00597201101 Femur CR POROUS FEM COMP SIZE AML Rx, Sterile; 00597201102 Femur CR POROUS FEM COMP SIZE AMR Rx, Sterile; 00597201201 Femur CR POROUS FEM COMP SIZE BML Rx, Sterile; 00597201202 Femur CR POROUS FEM COMP SIZE BMR Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    all codes
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

āš ļø Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Zimmer, Inc.
Reason for Recall:
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

00597201101 Femur CR POROUS FEM COMP SIZE AML Rx, Sterile; 00597201102 Femur CR POROUS FEM COMP SIZE AMR Rx, Sterile; 00597201201 Femur CR POROUS FEM COMP SIZE BML Rx, Sterile; 00597201202 Femur CR POROUS FEM COMP SIZE BMR Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.

Product Codes/Lot Numbers:

all codes

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2000-2013

Related Recalls

Zimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-EĀæ Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is a Size G 46 mm, and vice versa.

Sep 26, 2025 Implants & Prosthetics Nationwide View Details →

Zimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-EĀæ Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is a Size G 46 mm, and vice versa.

Sep 26, 2025 Implants & Prosthetics Nationwide View Details →