OEC 9800, OEC FluoroTrak 9800 Plus, OEC 9800 Plus, OEC 9800 MD Motorized C-arm System designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. The OEC 9800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. The system includes features specifically designed for use in diagnostic and interventional cardiac imaging procedures, and is also intended for cholangiography, endoscopic, urologic, orthopedic, neurologic, vascular, critical care and emergency room procedures. It may be used for other imaging applications at the physician's discretion.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Part numbers: 00-884909-01, 00-884777-01, 00-884911-01, 00-885553-01, 00-884913-01, 00-885554-01, 00-885555-01, 00-886289-01. All serial numbers.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- GE OEC Medical Systems, Inc
- Reason for Recall:
- Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC recalled imaging devices OEC 9900, OEC 9800, and OEC 8800 as a result of an FDA inspection identifying that the vertical lift column power supply in the mainframe C-arm is defective and subject to early life failure.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
OEC 9800, OEC FluoroTrak 9800 Plus, OEC 9800 Plus, OEC 9800 MD Motorized C-arm System designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. The OEC 9800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. The system includes features specifically designed for use in diagnostic and interventional cardiac imaging procedures, and is also intended for cholangiography, endoscopic, urologic, orthopedic, neurologic, vascular, critical care and emergency room procedures. It may be used for other imaging applications at the physician's discretion.
Product Codes/Lot Numbers:
Part numbers: 00-884909-01, 00-884777-01, 00-884911-01, 00-885553-01, 00-884913-01, 00-885554-01, 00-885555-01, 00-886289-01. All serial numbers.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2026-2012
Related Recalls
Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.
Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.
Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.