Zyno Medical Administration Set -105"ADMIN SET 20DP, W/2 Y SITE, CVCLMPS- For Use with Zyno Medical Infusion pumps or Gravity Feed Product Code:B2-70072
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot Numbers:18037128,18065375, 19015969 UDl/UPC/GTIN: 00814371020198
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Zyno Medical LLC
- Reason for Recall:
- Administration sets assembled with inverted pinch clamp cause the fluid from the distal end of the administration set to be pumped to the proximal end of the tubing when secondary set is used with the primary set. Thus, instead of delivering fluid to the patient, blood may be drawn from the patient if there is space in the drug container to hold extra volume of fluid, under the condition that the reversed pinch clamp is not identified
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Zyno Medical Administration Set -105"ADMIN SET 20DP, W/2 Y SITE, CVCLMPS- For Use with Zyno Medical Infusion pumps or Gravity Feed Product Code:B2-70072
Product Codes/Lot Numbers:
Lot Numbers:18037128,18065375, 19015969 UDl/UPC/GTIN: 00814371020198
Distribution:
Distributed in: MA
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2058-2019
Related Recalls
Unreleased software versions were installed on distributed devices without verification or validation.
Unreleased software versions were installed on distributed devices without verification or validation.
Unreleased software versions were installed on distributed devices without verification or validation.