Ra Medical Systems DABRA Catheter 5F (1.5mm), REF: 1222-5000-01, LOT: 18-090400-094, USE BY: 2019-09-24, MFR DATE: 2018-09-24, CE, Laser Radiation, STERILE (Gamma Sterilized), Single Use Only Revised Label: Product Label - Ra Medical Systems DABRA Catheter 5F (1.5mm), REF: 1222-5000-01, LOT: 19-092400-041, USE BY: 2019-12-13, MFR DATE: 2019-10-14, CE 2797, Laser Radiation, STERILE (Gamma Sterilized), Single Use Only Package Label - DABRA 101 5 French, Ref: 1222-5000-01, Lot: LOT: 19-092400-041, USE BY: 2019-12-13, UDI's (01)00842362100031, (17)20191213(21)2019092400041

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    DABRA 101 Catheter. Ref: 1222-5000-01 Serial Number(s): Manufacturing dates 2018-09-03 to 2019-08-26. Expiration dates: September 2019 to August 2020
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Ra Medical Systems, Inc.
Reason for Recall:
The firm has become aware there is a potential problem with its DABRA catheters resulting in higher catheter calibration failure rate due to the product shelf life is currently less than the labeled shelf life.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Ra Medical Systems DABRA Catheter 5F (1.5mm), REF: 1222-5000-01, LOT: 18-090400-094, USE BY: 2019-09-24, MFR DATE: 2018-09-24, CE, Laser Radiation, STERILE (Gamma Sterilized), Single Use Only Revised Label: Product Label - Ra Medical Systems DABRA Catheter 5F (1.5mm), REF: 1222-5000-01, LOT: 19-092400-041, USE BY: 2019-12-13, MFR DATE: 2019-10-14, CE 2797, Laser Radiation, STERILE (Gamma Sterilized), Single Use Only Package Label - DABRA 101 5 French, Ref: 1222-5000-01, Lot: LOT: 19-092400-041, USE BY: 2019-12-13, UDI's (01)00842362100031, (17)20191213(21)2019092400041

Product Codes/Lot Numbers:

DABRA 101 Catheter. Ref: 1222-5000-01 Serial Number(s): Manufacturing dates 2018-09-03 to 2019-08-26. Expiration dates: September 2019 to August 2020

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2006-2020

Related Recalls

Due to footswitch not meeting required specification for protection from ingress of solids or liquids. If spilling of liquids near or on the footswitch, it may fail in the 'on' position and allow the laser to continue to lase.

Oct 16, 2020 Infusion Pumps View Details →

A software issue was identified which could result in user or patient injury, or may adversely impact laser performance. It is possible that an unintended release of laser light (radiation) may occur, injuring the user or patient.