Philips C5-2 Transducer.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model No.989605389242, 989605454101, 989605398211, 989605450353, 989605451181, 989605477131, 989605450351, 989605450352; UDI: (01)00884838067424(21)B36LWB, (01)00884838067424(21)B2YZPM, (01)00884838067424(21)B34JQB, (01)00884838067424(21)F01L9N, (01)00884838067424(21)B3V8CH, (01)00884838067424(21)B3LJYB, (01)00884838067424(21)B2C4FQ, (01)00884838067424(21)F010MV, (01)00884838067424(21)B3VGXG, (01)00884838067424(21)F03968, (01)00884838067424(21)B31MT5, (01)00884838067424(21)B28T00, (01)00884838067424(21)B2C44Y, (01)00884838067424(21)B3D56M, (01)00884838067424(21)B22P69, (01)00884838067424(21)B3Q81Z, (01)00884838067424(21)B20145, (01)00884838067424(21)B3R9R3, (01)00884838067424(21)B3G9N5, (01)00884838067424(21)F0174Y, (01)00884838067424(21)B200TW, (01)00884838067424(21)B224WW, (01)00884838067424(21)B3C3W7, (01)00884838067424(21)B2JX46, (01)00884838067424(21)B2VM4B, (01)00884838067424(21)B30VBF, (01)00884838067424(21)B3WD3C, (01)00884838067424(21)B39XCG, (01)00884838067424(21)B2L1X5, (01)00884838067424(21)B2VLYK, (01)00884838067424(21)B3LZ50, (01)00884838067424(21)B3V9LJ, (01)00884838067424(21)B25N1F, (01)00884838067424(21)B231L5, (01)00884838067424(21)B31MT1, (01)00884838067424(21)B2C3W9, (01)00884838067424(21)F016D1, (01)00884838067424(21)B22NVD, (01)00884838067424(21)F008M1, (01)00884838067424(21)B1XBNC, (01)00884838067424(21)B30V7Y, (01)00884838067424(21)B34JXT, (01)00884838067424(21)B3QQX0, (01)00884838067424(21)B2JWTF, (01)00884838067424(21)B1Y1D9, (01)00884838067424(21)F02QD4, (01)00884838067424(21)B34T00, (01)00884838067424(21)F01K6Z, (01)00884838067424(21)B3P65C, (01)00884838067424(21)B224KM, (01)00884838067424(21)B33706, (01)00884838067424(21)B32B9T, (01)00884838067424(21)B2ND6N, (01)00884838067424(21)B39XDX, (01)00884838067424(21)B2WDTZ, (01)00884838067424(21)B2HNP4, (01)00884838067424(21)B2YZT6, (01)00884838067424(21)B30CJT, (01)00884838067424(21)B39ZX8, (01)00884838067424(21)B200T8, (01)00884838067424(21)B37D0B, (01)00884838067424(21)B2WZYX, (01)00884838067424(21)B2JXDC, (01)00884838067424(21)B2FY26, (01)00884838067424(21)B34MHH, (01)00884838067424(21)B2Q8W4, (01)00884838067424(21)B2XBFY, (01)00884838067424(21)B2L1W0, (01)00884838067424(21)B22P40, (01)00884838067424(21)B2C5P4, (01)00884838061385(21)B127JJ, (01)00884838061385(21)B11G3D, (01)00884838067424(21)B224KD, (01)00884838061385(21)B0JJLQ, (01)00884838061385(21)B0NG16, (01)00884838067424(21)F019T7, (01)00884838067424(21)B3H966, (01)00884838067424(21)F01GFM, (01)00884838067424(21)F0119H, (01)00884838067424(21)B39FB8, (01)00884838067424(21)F002CB, (01)00884838067424(21)B36B5B, (01)00884838067424(21)B23VFT, (01)00884838067424(21)F013HC, (01)00884838067424(21)B2R8T8, (01)00884838067424(21)B3G8ZY, (01)00884838067424(21)B2NDGV, (01)00884838067424(21)B37D8Y, (01)00884838067424(21)F005XT, (01)00884838067424(21)B2Y9YX, (01)00884838067424(21)B3RKBD, (01)00884838067424(21)B37D2W, (01)00884838067424(21)F02QD3, (01)00884838067424(21)B3V9NL, (01)00884838067424(21)B1XB2Q, (01)00884838067424(21)B2C43R, (01)00884838067424(21)B2L1T9, (01)00884838067424(21)B3Q81N, (01)00884838067424(21)B37D34, (01)00884838067424(21)B34JXC, (01)00884838067424(21)B3M9F3, (01)00884838067424(21)B2C5FL, (01)00884838067424(21)B35YBH, (01)00884838067424(21)B2M84V, (01)00884838067424(21)F004KL, (01)00884838067424(21)B3DQCW, (01)00884838067424(21)F02DPJ, (01)00884838067424(21)B2C40T, (01)00884838067424(21)B224RL, (01)00884838067424(21)B231GL, (01)00884838067424(21)B3VGXP, (01)00884838067424(21)B3RJXK, (01)00884838067424(21)B224JV, (01)00884838067424(21)F01C47, (01)00884838067424(21)B3VCQD, (01)00884838067424(21)B28RPC, (01)00884838067424(21)B3QR18, (01)00884838067424(21)F002CD, (01)00884838067424(21)B20170, (01)00884838067424(21)F00LD5, (01)00884838067424(21)B2C449, (01)00884838067424(21)B34T3Z, (01)00884838067424(21)B32B77, (01)00884838067424(21)B2YZXG, (01)00884838067424(21)B2M78R, (01)00884838067424(21)B2HN85, (01)00884838067424(21)B2YZQF, (01)00884838067424(21)B25M1W, (01)00884838067424(21)B38WW3, (01)00884838067424(21)B39XG1, (01)00884838067424(21)B2L1Y9, (01)00884838061385(21)B0T4G8, (01)00884838067424(21)B1XBJZ, (01)00884838061385(21)B0HNH9, (01)00884838061385(21)B0WNP3, (01)00884838061385(21)B12WHD, (01)00884838061385(21)B0QVLH, (01)00884838067424(21)B2BBKQ, (01)00884838067424(21)B28RNX, (01)00884838067424(21)B3M9CY, (01)00884838067424(21)B2JWX4, (01)00884838067424(21)B200R1, (01)00884838067424(21)B2VLNB, (01)00884838067424(21)B25N07, (01)00884838067424(21)B3VGZP, (01)00884838067424(21)B38WQ2, (01)00884838067424(21)B3RKB7, (01)00884838067424(21)B37DL7, (01)00884838067424(21)B34JQR, (01)00884838067424(21)B23W05, (01)00884838067424(21)B3V9FD, (01)00884838067424(21)F0103N, (01)00884838067424(21)F014WR, (01)00884838067424(21)B36LRF, (01)00884838067424(21)B35YFJ, (01)00884838067424(21)B307T7, (01)00884838067424(21)B31MQ3, (01)00884838067424(21)B3JQTT, (01)00884838067424(21)B1XB6C, (01)00884838067424(21)B3F4T5, (01)00884838067424(21)B39X4R, (01)00884838067424(21)B3C3PN, (01)00884838067424(21)B3F508, (01)00884838067424(21)B3VCQT, (01)00884838067424(21)F027KD, (01)00884838067424(21)B20156, (01)00884838067424(21)B3RB11, (01)00884838067424(21)F00LDG, (01)00884838067424(21)B2BBKL, (01)00884838067424(21)B3R49J, (01)00884838067424(21)F00PJ0, (01)00884838067424(21)B23W0X, (01)00884838067424(21)F016L9, (01)00884838067424(21)B2C42R, (01)00884838067424(21)B217CD, (01)00884838067424(21)B2NDNG, (01)00884838067424(21)B3F5DY, (01)00884838067424(21)B3V9JP, (01)00884838067424(21)B2HN60, (01)00884838067424(21)B22P38, (01)00884838067424(21)B2Z0NZ, (01)00884838067424(21)B27XBQ, (01)00884838067424(21)B2L1VZ, (01)00884838067424(21)B224QQ, (01)00884838067424(21)B2YZQC, (01)00884838067424(21)B200VX, (01)00884838067424(21)B2C4XN, (01)00884838067424(21)B2JX23, (01)00884838067424(21)B2VLYV, (01)00884838067424(21)B38WRX, (01)00884838067424(21)B2885B, (01)00884838067424(21)B2M88N, (01)00884838067424(21)B34MFB, (01)00884838067424(21)B2Y9NP, (01)00884838067424(21)B37CZZ, (01)00884838067424(21)B2YBHP, (01)00884838061385(21)B09PPC, (01)00884838061385(21)B0JJR9, (01)00884838067424(21)B2016G, (01)00884838061385(21)B0NM53, (01)00884838061385(21)B07ZMN, (01)00884838061385(21)B0T1D9, (01)00884838067424(21)B2HNMQ, (01)00884838067424(21)B3NTB4, (01)00884838067424(21)B35YCB, (01)00884838067424(21)B3M8K5, (01)00884838067424(21)B3VTKH, (01)00884838067424(21)B29W4H, (01)00884838067424(21)B3KT38, (01)00884838067424(21)B2Y9W4, (01)00884838067424(21)B224VG, (01)00884838067424(21)B3DQFC, (01)00884838067424(21)B2Q8TB, (01)00884838067424(21)B28GZJ, (01)00884838067424(21)B1Y1KN, (01)00884838067424(21)B3M8JM, (01)00884838067424(21)F01BFY, (01)00884838067424(21)B224P8, (01)00884838067424(21)B3MGQ2, (01)00884838067424(21)B28MJ4, (01)00884838067424(21)B28RTL, (01)00884838067424(21)B2XQ1T, (01)00884838067424(21)B3QZ1R, (01)00884838067424(21)B2M84T, (01)00884838067424(21)B224RW, (01)00884838067424(21)F01C44, (01)00884838067424(21)F00BR0, (01)00884838067424(21)B2C4FW, (01)00884838067424(21)B3777H, (01)00884838067424(21)F02T5G, (01)00884838067424(21)B3KHV2, (01)00884838067424(21)B2018Y, (01)00884838067424(21)B3RB2L, (01)00884838067424(21)F01338, (01)00884838067424(21)B3C4PZ, (01)00884838067424(21)B3JWP9, (01)00884838067424(21)B25N0D, (01)00884838067424(21)B3LZ0Q, (01)00884838067424(21)F012L3, (01)00884838067424(21)B1Y18D, (01)00884838067424(21)B2JWW0, (01)00884838067424(21)B23VFZ, (01)00884838067424(21)B3CWQ6, (01)00884838067424(21)B1Y1HR, (01)00884838067424(21)B34JZH, (01)00884838067424(21)B3CM0P, (01)00884838067424(21)B29JJ4, (01)00884838067424(21)B37DDK, (01)00884838067424(21)B2C5JP, (01)00884838067424(21)F01C4C, (01)00884838067424(21)B37D99, (01)00884838067424(21)F03K4B, (01)00884838067424(21)B3RK4B, (01)00884838067424(21)B200XL, (01)00884838067424(21)B3TXH3, (01)00884838067424(21)B2JX1L, (01)00884838067424(21)F04C24, (01)00884838067424(21)B307RM, (01)00884838067424(21)B2YZWK, (01)00884838067424(21)B3F50C, (01)00884838067424(21)B200W2, (01)00884838067424(21)B35YBN, (01)00884838067424(21)B307RD, (01)00884838067424(21)B30V8H, (01)00884838067424(21)B2C4LF, (01)00884838067424(21)B2C3P7, (01)00884838067424(21)B23VDR, (01)00884838067424(21)B2C529, (01)00884838067424(21)B3KHV3, (01)00884838067424(21)B287TP, (01)00884838061385(21)B0Q58D, (01)00884838061385(21)B14YNH, (01)00884838061385(21)B0BP7R, (01)00884838061385(21)B0K5HN, (01)00884838061385(21)B0T41K, (01)00884838067424(21)B25LTY, (01)00884838067424(21)B29VVP, (01)00884838085176(21)B279QK, (01)00884838085176(21)B1MC5Z, (01)00884838085176(21)B1FN91, (01)00884838085176(21)B1MBQK, 010088483808517621B27BF6, (01)00884838085176(21)B1W19Z, (01)00884838085176(21)B1W1BL, (01)00884838085176(21)B1MCBL, (01)00884838085176(21)B1QHPV, (01)00884838085176(21)B1MC33, (01)00884838085176(21)B25WL3, (01)00884838085176(21)B1FN3L, (01)00884838085176(21)B1W1BG, (01)00884838085176(21)B279L4, 010088483808517621B279Z7, (01)00884838072312(21)B1QJ53, (01)00884838072312(21)B25F17, (01)00884838072312(21)B2395Q, (01)00884838085176(21)B1MC3B, (01)00884838072312(21)B23953, 010088483808517621B1W1FP, (01)00884838086555(21)B2C80T, (01)00884838072312(21)B2394F, (01)00884838085176(21)B279T5, (01)00884838085176(21)B1JNKV, 010088483808517621B1QHCV, (01)00884838072312(21)B1XL56, (01)00884838072312(21)B2394T, (01)00884838072312(21)B1XL4D, (01)00884838085176(21)B1MBT1, (01)00884838085176(21)B1QHKW, (01)00884838085176(21)B27B7C, (01)00884838085176(21)B1QHLG, (01)00884838072312(21)B2393M, (01)00884838061828(21)B1FN5R, (01)00884838085176(21)B2Z0T1, (01)00884838068766(21)B1QHKV, (01)00884838061828(21)B1MC27, (01)00884838061828(21)B1FN4V, (01)00884838068766(21)B1QHPC, (01)00884838085176(21)B1MQQG, (01)00884838061828(21)B1MC31, (01)00884838068766(21)B1QHM3, (01)00884838085176(21)B1MBZK, (01)00884838085176(21)B27BGM, (01)00884838085176(21)B279VD, (01)00884838085176(21)B1W1RW, (01)00884838085176(21)B1QHLK, (01)00884838085176(21)B1MBT8, 010088483808517621B1MC98, (01)00884838085176(21)B27B8P, (01)00884838085176(21)B1MC40, (01)00884838085176(21)B1MBTK; Serial No. B36LWB, B2YZPM, B34JQB, F01L9N, B3V8CH, B3LJYB, B2C4FQ, F010MV, B3VGXG, F03968, B31MT5, B28T00, B2C44Y, B3D56M, B22P69, B3Q81Z, B20145, B3R9R3, B3G9N5, F0174Y, B200TW, B224WW, B3C3W7, B2JX46, B2VM4B, B30VBF, B3WD3C, B39XCG, B2L1X5, B2VLYK, B3LZ50, B3V9LJ, B25N1F, B231L5, B31MT1, B2C3W9, F016D1, B22NVD, F008M1, B1XBNC, B30V7Y, B34JXT, B3QQX0, B2JWTF, B1Y1D9, F02QD4, B34T00, F01K6Z, B3P65C, B224KM, B33706, B32B9T, B2ND6N, B39XDX, B2WDTZ, B2HNP4, B2YZT6, B30CJT, B39ZX8, B200T8, B37D0B, B2WZYX, B2JXDC, B2FY26, B34MHH, B2Q8W4, B2XBFY, B2L1W0, B22P40, B2C5P4, B127JJ, B11G3D, B224KD, B0JJLQ, B0NG16, F019T7, B3H966, F01GFM, F0119H, B39FB8, F002CB, B36B5B, B23VFT, F013HC, B2R8T8, B3G8ZY, B2NDGV, B37D8Y, F005XT, B2Y9YX, B3RKBD, B37D2W, F02QD3, B3V9NL, B1XB2Q, B2C43R, B2L1T9, B3Q81N, B37D34, B34JXC, B3M9F3, B2C5FL, B35YBH, B2M84V, F004KL, B3DQCW, F02DPJ, B2C40T, B224RL, B231GL, B3VGXP, B3RJXK, B224JV, F01C47, B3VCQD, B28RPC, B3QR18, F002CD, B20170, F00LD5, B2C449, B34T3Z, B32B77, B2YZXG, B2M78R, B2HN85, B2YZQF, B25M1W, B38WW3, B39XG1, B2L1Y9, B0T4G8, B1XBJZ, B0HNH9, B0WNP3, B12WHD, B0QVLH, B2BBKQ, B28RNX, B3M9CY, B2JWX4, B200R1, B2VLNB, B25N07, B3VGZP, B38WQ2, B3RKB7, B37DL7, B34JQR, B23W05, B3V9FD, F0103N, F014WR, B36LRF, B35YFJ, B307T7, B31MQ3, B3JQTT, B1XB6C, B3F4T5, B39X4R, B3C3PN, B3F508, B3VCQT, F027KD, B20156, B3RB11, F00LDG, B2BBKL, B3R49J, F00PJ0, B23W0X, F016L9, B2C42R, B217CD, B2NDNG, B3F5DY, B3V9JP, B2HN60, B22P38, B2Z0NZ, B27XBQ, B2L1VZ, B224QQ, B2YZQC, B200VX, B2C4XN, B2JX23, B2VLYV, B38WRX, B2885B, B2M88N, B34MFB, B2Y9NP, B37CZZ, B2YBHP, B09PPC, B0JJR9, B2016G, B0NM53, B07ZMN, B0T1D9, B2HNMQ, B3NTB4, B35YCB, B3M8K5, B3VTKH, B29W4H, B3KT38, B2Y9W4, B224VG, B3DQFC, B2Q8TB, B28GZJ, B1Y1KN, B3M8JM, F01BFY, B224P8, B3MGQ2, B28MJ4, B28RTL, B2XQ1T, B3QZ1R, B2M84T, B224RW, F01C44, F00BR0, B2C4FW, B3777H, F02T5G, B3KHV2, B2018Y, B3RB2L, F01338, B3C4PZ, B3JWP9, B25N0D, B3LZ0Q, F012L3, B1Y18D, B2JWW0, B23VFZ, B3CWQ6, B1Y1HR, B34JZH, B3CM0P, B29JJ4, B37DDK, B2C5JP, F01C4C, B37D99, F03K4B, B3RK4B, B200XL, B3TXH3, B2JX1L, F04C24, B307RM, B2YZWK, B3F50C, B200W2, B35YBN, B307RD, B30V8H, B2C4LF, B2C3P7, B23VDR, B2C529, B3KHV3, B287TP, B0Q58D, B14YNH, B0BP7R, B0K5HN, B0T41K, 03DRXN, 03DMNQ, B08G95, 0387T2, B021BX, 03DQ4W, B25LTY, B29VVP, B279QK, B1MC5Z, B1FN91, B1MBQK, B27BF6, B1W19Z, B1W1BL, B1MCBL, B1QHPV, B1MC33, B25WL3, B1FN3L, B1W1BG, B279L4, B279Z7, B1QJ53, B25F17, B2395Q, B1MC3B, B23953, B1W1FP, B2C80T, B2394F, B279T5, B1JNKV, B1QHCV, B1XL56, B2394T, B1XL4D, B1MBT1, B1QHKW, B27B7C, B1QHLG, B2393M, B1FN5R, B2Z0T1, B1QHKV, B1MC27, B1FN4V, B1QHPC, B1MQQG, B1MC31, B1QHM3, B1MBZK, B27BGM, B279VD, B1W1RW, B1QHLK, B1MBT8, B1MC98, B27B8P, B1MC40, B1MBTK.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Philips Ultrasound, Inc
- Reason for Recall:
- Ultrasound transducer devices were refurbished beyond their useful life.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Philips C5-2 Transducer.
Product Codes/Lot Numbers:
Model No.989605389242, 989605454101, 989605398211, 989605450353, 989605451181, 989605477131, 989605450351, 989605450352; UDI: (01)00884838067424(21)B36LWB, (01)00884838067424(21)B2YZPM, (01)00884838067424(21)B34JQB, (01)00884838067424(21)F01L9N, (01)00884838067424(21)B3V8CH, (01)00884838067424(21)B3LJYB, (01)00884838067424(21)B2C4FQ, (01)00884838067424(21)F010MV, (01)00884838067424(21)B3VGXG, (01)00884838067424(21)F03968, (01)00884838067424(21)B31MT5, (01)00884838067424(21)B28T00, (01)00884838067424(21)B2C44Y, (01)00884838067424(21)B3D56M, (01)00884838067424(21)B22P69, (01)00884838067424(21)B3Q81Z, (01)00884838067424(21)B20145, (01)00884838067424(21)B3R9R3, (01)00884838067424(21)B3G9N5, (01)00884838067424(21)F0174Y, (01)00884838067424(21)B200TW, (01)00884838067424(21)B224WW, (01)00884838067424(21)B3C3W7, (01)00884838067424(21)B2JX46, (01)00884838067424(21)B2VM4B, (01)00884838067424(21)B30VBF, (01)00884838067424(21)B3WD3C, (01)00884838067424(21)B39XCG, (01)00884838067424(21)B2L1X5, (01)00884838067424(21)B2VLYK, (01)00884838067424(21)B3LZ50, (01)00884838067424(21)B3V9LJ, (01)00884838067424(21)B25N1F, (01)00884838067424(21)B231L5, (01)00884838067424(21)B31MT1, (01)00884838067424(21)B2C3W9, (01)00884838067424(21)F016D1, (01)00884838067424(21)B22NVD, (01)00884838067424(21)F008M1, (01)00884838067424(21)B1XBNC, (01)00884838067424(21)B30V7Y, (01)00884838067424(21)B34JXT, (01)00884838067424(21)B3QQX0, (01)00884838067424(21)B2JWTF, (01)00884838067424(21)B1Y1D9, (01)00884838067424(21)F02QD4, (01)00884838067424(21)B34T00, (01)00884838067424(21)F01K6Z, (01)00884838067424(21)B3P65C, (01)00884838067424(21)B224KM, (01)00884838067424(21)B33706, (01)00884838067424(21)B32B9T, (01)00884838067424(21)B2ND6N, (01)00884838067424(21)B39XDX, (01)00884838067424(21)B2WDTZ, (01)00884838067424(21)B2HNP4, (01)00884838067424(21)B2YZT6, (01)00884838067424(21)B30CJT, (01)00884838067424(21)B39ZX8, (01)00884838067424(21)B200T8, (01)00884838067424(21)B37D0B, (01)00884838067424(21)B2WZYX, (01)00884838067424(21)B2JXDC, (01)00884838067424(21)B2FY26, (01)00884838067424(21)B34MHH, (01)00884838067424(21)B2Q8W4, (01)00884838067424(21)B2XBFY, (01)00884838067424(21)B2L1W0, (01)00884838067424(21)B22P40, (01)00884838067424(21)B2C5P4, (01)00884838061385(21)B127JJ, (01)00884838061385(21)B11G3D, (01)00884838067424(21)B224KD, (01)00884838061385(21)B0JJLQ, (01)00884838061385(21)B0NG16, (01)00884838067424(21)F019T7, (01)00884838067424(21)B3H966, (01)00884838067424(21)F01GFM, (01)00884838067424(21)F0119H, (01)00884838067424(21)B39FB8, (01)00884838067424(21)F002CB, (01)00884838067424(21)B36B5B, (01)00884838067424(21)B23VFT, (01)00884838067424(21)F013HC, (01)00884838067424(21)B2R8T8, (01)00884838067424(21)B3G8ZY, (01)00884838067424(21)B2NDGV, (01)00884838067424(21)B37D8Y, (01)00884838067424(21)F005XT, (01)00884838067424(21)B2Y9YX, (01)00884838067424(21)B3RKBD, (01)00884838067424(21)B37D2W, (01)00884838067424(21)F02QD3, (01)00884838067424(21)B3V9NL, (01)00884838067424(21)B1XB2Q, (01)00884838067424(21)B2C43R, (01)00884838067424(21)B2L1T9, (01)00884838067424(21)B3Q81N, (01)00884838067424(21)B37D34, (01)00884838067424(21)B34JXC, (01)00884838067424(21)B3M9F3, (01)00884838067424(21)B2C5FL, (01)00884838067424(21)B35YBH, (01)00884838067424(21)B2M84V, (01)00884838067424(21)F004KL, (01)00884838067424(21)B3DQCW, (01)00884838067424(21)F02DPJ, (01)00884838067424(21)B2C40T, (01)00884838067424(21)B224RL, (01)00884838067424(21)B231GL, (01)00884838067424(21)B3VGXP, (01)00884838067424(21)B3RJXK, (01)00884838067424(21)B224JV, (01)00884838067424(21)F01C47, (01)00884838067424(21)B3VCQD, (01)00884838067424(21)B28RPC, (01)00884838067424(21)B3QR18, (01)00884838067424(21)F002CD, (01)00884838067424(21)B20170, (01)00884838067424(21)F00LD5, (01)00884838067424(21)B2C449, (01)00884838067424(21)B34T3Z, (01)00884838067424(21)B32B77, (01)00884838067424(21)B2YZXG, (01)00884838067424(21)B2M78R, (01)00884838067424(21)B2HN85, (01)00884838067424(21)B2YZQF, (01)00884838067424(21)B25M1W, (01)00884838067424(21)B38WW3, (01)00884838067424(21)B39XG1, (01)00884838067424(21)B2L1Y9, (01)00884838061385(21)B0T4G8, (01)00884838067424(21)B1XBJZ, (01)00884838061385(21)B0HNH9, (01)00884838061385(21)B0WNP3, (01)00884838061385(21)B12WHD, (01)00884838061385(21)B0QVLH, (01)00884838067424(21)B2BBKQ, (01)00884838067424(21)B28RNX, (01)00884838067424(21)B3M9CY, (01)00884838067424(21)B2JWX4, (01)00884838067424(21)B200R1, (01)00884838067424(21)B2VLNB, (01)00884838067424(21)B25N07, (01)00884838067424(21)B3VGZP, (01)00884838067424(21)B38WQ2, (01)00884838067424(21)B3RKB7, (01)00884838067424(21)B37DL7, (01)00884838067424(21)B34JQR, (01)00884838067424(21)B23W05, (01)00884838067424(21)B3V9FD, (01)00884838067424(21)F0103N, (01)00884838067424(21)F014WR, (01)00884838067424(21)B36LRF, (01)00884838067424(21)B35YFJ, (01)00884838067424(21)B307T7, (01)00884838067424(21)B31MQ3, (01)00884838067424(21)B3JQTT, (01)00884838067424(21)B1XB6C, (01)00884838067424(21)B3F4T5, (01)00884838067424(21)B39X4R, (01)00884838067424(21)B3C3PN, (01)00884838067424(21)B3F508, (01)00884838067424(21)B3VCQT, (01)00884838067424(21)F027KD, (01)00884838067424(21)B20156, (01)00884838067424(21)B3RB11, (01)00884838067424(21)F00LDG, (01)00884838067424(21)B2BBKL, (01)00884838067424(21)B3R49J, (01)00884838067424(21)F00PJ0, (01)00884838067424(21)B23W0X, (01)00884838067424(21)F016L9, (01)00884838067424(21)B2C42R, (01)00884838067424(21)B217CD, (01)00884838067424(21)B2NDNG, (01)00884838067424(21)B3F5DY, (01)00884838067424(21)B3V9JP, (01)00884838067424(21)B2HN60, (01)00884838067424(21)B22P38, (01)00884838067424(21)B2Z0NZ, (01)00884838067424(21)B27XBQ, (01)00884838067424(21)B2L1VZ, (01)00884838067424(21)B224QQ, (01)00884838067424(21)B2YZQC, (01)00884838067424(21)B200VX, (01)00884838067424(21)B2C4XN, (01)00884838067424(21)B2JX23, (01)00884838067424(21)B2VLYV, (01)00884838067424(21)B38WRX, (01)00884838067424(21)B2885B, (01)00884838067424(21)B2M88N, (01)00884838067424(21)B34MFB, (01)00884838067424(21)B2Y9NP, (01)00884838067424(21)B37CZZ, (01)00884838067424(21)B2YBHP, (01)00884838061385(21)B09PPC, (01)00884838061385(21)B0JJR9, (01)00884838067424(21)B2016G, (01)00884838061385(21)B0NM53, (01)00884838061385(21)B07ZMN, (01)00884838061385(21)B0T1D9, (01)00884838067424(21)B2HNMQ, (01)00884838067424(21)B3NTB4, (01)00884838067424(21)B35YCB, (01)00884838067424(21)B3M8K5, (01)00884838067424(21)B3VTKH, (01)00884838067424(21)B29W4H, (01)00884838067424(21)B3KT38, (01)00884838067424(21)B2Y9W4, (01)00884838067424(21)B224VG, (01)00884838067424(21)B3DQFC, (01)00884838067424(21)B2Q8TB, (01)00884838067424(21)B28GZJ, (01)00884838067424(21)B1Y1KN, (01)00884838067424(21)B3M8JM, (01)00884838067424(21)F01BFY, (01)00884838067424(21)B224P8, (01)00884838067424(21)B3MGQ2, (01)00884838067424(21)B28MJ4, (01)00884838067424(21)B28RTL, (01)00884838067424(21)B2XQ1T, (01)00884838067424(21)B3QZ1R, (01)00884838067424(21)B2M84T, (01)00884838067424(21)B224RW, (01)00884838067424(21)F01C44, (01)00884838067424(21)F00BR0, (01)00884838067424(21)B2C4FW, (01)00884838067424(21)B3777H, (01)00884838067424(21)F02T5G, (01)00884838067424(21)B3KHV2, (01)00884838067424(21)B2018Y, (01)00884838067424(21)B3RB2L, (01)00884838067424(21)F01338, (01)00884838067424(21)B3C4PZ, (01)00884838067424(21)B3JWP9, (01)00884838067424(21)B25N0D, (01)00884838067424(21)B3LZ0Q, (01)00884838067424(21)F012L3, (01)00884838067424(21)B1Y18D, (01)00884838067424(21)B2JWW0, (01)00884838067424(21)B23VFZ, (01)00884838067424(21)B3CWQ6, (01)00884838067424(21)B1Y1HR, (01)00884838067424(21)B34JZH, (01)00884838067424(21)B3CM0P, (01)00884838067424(21)B29JJ4, (01)00884838067424(21)B37DDK, (01)00884838067424(21)B2C5JP, (01)00884838067424(21)F01C4C, (01)00884838067424(21)B37D99, (01)00884838067424(21)F03K4B, (01)00884838067424(21)B3RK4B, (01)00884838067424(21)B200XL, (01)00884838067424(21)B3TXH3, (01)00884838067424(21)B2JX1L, (01)00884838067424(21)F04C24, (01)00884838067424(21)B307RM, (01)00884838067424(21)B2YZWK, (01)00884838067424(21)B3F50C, (01)00884838067424(21)B200W2, (01)00884838067424(21)B35YBN, (01)00884838067424(21)B307RD, (01)00884838067424(21)B30V8H, (01)00884838067424(21)B2C4LF, (01)00884838067424(21)B2C3P7, (01)00884838067424(21)B23VDR, (01)00884838067424(21)B2C529, (01)00884838067424(21)B3KHV3, (01)00884838067424(21)B287TP, (01)00884838061385(21)B0Q58D, (01)00884838061385(21)B14YNH, (01)00884838061385(21)B0BP7R, (01)00884838061385(21)B0K5HN, (01)00884838061385(21)B0T41K, (01)00884838067424(21)B25LTY, (01)00884838067424(21)B29VVP, (01)00884838085176(21)B279QK, (01)00884838085176(21)B1MC5Z, (01)00884838085176(21)B1FN91, (01)00884838085176(21)B1MBQK, 010088483808517621B27BF6, (01)00884838085176(21)B1W19Z, (01)00884838085176(21)B1W1BL, (01)00884838085176(21)B1MCBL, (01)00884838085176(21)B1QHPV, (01)00884838085176(21)B1MC33, (01)00884838085176(21)B25WL3, (01)00884838085176(21)B1FN3L, (01)00884838085176(21)B1W1BG, (01)00884838085176(21)B279L4, 010088483808517621B279Z7, (01)00884838072312(21)B1QJ53, (01)00884838072312(21)B25F17, (01)00884838072312(21)B2395Q, (01)00884838085176(21)B1MC3B, (01)00884838072312(21)B23953, 010088483808517621B1W1FP, (01)00884838086555(21)B2C80T, (01)00884838072312(21)B2394F, (01)00884838085176(21)B279T5, (01)00884838085176(21)B1JNKV, 010088483808517621B1QHCV, (01)00884838072312(21)B1XL56, (01)00884838072312(21)B2394T, (01)00884838072312(21)B1XL4D, (01)00884838085176(21)B1MBT1, (01)00884838085176(21)B1QHKW, (01)00884838085176(21)B27B7C, (01)00884838085176(21)B1QHLG, (01)00884838072312(21)B2393M, (01)00884838061828(21)B1FN5R, (01)00884838085176(21)B2Z0T1, (01)00884838068766(21)B1QHKV, (01)00884838061828(21)B1MC27, (01)00884838061828(21)B1FN4V, (01)00884838068766(21)B1QHPC, (01)00884838085176(21)B1MQQG, (01)00884838061828(21)B1MC31, (01)00884838068766(21)B1QHM3, (01)00884838085176(21)B1MBZK, (01)00884838085176(21)B27BGM, (01)00884838085176(21)B279VD, (01)00884838085176(21)B1W1RW, (01)00884838085176(21)B1QHLK, (01)00884838085176(21)B1MBT8, 010088483808517621B1MC98, (01)00884838085176(21)B27B8P, (01)00884838085176(21)B1MC40, (01)00884838085176(21)B1MBTK; Serial No. B36LWB, B2YZPM, B34JQB, F01L9N, B3V8CH, B3LJYB, B2C4FQ, F010MV, B3VGXG, F03968, B31MT5, B28T00, B2C44Y, B3D56M, B22P69, B3Q81Z, B20145, B3R9R3, B3G9N5, F0174Y, B200TW, B224WW, B3C3W7, B2JX46, B2VM4B, B30VBF, B3WD3C, B39XCG, B2L1X5, B2VLYK, B3LZ50, B3V9LJ, B25N1F, B231L5, B31MT1, B2C3W9, F016D1, B22NVD, F008M1, B1XBNC, B30V7Y, B34JXT, B3QQX0, B2JWTF, B1Y1D9, F02QD4, B34T00, F01K6Z, B3P65C, B224KM, B33706, B32B9T, B2ND6N, B39XDX, B2WDTZ, B2HNP4, B2YZT6, B30CJT, B39ZX8, B200T8, B37D0B, B2WZYX, B2JXDC, B2FY26, B34MHH, B2Q8W4, B2XBFY, B2L1W0, B22P40, B2C5P4, B127JJ, B11G3D, B224KD, B0JJLQ, B0NG16, F019T7, B3H966, F01GFM, F0119H, B39FB8, F002CB, B36B5B, B23VFT, F013HC, B2R8T8, B3G8ZY, B2NDGV, B37D8Y, F005XT, B2Y9YX, B3RKBD, B37D2W, F02QD3, B3V9NL, B1XB2Q, B2C43R, B2L1T9, B3Q81N, B37D34, B34JXC, B3M9F3, B2C5FL, B35YBH, B2M84V, F004KL, B3DQCW, F02DPJ, B2C40T, B224RL, B231GL, B3VGXP, B3RJXK, B224JV, F01C47, B3VCQD, B28RPC, B3QR18, F002CD, B20170, F00LD5, B2C449, B34T3Z, B32B77, B2YZXG, B2M78R, B2HN85, B2YZQF, B25M1W, B38WW3, B39XG1, B2L1Y9, B0T4G8, B1XBJZ, B0HNH9, B0WNP3, B12WHD, B0QVLH, B2BBKQ, B28RNX, B3M9CY, B2JWX4, B200R1, B2VLNB, B25N07, B3VGZP, B38WQ2, B3RKB7, B37DL7, B34JQR, B23W05, B3V9FD, F0103N, F014WR, B36LRF, B35YFJ, B307T7, B31MQ3, B3JQTT, B1XB6C, B3F4T5, B39X4R, B3C3PN, B3F508, B3VCQT, F027KD, B20156, B3RB11, F00LDG, B2BBKL, B3R49J, F00PJ0, B23W0X, F016L9, B2C42R, B217CD, B2NDNG, B3F5DY, B3V9JP, B2HN60, B22P38, B2Z0NZ, B27XBQ, B2L1VZ, B224QQ, B2YZQC, B200VX, B2C4XN, B2JX23, B2VLYV, B38WRX, B2885B, B2M88N, B34MFB, B2Y9NP, B37CZZ, B2YBHP, B09PPC, B0JJR9, B2016G, B0NM53, B07ZMN, B0T1D9, B2HNMQ, B3NTB4, B35YCB, B3M8K5, B3VTKH, B29W4H, B3KT38, B2Y9W4, B224VG, B3DQFC, B2Q8TB, B28GZJ, B1Y1KN, B3M8JM, F01BFY, B224P8, B3MGQ2, B28MJ4, B28RTL, B2XQ1T, B3QZ1R, B2M84T, B224RW, F01C44, F00BR0, B2C4FW, B3777H, F02T5G, B3KHV2, B2018Y, B3RB2L, F01338, B3C4PZ, B3JWP9, B25N0D, B3LZ0Q, F012L3, B1Y18D, B2JWW0, B23VFZ, B3CWQ6, B1Y1HR, B34JZH, B3CM0P, B29JJ4, B37DDK, B2C5JP, F01C4C, B37D99, F03K4B, B3RK4B, B200XL, B3TXH3, B2JX1L, F04C24, B307RM, B2YZWK, B3F50C, B200W2, B35YBN, B307RD, B30V8H, B2C4LF, B2C3P7, B23VDR, B2C529, B3KHV3, B287TP, B0Q58D, B14YNH, B0BP7R, B0K5HN, B0T41K, 03DRXN, 03DMNQ, B08G95, 0387T2, B021BX, 03DQ4W, B25LTY, B29VVP, B279QK, B1MC5Z, B1FN91, B1MBQK, B27BF6, B1W19Z, B1W1BL, B1MCBL, B1QHPV, B1MC33, B25WL3, B1FN3L, B1W1BG, B279L4, B279Z7, B1QJ53, B25F17, B2395Q, B1MC3B, B23953, B1W1FP, B2C80T, B2394F, B279T5, B1JNKV, B1QHCV, B1XL56, B2394T, B1XL4D, B1MBT1, B1QHKW, B27B7C, B1QHLG, B2393M, B1FN5R, B2Z0T1, B1QHKV, B1MC27, B1FN4V, B1QHPC, B1MQQG, B1MC31, B1QHM3, B1MBZK, B27BGM, B279VD, B1W1RW, B1QHLK, B1MBT8, B1MC98, B27B8P, B1MC40, B1MBTK.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2050-2025
Related Recalls
Lumify Diagnostic Ultrasound System models : 795005, 795161, 989605449841,
Philips Ultrasound
Ultrasound system compatibility issues with Apple devices running iOS 18 may cause a failure to perform live imagining.
TRANSDUCER L12-5 Transducer Probe
Philips Ultrasound
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
C6-2 Transducer Probe
Philips Ultrasound
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.