Heraeus, PALAMIX uno. Material Number: 66057893.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Material Number: 66057893; UDI-DI: 04260102133587; Batch: all batches with numbers below 223060;
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Heraeus Medical GmbH (Dental Division)
- Reason for Recall:
- Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Heraeus, PALAMIX uno. Material Number: 66057893.
Product Codes/Lot Numbers:
Material Number: 66057893; UDI-DI: 04260102133587; Batch: all batches with numbers below 223060;
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1997-2026
Related Recalls
Heraeus, PALAMIX duo. Material Number: 66057897.
Heraeus Medical GmbH (Dental Division)
Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).
Brand Name: PALACOS Product Name: PALACOS R+G pro 40; PALACOS R+G pro 80 Model/Catalog Number: 5081288; 5081289 Software Version: N/A Product Description: Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system, suitable for use with or without vacuum (ready to mix). Component: N/A
Heraeus Medical GmbH (Dental Division)
increase of complaints regarding the breakage of the ampoules within their device systems. An extensive investigation revealed that multiple factors have an impact on the ampoule breakage. An unfortunate combination of these influence factors can lead to unbreakable ampoules. If none or not all ampoules break, the product cannot be used on the patient since no or no appropriate cement dough is formed.