Physician Order Form (PL-15400-00, Nov. 2018) distributed to end user used with the following: Prometra Programmable Infusion Pump System: MODEL/CATALOG #(s): Catalog No. US Model GTIN-14 11827 Prometra Programmable 20mL Pump 00810335020082 13827 Prometra II Programmable 20mLPump 00810335020228 16827 Prometra II Programmable 40mLPump 00810335020242
Class I - DangerousWhat Should You Do?
- Check if you have this product: All Physician Order Form (PL-15400-00, Nov. 2018)
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Flowonix Medical Inc
- Reason for Recall:
- Physician Order Form (PL-15400-00 Nov 2018) distributed is discontinued due to the text 'Refill the pump and program refill with the same medication removed prior to MRI' may be misinterpreted as 'reuse the same drug solution extracted from the pump prior to the MRI procedure', creating a possible risk of infection
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Physician Order Form (PL-15400-00, Nov. 2018) distributed to end user used with the following: Prometra Programmable Infusion Pump System: MODEL/CATALOG #(s): Catalog No. US Model GTIN-14 11827 Prometra Programmable 20mL Pump 00810335020082 13827 Prometra II Programmable 20mLPump 00810335020228 16827 Prometra II Programmable 40mLPump 00810335020242
Product Codes/Lot Numbers:
All Physician Order Form (PL-15400-00, Nov. 2018)
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1996-2021
Related Recalls
Prometra II 20mL Pump. Catalog No. 13827 The device is an implantable infusion pump. GTIN-14 00810335020228
Flowonix Medical
A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system.
A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system.
US Clinical 20mL Programmable Pump. Catalog No. 01827 The device is an implantable infusion pump. GTIN-14 00810335020082
Flowonix Medical
A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system.