Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number: (1) 6802413, (2) 6802915

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    VITROS 5600 Integrated System; Product Code: 6802413; UDI: 10758750002740; VITROS 5600 Integrated System - Certified/Refurbished; Product Code: 6802915; UDI: 10758750002740; Range of installed serial numbers: 56000143-56005096 Note1: The catalogue number 6802413 is used to denote the VITROS 5600 Integrated System. Catalogue number 6802915 is used to denote those VITROS 5600 Integrated Systems that have been refurbished, these systems can also be referred to as certified. A VITROS 5600 Integrated System maintains its original serial number through the refurbishment/certification process.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Ortho-Clinical Diagnostics, Inc.
Reason for Recall:
A software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the system treats as expired even when the Use Expired Reagents setting is not enabled if a shelf expiration date is omitted during manual loading. The issue was identified by a customer complaint, in which a CREA Urine test was processed with an RE (Reagent Expired) code, even though the diluent pack was not expired and the use-expired-reagents setting was disabled. This can result in erroneous test results and/or delaying test results.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number: (1) 6802413, (2) 6802915

Product Codes/Lot Numbers:

VITROS 5600 Integrated System; Product Code: 6802413; UDI: 10758750002740; VITROS 5600 Integrated System - Certified/Refurbished; Product Code: 6802915; UDI: 10758750002740; Range of installed serial numbers: 56000143-56005096 Note1: The catalogue number 6802413 is used to denote the VITROS 5600 Integrated System. Catalogue number 6802915 is used to denote those VITROS 5600 Integrated Systems that have been refurbished, these systems can also be referred to as certified. A VITROS 5600 Integrated System maintains its original serial number through the refurbishment/certification process.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2038-2025

Related Recalls

VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.

Jul 29, 2025 Diagnostic Equipment Nationwide View Details →

VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.

Jul 29, 2025 Diagnostic Equipment Nationwide View Details →
Class I - Dangerous

Generations 67 and above of VITROS Ca Slides are experiencing an increased occurrence of Condition Code TH4-63J during calibration on VITROS XT 3400 and VITROS XT 7600 Systems, causing calibration failures that delay calcium test results and potentially impact patient management. The imaging reflectometer algorithm detects excessive "spikes" (imperfections) in slide images and generates the TH4-63J condition code when the threshold of 600 spikes is exceeded. Updated analysis revealed a 19-fold increase in failure rates from 0.0043% (2024 baseline) to 0.082% for Generation 67+ slides in 2025, with 100% of calibrator failures occurring at the lowest concentration level. This can result in delayed diagnosis and treatment of conditions such as hypercalcemic crisis, parathyroid disorders, and other calcium-related emergencies requiring urgent intervention.

Jul 9, 2025 Other Medical Devices Nationwide View Details →