PSA Assays for use in the IMMULITE/IMMULITE 1000 Systems Analyzers, Catalog Numbers LKPS1, LKPS5 and LKPTS1(D); and in the IMMULITE 2000/2000XPi Systems Analyzers, Catalog Numbers L2KPS2, L2KPS6, L2KPTS2(D) and L2KPTS6(D). IVD Immunoassay kits in the following sizes: 100, 200, 500 and 600 test kits; For in vitro diagnostic use with the IMMULITE/IMMULITE 1000 and the IMMULITE 2000/ 2000 XPi Systems Analyzers. PMA # P930027/ Device Listing # D002374. Product Usage: For the quantitative measurement of prostate-specific antigen (PSA) in human serum, as an aid in the detection of prostate cancer when used in conjunction with digital rectal examination (DRE) in men aged 50 years or older. This assay is further indicated as an adjunctive test to aid in the management of prostate cancer patients. For in vitro diagnostic use with the IMMULITE/IMMULITE 1000 and the IMMULITE 2000/ 2000 XPI Systems Analyzers

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Product Code/Lot # LKPS1 424, Expiration Date 2013-08-31; Product Code/Lot # LKPS1 425, Expiration Date 2013-10-31; Product Code/Lot # LKPS1 426, Expiration Date 2013-10-31; Product Code/Lot # LKPS5 424, Expiration Date 2013-08-31; Product Code/Lot # LKPS5 425, Expiration Date 2013-10-31; Product Code/Lot # LKPS5 426, Expiration Date 2013-10-31; Product Code/Lot # LKPTS1(D) D0104, Expiration Date 2013-08-31; Product Code/Lot # L2KPS2 379, Expiration Date 2013-07-31; Product Code/Lot # L2KPS2 380, Expiration Date 2013-08-31; Product Code/Lot # L2KPS2 381, Expiration Date 2013-10-31; Product Code/Lot # L2KPS2 382, Expiration Date 2013-10-31; Product Code/Lot # L2KPS2 383, Expiration Date 2013-10-31; Product Code/Lot # L2KPS2 384, Expiration Date 2013-10-31; Product Code/Lot # L2KPS6 379, Expiration Date 2013-07-31; Product Code/Lot # L2KPS6 380, Expiration Date 2013-08-31; Product Code/Lot # L2KPS6 381, Expiration Date 2013-10-31; Product Code/Lot # L2KPS6 382, Expiration Date 2013-10-31; Product Code/Lot # L2KPS6 383, Expiration Date 2013-10-31; Product Code/Lot # L2KPS6 384, Expiration Date 2013-10-31; Product Code/Lot # L2KPTS2(D) D107, Expiration Date 2013-07-31; Product Code/Lot # L2KPTS2(D) D108, Expiration Date 2013-10-31; Product Code/Lot # L2KPTS2(D) D109, Expiration Date 2013-10-31; Product Code/Lot # L2KPTS6(D) D107, Expiration Date 2013-07-31; Product Code/Lot # L2KPTS6(D) D108, Expiration Date 2013-10-31; Product Code/Lot # L2KPTS6(D) D109, Expiration Date 2013-10-31.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Siemens Healthcare Diagnostics
Reason for Recall:
Customers complained of a positive bias in patient results on the IMMULITE/ IMMULITE 1000 and the IMMULITE 2000/ 2000 XPi PSA assay when compared against the IMMULITE /IMMULITE 1000 3rd Generation PSA assay. Siemens has confirmed an overall average positive bias of approximately 20% to 23% across the assay range relative to the WHO 96/670 with the IMMMULITE Systems PSA Assay. This positive bias is observed in patient values and the Siemens Tumor Marker controls (TMCO). Other commercially available controls may show this bias.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

PSA Assays for use in the IMMULITE/IMMULITE 1000 Systems Analyzers, Catalog Numbers LKPS1, LKPS5 and LKPTS1(D); and in the IMMULITE 2000/2000XPi Systems Analyzers, Catalog Numbers L2KPS2, L2KPS6, L2KPTS2(D) and L2KPTS6(D). IVD Immunoassay kits in the following sizes: 100, 200, 500 and 600 test kits; For in vitro diagnostic use with the IMMULITE/IMMULITE 1000 and the IMMULITE 2000/ 2000 XPi Systems Analyzers. PMA # P930027/ Device Listing # D002374. Product Usage: For the quantitative measurement of prostate-specific antigen (PSA) in human serum, as an aid in the detection of prostate cancer when used in conjunction with digital rectal examination (DRE) in men aged 50 years or older. This assay is further indicated as an adjunctive test to aid in the management of prostate cancer patients. For in vitro diagnostic use with the IMMULITE/IMMULITE 1000 and the IMMULITE 2000/ 2000 XPI Systems Analyzers

Product Codes/Lot Numbers:

Product Code/Lot # LKPS1 424, Expiration Date 2013-08-31; Product Code/Lot # LKPS1 425, Expiration Date 2013-10-31; Product Code/Lot # LKPS1 426, Expiration Date 2013-10-31; Product Code/Lot # LKPS5 424, Expiration Date 2013-08-31; Product Code/Lot # LKPS5 425, Expiration Date 2013-10-31; Product Code/Lot # LKPS5 426, Expiration Date 2013-10-31; Product Code/Lot # LKPTS1(D) D0104, Expiration Date 2013-08-31; Product Code/Lot # L2KPS2 379, Expiration Date 2013-07-31; Product Code/Lot # L2KPS2 380, Expiration Date 2013-08-31; Product Code/Lot # L2KPS2 381, Expiration Date 2013-10-31; Product Code/Lot # L2KPS2 382, Expiration Date 2013-10-31; Product Code/Lot # L2KPS2 383, Expiration Date 2013-10-31; Product Code/Lot # L2KPS2 384, Expiration Date 2013-10-31; Product Code/Lot # L2KPS6 379, Expiration Date 2013-07-31; Product Code/Lot # L2KPS6 380, Expiration Date 2013-08-31; Product Code/Lot # L2KPS6 381, Expiration Date 2013-10-31; Product Code/Lot # L2KPS6 382, Expiration Date 2013-10-31; Product Code/Lot # L2KPS6 383, Expiration Date 2013-10-31; Product Code/Lot # L2KPS6 384, Expiration Date 2013-10-31; Product Code/Lot # L2KPTS2(D) D107, Expiration Date 2013-07-31; Product Code/Lot # L2KPTS2(D) D108, Expiration Date 2013-10-31; Product Code/Lot # L2KPTS2(D) D109, Expiration Date 2013-10-31; Product Code/Lot # L2KPTS6(D) D107, Expiration Date 2013-07-31; Product Code/Lot # L2KPTS6(D) D108, Expiration Date 2013-10-31; Product Code/Lot # L2KPTS6(D) D109, Expiration Date 2013-10-31.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2047-2013

Related Recalls

The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.

Oct 6, 2025 Other Medical Devices Nationwide View Details →

The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.

Oct 6, 2025 Infusion Pumps Nationwide View Details →

The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.

Oct 6, 2025 Other Medical Devices Nationwide View Details →