Sagittal MICRO-SAW

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    All serial numbers shipped to customers prior to 09/27/2016; DFU 18-0095; Part No. S-SAW.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
The Anspach Effort, Inc.
Reason for Recall:
Supplied Directions for Use (DFU) did not contain a recommended service interval.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Sagittal MICRO-SAW

Product Codes/Lot Numbers:

All serial numbers shipped to customers prior to 09/27/2016; DFU 18-0095; Part No. S-SAW.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2031-2017

Related Recalls

Class I - Dangerous

Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury

Dec 7, 2022 Other Medical Devices Nationwide View Details →

Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury

Dec 7, 2022 Other Medical Devices Nationwide View Details →