BD FACS Canto and associated Fluidics Cart Product Usage: Flow cytometer for use in In vitro Diagnostics.
Class I - DangerousWhat Should You Do?
- Check if you have this product: FACS Canto: model number 657338; Associates Fluidic Cart (catalog number 6657660)- IVD; Serial numbers: V65733800001, V65733800002, V65733800003, V65733800004 V65733800005, V65733800006, V65733800007, V65733800008, V65733800009, V657338000010, V657338000011, V657338000012, V657338000013, V657338000014, V657338000017, V657338000018, V657338000019, V657338000020, V657338000021, V657338000022, V657338000023, V657338000024, V657338000025, V657338000026, V657338000028, V657338000029, V657338000031, V657338000032, V657338000033, V657338000034, V65733800035, V65733800036, V65733800037, V65733800038, V65733800039, V657338000040 V657338000043, V657338000044, V657338000045, V657338000046, V657338000047, V657338000048, V657338000049, V657338000050, V657338000051, V657338000052, V657338000053, V657338000054, V657338000055, V657338000056, V657338000057, V657338000058, V657338000060.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- BD Biosciences, Systems & Reagents
- Reason for Recall:
- Laser cooling fans and internal fan of the associated fluidics cart are not operational.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
BD FACS Canto and associated Fluidics Cart Product Usage: Flow cytometer for use in In vitro Diagnostics.
Product Codes/Lot Numbers:
FACS Canto: model number 657338; Associates Fluidic Cart (catalog number 6657660)- IVD; Serial numbers: V65733800001, V65733800002, V65733800003, V65733800004 V65733800005, V65733800006, V65733800007, V65733800008, V65733800009, V657338000010, V657338000011, V657338000012, V657338000013, V657338000014, V657338000017, V657338000018, V657338000019, V657338000020, V657338000021, V657338000022, V657338000023, V657338000024, V657338000025, V657338000026, V657338000028, V657338000029, V657338000031, V657338000032, V657338000033, V657338000034, V65733800035, V65733800036, V65733800037, V65733800038, V65733800039, V657338000040 V657338000043, V657338000044, V657338000045, V657338000046, V657338000047, V657338000048, V657338000049, V657338000050, V657338000051, V657338000052, V657338000053, V657338000054, V657338000055, V657338000056, V657338000057, V657338000058, V657338000060.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1977-2014
Related Recalls
Instruments installed between June 2007 and September 2011 do not contain labeling pursuant to the Food and Drug Administration's (FDA's) Guidance on Laser Products- Conformance with IEC 60825-1 and IEC 60601-2-22 (Laser Notice No. 50).
CD123 (9F5) PE Catalog number 649453 Analyte Specific Reagent
BD Biosciences, Systems & Reagents
One lot of BD CD123 PE (ASR) has been determined to contain a low amount of CD4 antibody and may result in an unexpected staining pattern.
BD FACSCalibur; Catalog Numbers 342973, 342975, 34976 In vitro Diagnostics Use.
BD Biosciences, Systems & Reagents
The circuit breakers on some BD FACSCalibur instruments may have the incorrect rated amperage (20 amps instead of 2 amps) causing the circuit breaker to fail to open in an over-current situation.