Medtronic Nexframe Stereotactic System and StealthStation Cranial software version 3.0 or newer with StealthStation DBS License or StealthStation S8 Software with Stealth DBS License and O-arm Imaging System utilizing the auto-registration (fiducial- less) workflow used in combination during a DBS (deep brain stimulation) procedure. Neurological stereotaxic Instrument.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Medtronic Nexframe Stereotactic System and StealthStation Cranial software version 3.0 or newer with StealthStation DBS License or StealthStation S8 Software with Stealth DBS License and O-arm Imaging System utilizing the auto-registration (fiducial- less) workflow used in combination during a DBS procedure. Stealthware Station Cranial Software GTIN(s): 00643169878204, 00643169625549 O-Arm Imaging System GTIN(s): 00763000068073, 00763000068080, 00763000068097, 00763000068103, 00763000074050, 00763000074067, 00763000074074, 00763000074081, 00763000074098, 00763000074104, 00763000074111, 00763000074128, 00763000074135, 00763000079000, 00763000079017, 00763000079024, 00763000079031, 00763000079048, 00763000079055, 00763000079062, 00763000079079, 00763000079086, 00763000079093, 00763000079109, 00763000079215, 00763000079246, 00763000079260, 00763000080969, 00763000080976, 00763000080983, 00763000080990, 00763000081003, 00763000081010, 00763000081027, 00763000081034, 00763000081041, 00763000081058, 00763000081065, 00763000081072, 00763000081089, 00763000081096 and 00763000081102. Nexframe Stereotactic System GTIN(s): DB-2040, 00643169214811; MI-1000, 00643169986763; MI-2000, 00643169986770; NP-1000, 00613994747877.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Medtronic Navigation, Inc.
- Reason for Recall:
- Entry point and lead placement inaccuracies during deep brain stimulation lead implantation procedures may occur when using a specific combination of the firm's Steriotactic System and auto-registration feature with a specific imaging system (also known as a fiducial-less procedure). Minor patient movement may not be initially detected by the user or the software during the auto-registration scan process potentially resulting in inaccuracies and risks for the patient including: inaccurate lead placement, delay of surgery, aborted surgery, or additional intervention (including revision of the lead placement and subsequent imaging).
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Medtronic Nexframe Stereotactic System and StealthStation Cranial software version 3.0 or newer with StealthStation DBS License or StealthStation S8 Software with Stealth DBS License and O-arm Imaging System utilizing the auto-registration (fiducial- less) workflow used in combination during a DBS (deep brain stimulation) procedure. Neurological stereotaxic Instrument.
Product Codes/Lot Numbers:
Medtronic Nexframe Stereotactic System and StealthStation Cranial software version 3.0 or newer with StealthStation DBS License or StealthStation S8 Software with Stealth DBS License and O-arm Imaging System utilizing the auto-registration (fiducial- less) workflow used in combination during a DBS procedure. Stealthware Station Cranial Software GTIN(s): 00643169878204, 00643169625549 O-Arm Imaging System GTIN(s): 00763000068073, 00763000068080, 00763000068097, 00763000068103, 00763000074050, 00763000074067, 00763000074074, 00763000074081, 00763000074098, 00763000074104, 00763000074111, 00763000074128, 00763000074135, 00763000079000, 00763000079017, 00763000079024, 00763000079031, 00763000079048, 00763000079055, 00763000079062, 00763000079079, 00763000079086, 00763000079093, 00763000079109, 00763000079215, 00763000079246, 00763000079260, 00763000080969, 00763000080976, 00763000080983, 00763000080990, 00763000081003, 00763000081010, 00763000081027, 00763000081034, 00763000081041, 00763000081058, 00763000081065, 00763000081072, 00763000081089, 00763000081096 and 00763000081102. Nexframe Stereotactic System GTIN(s): DB-2040, 00643169214811; MI-1000, 00643169986763; MI-2000, 00643169986770; NP-1000, 00613994747877.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1995-2020
Related Recalls
Due to an increase in complaint that the percutaneous reference pin would not fit into the patient reference frame or percutaneous pin adapter.
Due to two software anomalies that can occur within a spine procedure. Two issues are: 1. Navigation Orientation - when lateral orientation radiographic images are taken, information may appear flipped when utilizing the system (180 degree flipped). 2. Pop-up Message - when using CT+Fluouroscopy images with a slice spacing thickness greater than 2mm, the system does not display a pop-up message indicating the slice spacing is not optimal.
Due to a software issue, there is the potential for a missing digit, letter, or decimal in either the "distance to target" or "tip stop point" display during navigation.