Flexible Drill Shaft, FMP Acetabular System Instrumentation REF 803-05-239 RX Encore Medical, L.P.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot # 262667L01
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Encore Medical, Lp
- Reason for Recall:
- Manufacturing error, the drill bit shaft is unable to properly capture the drill bit pin. Failure of the drill bit shaft to capture the drill bit pin could result in the drill bit coming loose which could result in a delay in surgery or adverse patient reaction resulting in a revision surgery.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Flexible Drill Shaft, FMP Acetabular System Instrumentation REF 803-05-239 RX Encore Medical, L.P.
Product Codes/Lot Numbers:
Lot # 262667L01
Distribution:
Distributed in: US, MS, NY, MO, OK, UT, AZ
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2090-2019
Related Recalls
Their is a potential that the reamer may kick or bind up during or immediately prior to use.
Their is a potential that the reamer may kick or bind up during or immediately prior to use.
Their is a potential that the reamer may kick or bind up during or immediately prior to use.