ARCHITECT BNP Controls Ptoduct Usage: The ARCHITECT BNP Controls are for the estimation of test precision and the detection of systematic analytical deviations of the ARCHITECT iSystem (reagents, calibrators and instrument), when used for the quantitative determination of human B type natriuretic peptide (BNP) in human EDTA plasma.
Class I - DangerousWhat Should You Do?
- Check if you have this product: a. List Number 8K28 11 - Lot Number (Expiration Date) UDI: 44K78918 (18JUN2019) (01)00380740017477(17)190618(10)44K78918, 44K80918 (05SEP2019) (01)00380740017477(17)190905(10)44K80918, 44K82718 (15NOV2019) (01)00380740017477(17)191115(10)44K82718; b. List Number 8K28 12 - Lot Number (Expiration Date) UDI: 44K79518 (23Jul2019) (01)00380740145262(17)190723(10)44K79518, 44K81118 (5Sep2019) (01)00380740145262(17)190905(10)44K81118, 44K82818 (15Nov2019) (01)00380740145262(17)191115(10)44K82818, 44K84219 (17Dec2019) (01)00380740145262(17)191217(10)44K84219
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Abbott Laboratories
- Reason for Recall:
- Presence of a time dependent, stability drift in patient and control results returned from ARCHITECT BNP testing.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
ARCHITECT BNP Controls Ptoduct Usage: The ARCHITECT BNP Controls are for the estimation of test precision and the detection of systematic analytical deviations of the ARCHITECT iSystem (reagents, calibrators and instrument), when used for the quantitative determination of human B type natriuretic peptide (BNP) in human EDTA plasma.
Product Codes/Lot Numbers:
a. List Number 8K28 11 - Lot Number (Expiration Date) UDI: 44K78918 (18JUN2019) (01)00380740017477(17)190618(10)44K78918, 44K80918 (05SEP2019) (01)00380740017477(17)190905(10)44K80918, 44K82718 (15NOV2019) (01)00380740017477(17)191115(10)44K82718; b. List Number 8K28 12 - Lot Number (Expiration Date) UDI: 44K79518 (23Jul2019) (01)00380740145262(17)190723(10)44K79518, 44K81118 (5Sep2019) (01)00380740145262(17)190905(10)44K81118, 44K82818 (15Nov2019) (01)00380740145262(17)191115(10)44K82818, 44K84219 (17Dec2019) (01)00380740145262(17)191217(10)44K84219
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2079-2019
Related Recalls
Software issue for hq analyzer results in system not visibly applying appropriate flagging to results when saturation is present which may result in incorrect results.
Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-01 Software Version: All software versions Product Description: CELL-DYN Ruby Component: No
Abbott Laboratories
When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.
Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-10 Software Version: All software versions Product Description: CELL-DYN Ruby Component: No
Abbott Laboratories
When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.