Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH Nuclear Medicine System

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Serial Numbers: Millennium VG 1074; "Varicam Millennium VG Discovery VH" 1003 1092 10147 10242 10374 10378 10539 10543 10550 10553 10561 10569 10570 10571 10579 10582 10587 10599 10604 3101-1502
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Reason for Recall:
Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH Nuclear Medicine System

Product Codes/Lot Numbers:

Serial Numbers: Millennium VG 1074; "Varicam Millennium VG Discovery VH" 1003 1092 10147 10242 10374 10378 10539 10543 10550 10553 10561 10569 10570 10571 10579 10582 10587 10599 10604 3101-1502

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2098-2025

Related Recalls

Class I - Dangerous

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

Sep 12, 2025 Diagnostic Equipment View Details →
Class I - Dangerous

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

Sep 12, 2025 Diagnostic Equipment View Details →

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

Sep 12, 2025 Diagnostic Equipment View Details →