Finished good - Make, ASSY-LAU, Omnibed ENG S-Gr w/REUProbe CE Power Cord, Finished good - Make, ASSY-LAU, Omnibed Portuguese S-PH w/NIST O2w/Reusable Probe CE Power Cord, Finished good - Make. Incubator, Neonatal, Warmer, Infant radiant

Product Codes/Lot Numbers:

a. Model Number 2082844-001-XXX; GTIN Numbers: 010084068211686221, 010084068211686221; All Lot/Serial Numbers. b. Model Number 2074920-001-XXXXXX; No GTIN Number; Serial Numbers: HDGR52137, HDGL50099, HDGT60164, HDGT60143, HDGT60142, HDGT60139, HDGT60138, HDGT60137, HDGU60753, HDGT60652, HDGU60069, HDGT60141, HDGU60156, HDGU60120, HDGU60119, HDGT60163, HDGT60703, HDGT60245, HDGT60160, HDGT60459, HDGT60246, HDGT60235, HDGU67645, HDGU60065, HDGU60063, HDGU60061, HDGU63122, HDGU63121, HDGU60062, HDGU57534, HDGU67533, HDGU67534, HDGT60460, HDGU67236, HDGT60249, HDGU60201, HDGU60570, HDGU60620, HDGU60023, HDGU60020, HDGT60612, HDGU60022, HDGU60571, HDGU60686, HDGU60202, HDGV60190, HDGW60244, HDGU67677, HDGU67676, HDGU67675, HDGU67280, HDGU67416, HDGU67415, HDGU67382, HDGU67377, HDGU67369, HDGU67366, HDGU67362, HDGU67289, HDGU67284, HDGU67282, HDGU63165, HDGU63167, HDGU63166, HDGV61084, HDGU67436, HDGU67423, HDGU67422, HDGU67421, HDGU67420, HDGU67419, HDGU67418, HDGU67414, HDGU67413, HDGU67381, HDGU67380, HDGU67379, HDGU67378, HDGU67370, HDGU67368, HDGU67367, HDGU67365, HDGU67363, HDGU67361, HDGU67360, HDGU67359, HDGU67358, HDGU67357, HDGU67304, HDGU67303, HDGU67302, HDGU67301, HDGU67300, HDGU67299, HDGU67298, HDGU67297, HDGU67296, HDGU67295, HDGU67294, HDGU67293, HDGU67292, HDGU67291, HDGU67290, HDGU67288, HDGU67287, HDGU67286, HDGU67285, HDGU67283, HDGU67281, HDGJ51159, HDGT60613, HDGT60614, HDGU60700, HDGU60703. c. Model Number M1050205; No GTIN Number; Serial Numbers: HDGV60148, HDGU60656, HDGU60658, HDGT60481, HDGS50047, HDGV60146. d. Model Number M1135822; No GTIN Number; Serial Numbers: HDGL51737, HDGL51738, HDGL51734, HDGM50736, HDGM50841, HDGL51735. e. Model Number M1145475; No GTIN Number; Serial Numbers: HDGM51813, HDGM52120, HDGM52119, HDGN50474, HDGN50472, HDGN50471, HDGN50469, HDGN50466, HDGN50464, HDGN50848, HDGN50847, HDGN50846, HDGN50845, HDGN50844, HDGQ50049, HDGR53203, HDGS51654, HDGS51653, HDGS51652, HDGS51651, HDGS51650, HDGS51649, HDGS51648, HDGS51647, HDGS51646, HDGS51645, HDGM51179, HDGQ50425, HDGN50465, HDGN50849, HDGN50467, HDGN50470, HDGN50475, HDGR52324, HDGR52323, HDGR52322, HDGR52321, HDGN50468. f. Model Number M1145476; No GTIN Number; Serial Numbers: HDGR52896. g. Model Number M1235957; No GTIN Number; Serial Numbers: HDGT60222. h. Model Number M1149051; No GTIN Number; Serial Numbers: HDGR53018, HDGR53085, HDGR52927, HDGR53086, HDGR53021. i. Model Number M1146098; No GTIN Number; Serial Numbers: HDGS52233, HDGS52234.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2100-2023

Related Recalls

Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.