Canon Vantage Titan 3T, Premium Open-Bore MRI System, REF: MRT-3010 - Product Usage: use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Affected Serial Numbers: A5B1182002, A5B11Y2005, A5D14X2032, A5A1082001, A5C12Y2017, A5D13Y2025, A5C1412028, S7A1642006, A5D13Y2026, A5F1562034, A5B1192004, A5C1222010, A5C1212009, A5C12X2016, A5G1692038, A5G1642036, A5C1232011, A5C1252012, A5D13Z2027, A5D1492030, A5C1262013, A5E1522033, A5G1692039, A5D14X2031, A5B1192003, A5B11Y2006, A5D1312019, A5G1652037, A5D1362021, A5C1292015, A5D13X2024, A5C1422029, A5D1322020, A5B11Z2007, A5C1212008, A5C12Z2018, A5D1372022
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Canon Medical System, USA, INC.
Reason for Recall:
The MRI system contains a component that does not meet the appropriate flame retardancy-rating which could result in smoke or fire.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Canon Vantage Titan 3T, Premium Open-Bore MRI System, REF: MRT-3010 - Product Usage: use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.

Product Codes/Lot Numbers:

Affected Serial Numbers: A5B1182002, A5B11Y2005, A5D14X2032, A5A1082001, A5C12Y2017, A5D13Y2025, A5C1412028, S7A1642006, A5D13Y2026, A5F1562034, A5B1192004, A5C1222010, A5C1212009, A5C12X2016, A5G1692038, A5G1642036, A5C1232011, A5C1252012, A5D13Z2027, A5D1492030, A5C1262013, A5E1522033, A5G1692039, A5D14X2031, A5B1192003, A5B11Y2006, A5D1312019, A5G1652037, A5D1362021, A5C1292015, A5D13X2024, A5C1422029, A5D1322020, A5B11Z2007, A5C1212008, A5C12Z2018, A5D1372022

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2046-2021

Related Recalls

Class I - Dangerous

It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed

Dec 19, 2025 Other Medical Devices Nationwide View Details →
Class I - Dangerous

There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.

Dec 19, 2025 Other Medical Devices Nationwide View Details →
Class I - Dangerous

There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.

Dec 19, 2025 Other Medical Devices Nationwide View Details →