Recall Details
- Company:
- Randox Laboratories, Limited
- Reason for Recall:
- There was a report of a malfunction with the sensor which resulted in an increase of temperature leading to melting of internal wires and scorch damage to the internal structure of the instrument.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
RX Daytona, Model Nos. RX4040 and RX4041
Product Codes/Lot Numbers:
All serial numbers
Distribution:
Distributed in: CA, CT, MA, MD, MO, NC, NJ, NY, OH, PA, UT, WV, PR
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2083-2019
Related Recalls
Randox Lipoprotein (a) - In vitro diagnostic (IVD) use as Randox Lipoprotein (a) assay on automated analysers. Catalog Number: LP2757
Randox Laboratories, Limited
Lipoprotein (a) on Roche Cobas c501 with nmol/l claims does not have a 510(k) or PMA.
Randox Lipoprotein (a) - In vitro diagnostic (IVD) use as Randox Lipoprotein (a) assay on automated analysers. Catalog Number: LP3403
Randox Laboratories, Limited
Lipoprotein (a) on Roche Cobas c501 with nmol/l claims does not have a 510(k) or PMA.
RX Daytona Plus (with ISE/without ISE); Models Nos. RX4040 (w/ISE) and RX4041 (without ISE) - Product Usage: A fully automated, random access, clinical chemistry analyzer complete with dedicated analyzer software. For use in clinical laboratories.
Randox Laboratories, Limited
Software version UI2550642107 for the RX Daytona + instrument released to correct the following issues: 1) The order of samples displayed is not sequential according to SID (sample identification number) when searching the run/results screen; 2) Incorrect color flagging of IQC as "out of range."