VORTRAN Automatic Resuscitator, VARPlus with Extension Kit; Model PCE-5012; Qty: 10 each. Anesthesiology: This device is to be used by properly trained personnel to deliver emergency, short term, constant flow, pressure cycled ventilatory support.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model PCE 5012; Lot number/ expiry date: 1419 PCE , 2019-03-12; 1434 PCE, 2019-06-04.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Vortran Medical Technology 1, Inc
Reason for Recall:
Vortran received a customer complaint that the spin-nut-DSS connector was assembled in the reverse direction on the oxygen tubing; making the connector unable to be screwed on due to the wrong thread direction.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

VORTRAN Automatic Resuscitator, VARPlus with Extension Kit; Model PCE-5012; Qty: 10 each. Anesthesiology: This device is to be used by properly trained personnel to deliver emergency, short term, constant flow, pressure cycled ventilatory support.

Product Codes/Lot Numbers:

Model PCE 5012; Lot number/ expiry date: 1419 PCE , 2019-03-12; 1434 PCE, 2019-06-04.

Distribution:

Distributed in: CA, MS, FL

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2092-2017

Related Recalls

Vortran received a customer complaint that the spin-nut-DSS connector was assembled in the reverse direction on the oxygen tubing; making the connector unable to be screwed on due to the wrong thread direction.

Apr 20, 2017 Patient Monitors View Details →